DXA body composition corrective factors between Hologic Discovery models to conduct multicenter studies.

Thibault Sutter1, François Duboeuf2, Roland Chapurlat3, Bernard Cortet4, Eric Lespessailles5, Jean-Paul Roux3.   

Abstract

BACKGROUND: Dual X-ray absorptiometry body composition measurements are widely used for clinical and research settings. It is well known that measurements vary across instruments, needing caution for longitudinal monitoring or multicentric studies. This study was to quantify intra- and inter-center variability of bone mineral content, bone mineral density, fat and lean body composition measurements between Hologic Discovery models in order to calculate the corrective factors to be applied for multicenter research projects.
MATERIALS AND METHODS: A whole body phantom composed of materials representing the thickness and percentage of bone, lean and fat mass in the human physiological range was analyzed ten times in three different centers using dual energy X-ray absorptiometry scanners (Two Hologic Discovery QDR A and one QDR W). In addition, we used a morphometric vertebral phantom to monitor stability and a three steps block phantom to check accuracy.
RESULTS: We found a good long-term stability and accuracy for the three devices. Intra-center coefficients of variation were within the range of the manufacturer acceptable values (bone mineral density: 1.40%, bone mineral content: 1%, area: 1.50%, fat mass: 0.89%, lean mass: 0.76%, total mass: 0.12%). Whereas the inter-center coefficient of variation exceeded 8% (bone mineral density: 8.18%, bone mineral content: 3.03%, area: 8.63%: fat mass: 3,92%, lean mass: 7.89%, total mass: 2.85%).
CONCLUSION: Our study showed that the discrepancies across centers remain a major concern, particularly with regard to body composition results. Our study highlight the need of cross calibration between densitometers and proposes corrective factors evaluated from a whole body phantom to lead multicentric studies adjustment.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bone mineral density; Cross-calibration; DXA; Whole body composition

Mesh:

Year:  2020        PMID: 33045389     DOI: 10.1016/j.bone.2020.115683

Source DB:  PubMed          Journal:  Bone        ISSN: 1873-2763            Impact factor:   4.398


  4 in total

1.  Circulating Carboxylated Osteocalcin Correlates With Skeletal Muscle Mass and Risk of Fall in Postmenopausal Osteoporotic Women.

Authors:  Jacopo Antonino Vitale; Veronica Sansoni; Martina Faraldi; Carmelo Messina; Chiara Verdelli; Giovanni Lombardi; Sabrina Corbetta
Journal:  Front Endocrinol (Lausanne)       Date:  2021-05-05       Impact factor: 5.555

2.  Impact of appendicular and trunk skeletal muscle mass and back extensor strength on sagittal spinal alignment in Japanese women without vertebral fracture.

Authors:  Daisuke Kudo; Naohisa Miyakoshi; Michio Hongo; Yuji Kasukawa; Yoshinori Ishikawa; Takashi Mizutani; Yoichi Mizutani; Yoichi Shimada
Journal:  Osteoporos Sarcopenia       Date:  2021-03-04

3.  Comparison of Two DXA Systems, Hologic Horizon W and GE Lunar Prodigy, for Assessing Body Composition in Healthy Korean Adults.

Authors:  Seung Shin Park; Soo Lim; Hoyoun Kim; Kyoung Min Kim
Journal:  Endocrinol Metab (Seoul)       Date:  2021-12-16

4.  Musclin Is Related to Insulin Resistance and Body Composition, but Not to Body Mass Index or Cardiorespiratory Capacity in Adults.

Authors:  Yeliana L Sánchez; Manuela Yepes-Calderón; Luis Valbuena; Andrés F Milán; María C Trillos-Almanza; Sergio Granados; Miguel Peña; Mauricio Estrada-Castrillón; Juan C Aristizábal; Raúl Narvez-Sanchez; Jaime Gallo-Villegas; Juan C Calderón
Journal:  Endocrinol Metab (Seoul)       Date:  2021-10-21
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.