A 1.5-D Array for Acoustic Radiation Force (ARF)-Induced Peak Displacement-Based Tissue Anisotropy Assessment With a Row-Column Excitation Method.

Huaiyu Wu, Md Murad Hossain, Howuk Kim, Caterina M Gallippi, Xiaoning Jiang.   

Abstract

Many biological tissues, including muscle or kidney, are mechanically anisotropic, and the degree of anisotropy (DoA) in mechanical properties is diagnostically relevant. DoA can be assessed either using the ratio of shear wave velocities (SWVs) or acoustic radio forced impulse (ARFI)-induced peak displacements (PD) measured longitudinal over transverse orientations. Whether using SWV or PD as a basis, DoA expressed as the ratio of values requires 90° transducer rotation when a linear array is employed. This large rotation angle is prone to misalignment errors. One solution is the use of a fully sampled matrix array for electronic rotation of point spread function (PSF). However, the challenges of matrix array are its high fabrication cost and complicated fabrication procedures. The cheaper and simpler alternative of matrix array is the use of a row-column array. A 3×64 elements 1.5-D array with a row-column excitation mode is proposed to assess DoA in mechanical properties using the PD ratio. Different numbers of elements in elevational and lateral directions were selected to have orthogonal ARFI excitation beams without rotating the transducer. A custom-designed flex circuit was used to fabricate the array with a simpler electrode connection than a fully sampled matrix array. The performance of the array was evaluated in Field II simulation and experiment. The output pressure was 0.57-MPa output under a 40- [Formula: see text] excitation with a -6-dB point spread dimension of 14×4 mm2 in orthogonal directions. The PD was measured to be [Formula: see text] in an isotropic elastic phantom with Young's modulus of 5.4 kPa. These results suggest that the array is capable of assessing DoA using PD ratio without physical rotation of the transducer. The array has the potential to reduce the misalignment errors for DoA assessment.

Entities:  

Mesh:

Year:  2021        PMID: 33044921      PMCID: PMC8080255          DOI: 10.1109/TUFFC.2020.3030040

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  21 in total

1.  Evaluation of skin viscoelasticity and anisotropy by measurement of speed of shear wave propagation with viscoelasticity skin analyzer.

Authors:  A Vexler; I Polyansky; R Gorodetsky
Journal:  J Invest Dermatol       Date:  1999-11       Impact factor: 8.551

2.  Rapid tracking of small displacements with ultrasound.

Authors:  Gianmarco F Pinton; Jeremy J Dahl; Gregg E Trahey
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2006-06       Impact factor: 2.725

3.  Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers.

Authors:  J A Jensen; N B Svendsen
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  1992       Impact factor: 2.725

4.  A large aperture row column addressed probe for in vivo 4D ultrafast doppler ultrasound imaging.

Authors:  J Sauvage; M Flesch; G Férin; A Nguyen-Dinh; J Porée; M Tanter; M Pernot; T Deffieux
Journal:  Phys Med Biol       Date:  2018-10-24       Impact factor: 3.609

5.  Tissue mimicking materials for ultrasound phantoms.

Authors:  E L Madsen; J A Zagzebski; R A Banjavie; R E Jutila
Journal:  Med Phys       Date:  1978 Sep-Oct       Impact factor: 4.071

6.  Expression of full-length utrophin prevents muscular dystrophy in mdx mice.

Authors:  J Tinsley; N Deconinck; R Fisher; D Kahn; S Phelps; J M Gillis; K Davies
Journal:  Nat Med       Date:  1998-12       Impact factor: 53.440

7.  Diffusion tensor MRI: preliminary anisotropy measures and mapping of breast tumors.

Authors:  Savannah C Partridge; Ali Ziadloo; Revathi Murthy; Steven W White; Sue Peacock; Peter R Eby; Wendy B DeMartini; Constance D Lehman
Journal:  J Magn Reson Imaging       Date:  2010-02       Impact factor: 4.813

8.  Electronic Point Spread Function Rotation Using a Three-Row Transducer for ARFI-Based Elastic Anisotropy Assessment: In Silico and Experimental Demonstration.

Authors:  Md Murad Hossain; Caterina M Gallippi
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2021-02-25       Impact factor: 2.725

9.  Mechanical Anisotropy Assessment in Kidney Cortex Using ARFI Peak Displacement: Preclinical Validation and Pilot In Vivo Clinical Results in Kidney Allografts.

Authors:  Md Murad Hossain; Randal K Detwiler; Emily H Chang; Melissa C Caughey; Melrose W Fisher; Timothy C Nichols; Elizabeth P Merricks; Robin A Raymer; Margaret Whitford; Dwight A Bellinger; Lauren E Wimsey; Caterina M Gallippi
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-08-13       Impact factor: 2.725

10.  On the Quantitative Potential of Viscoelastic Response (VisR) Ultrasound Using the One-Dimensional Mass-Spring-Damper Model.

Authors:  Mallory R Selzo; Christopher J Moore; Md Murad Hossain; Mark L Palmeri; Caterina M Gallippi
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2016-03-08       Impact factor: 2.725

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.