Anna C Chlebowski1,2, Glen E Kisby1,3. 1. Western University of Health Sciences College of Osteopathic Medicine of the Pacific-Northwest, Lebanon, OR 97355, USA. 2. Technical Contact. 3. Lead Contact.
Abstract
Dot blotting allows for the rapid screening of a larger number of samples and/or targets than more traditional methods, such as a western blot or in-tissue-based methods. We have developed a dot-blot assay specifically for use with a LiCor Odyssey CLx imager, which allows for sensitive detection of proteins in the infrared range. Here, we provide a detailed protocol for the preparation of brain tissue and neural cell culture lysates for analysis of protein targets by dot blotting.
Dot blotting allows for the rapid screening of a larger number of samples and/or targets than more traditional methods, such as a western blot or in-tissue-based methods. We have developed a dot-blot assay specifically for use with a LiCor Odyssey CLx imager, which allows for sensitive detection of proteins in the infrared range. Here, we provide a detailed protocol for the preparation of brain tissue and neural cell culture lysates for analysis of protein targets by dot blotting.
Timing: 1+ h, depending on number of samplesCultivate cell type as appropriate. Further details on basic cell culture should be sought elsewhere.Once cell cultures are ready for collection (80%–90% confluency), media over the cells is aspirated, accutase (STEMCELL™ Technologies or Sigma) is then added and the cells treated for 5 min in a CO2 incubator at 37°C.Check for detachment by light microscopy after 5 min and incubate for an additional 5 min for adequate detachment (if necessary).1 mL of accutase is sufficient for one well of a 6-well plate or 500 μL/well of a 12-well plate.Other dissociation methods can be used as appropriate for cell type and preferred culture protocols.Using a pipette tip, transfer the detached cells into a microcentrifuge tube (1.7 mL or larger, depending on size of well plate).Using a fresh pipette tip, rinse each well with the same volume of cell culture media or DMEM (500 μL/12-well plate, 1 mL/6-well plate) and then add it to the tube containing the detached cells. Repeat for all sample wells.Centrifuge the tubes at 750 × g for 5 min at 20°C–22°C (room temperature).Carefully aspirate all media over the cell pellet and then transfer tubes to an ice bucket.Cell pellets can be washed with PBS prior to snap-freezing. Care should be taken to not agitate or aspirate the cell pellet. This step is routinely omitted when the cell pellet is small, to minimize loss of material.Pause Point: At this point, cell pellets can be snap-frozen in liquid nitrogen and stored at −80°C.Resuspend cell pellet in 25 μL of cOmplete™ Mini EDTA-Free Protease Inhibitor cocktail (Roche).To make the protease inhibitor cocktail, dissolve 1 cOmplete™ Mini EDTA-free tablet in 10 mL PBS (pH 7.4) and mix well.Homogenize the samples on ice by sonication (we use a Qsonica, Newton, CT), to obtain a homogeneous solution.Program used: Pulse 1 s on/5 s off at 20% amp.Typically 2–4 pulses is adequate to achieve homogenization. Monitor for a uniformly turbid solution and no visible cell clumps.Place the microcentrifuge tube in a 50 mL conical tube containing ice to prevent overheating of samples during sonication.Wash sonicator tip with ddH2O before starting, between samples, and after sonication.Store cell homogenates on ice.Wear hearing protection as appropriate for the sonication instrument used.Centrifuge cell homogenates for 1 min at 750 × g.Transfer the supernatant to a clean microcentrifuge tube and store the remaining insoluble material at −80°C.Bring final volume of the supernatant to 150 μL with PBS.Final volume of 150 μL was appropriate for the amount of protein collected from 200,000 or more neural stem cells (NPCs). Smaller numbers of cells may warrant a smaller final volume. Likewise, a larger volume or additional dilutions may be required for samples with a higher cell count and/or protein yield.Diluted cell supernatants are stored at −80°C until protein quantification.Pause Point: Samples can be stored at −80°C before and/or after protein quantification. However, we recommend analyzing supernatants after centrifugation to minimize freeze-thaw effects on samples.
Option 2: Brain Tissue Preparation
Timing: 3+ h, depending on number of samplesFreshly frozen brain tissue is placed in a microcentrifuge tube (previously weighed) and then immediately placed on dry ice. The tube containing the brain tissue is weighed and the wet weight determined. All tubes are kept on ice during this procedure.Murine hemi-brains were dissected on ice into cortex, hippocampus, and cerebellum and each region placed in separate tubes before processing the tissue samples.Add extraction buffer to each sample tube containing tissue (see recipe for extraction buffer in Table 1).
Table 1
Extraction Buffer Recipe for the Extraction of Proteins from Brain Tissue
Reagent
Final Concentration
Amount
Tris HCl (100 mM) pH 7.8
10 mM
1.0 mL
Dithiothreitol (DTT) (50 mM)
0.5 mM
0.1 mL
MgCl2 (50 mM)
5.0 mM
1.0 mL
Adenosine triphosphate (ATP)
3.08 mg/mL
30.8 mg
cOmplete™ Mini Protease inhibitor, EDTA-free
1×
1 tablet
ddH2O
n/a
7.8
Total
n/a
10.0
Extraction Buffer Recipe for the Extraction of Proteins from Brain TissueThe amount of extraction buffer depends on the brain region: 500 μL for cortex, 200 μL for hippocampus, and 50 μL for cerebellum from the hemi-brain of an adult mouse. Volumes may need to be adjusted for other regions.Homogenize the tissue on ice by sonication (Qsonica, Newton, CT), to obtain a uniform turbid solution.Program used: Pulse 1 s on/5 s off at 30% amp.2–10 pulses with the sonicator may be required, depending on the amount and type of tissue used. Monitor for a uniformly turbid solution and no visible tissue chunks.Place the microcentrifuge tube in a 50 mL conical tube containing ice to prevent overheating of samples during sonication.Wash sonicator tip with ddH2O before starting, between samples, and after sonication.Store cell homogenates on ice.After sonication, place the tissue homogenates on ice for 30 min to lyse the tissue.Transfer tissue lysates to a new 1.7 mL microcentrifuge tube (previously weighed). Weigh tube to determine the wet brain tissue weight.Centrifuge the lysates at 15,000 × g for 90 min at 4°C.Transfer the supernatant into a new tube (soluble proteins). The pellet (insoluble fraction) is stored at −80°C until use.Supernatants are stored at −80°C until protein quantification.Use excess extraction buffer as diluent for protein quantification technique of choice.Pause Point: Supernatants can be stored at −80°C before and/or after protein quantification. However, we recommend analyzing supernatants after centrifugation to minimize freeze-thaw effects on samples.
Key Resources Table
Resource Availability
Lead Contact
Further information and requests for resources and reagents should be directed to and will be fulfilled by the Lead Contact, Dr Glen Kisby (gkisby@westernu.edu).
Materials Availability
This study did not generate any unique materials or reagents.
Data and Code Availability
This study did not generate any unique datasets or code.
Materials and Equipment
Primary antibodies as appropriate for sample/targets of interest should be selected, and the appropriate IRDye® Secondary antibodies (LiCor) selected. IR-wavelength secondary antibodies are also available from other vendors and could potentially be utilized as well.For generation of neural stem cells and neurons, iPSCs (cell line ND50031 from RUCDR Infinite Biologics) were differentiated using protocols commercially available from STEMCELL™ Technologies. Similar cells can be purchased commercially or derived through other differentiation processes.Tris-buffered saline (TBS) and TBS + 0.1% Tween-20 (TBST) are required. These can be made in-house, or purchased commercially (as listed in the Key Resources Table). This buffer system is for the extraction of soluble native proteins from cells or tissues. If analysis of other protein fractions is desired, alternate buffer formulations can be used as appropriate. 1x TBS should contain 25 mM Tris and 0.15 mM NaCl, buffered to pH 7.6.The Bradford Assay (Bio-Rad, Cat# 5000201) can be used for protein quantification, but any other similar method could also be used.The Bio-Dot® Microfiltration Apparatus (Bio-Rad) is made of polysulfone with a silicone gasket, designed for the application of samples to a membrane using vacuum. Adapters are available for the application of either 96 “dots” in a 12x8 array, or as rectangular “slots” in a 6x8 array. The current protocol is designed for use with the 96 “dot” array, but could be modified for the slot array (Bronowicka-Szydełko et al., 2020). Details on setup and cleaning of the apparatus can be found in the product manual (https://www.bio-rad.com/webroot/web/pdf/lsr/literature/M1706545.pdf). Similar vacuum apparatuses are available through other manufacturers (Guillemin et al., 2009), the specific user manuals for these should be referenced for any necessary modifications.Alternatively, dots can be applied freehand, as has been published by other laboratories (Au - Chunhui et al., 2018, Tizon et al., 2010). In this case, the user should take care to keep the membrane as clean as possible (the infrared detection system is more sensitive to background contamination than other methods such as chemiluminescence (Schutz-Geschwender et al., 2004)). A template, such as a pipette tip rack, would be helpful for aligning the dots on the membrane in a uniform manner, as analysis in the ImageStudio™ software can be done using a 96-well template rather than individual selection of each dot (Wehr and Levine, 2012).PVDF membranes are not recommended for use with the Bio-Dot® Microfiltration Apparatus, as per the product manual. Specific inquiries regarding membrane types other than nitrocellulose should be investigated with the manufacturer.CRITICAL: Methanol (CAS# 67-56-1) is required for the Revert™ 700 Total Protein Stain Kit. Avoid direct skin contact and inhalation of vapors of methanol and Revert™ Kit reagents. We recommend keeping the kit reagents contained in a chemical safety hood or a closed container. Use appropriate personal protective equipment (gloves, eye protection) when handling these reagents.CRITICAL: The Revert™ 700 Total Protein Stain Kit components contain acetic acid (CAS# 64-19-7) and sodium hydroxide (CAS# 1310-73-2). Avoid contact with skin, eyes, and inhalation of vapors. We recommend keeping the kit reagents in a chemical safety hood or a closed container. Review the SDS document for these reagents before use. Use appropriate personal protective equipment (gloves, eye protection) when handling of these reagents.
Step-By-Step Method Details
Sample Loading onto Membranes
Timing: 2.5+ h/membrane, plus thawing samplesFollowing protein determination, samples are loaded on a nitrocellulose membrane by manual pipetting and allowed to dry thoroughly prior to staining. Careful dilution of samples and application to the membrane is important for uniform results. The use of a 96-well dot array allows for high-throughput determination of a larger number of samples (96 individual samples, or 48 samples analyzed in duplicate).Thaw protein samples on ice. Quantify using a Bradford Assay or similar technique for total protein quantification.Dilute the protein sample before applying to the membrane.Each sample must be diluted to a final concentration of 1 μg of protein in a total volume of 10 μL (see Troubleshooting section for method optimization and selection of protein loading amount).Store the diluted samples on ice before loading on the membrane.We recommend replicates of each sample be applied to the membrane for greater accuracy. We also suggest making 5–10 μL extra of each sample solution before application to ensure an adequate volume for all dots.Using scissors cleaned with alcohol, cut a nitrocellulose membrane (0.45 μm) to size for the required number of samples to be dotted.Place the cut membrane(s) in a container with a lid (e.g., clean pipette tip box) and cover them with TBS.Allow the membrane(s) to rehydrate for a minimum of 10 min before applying samples.Multiple membranes can be placed in the same box for rehydration with TBS.Place a pre-wetted membrane in the 96-well Bio-Dot® Microfiltration Apparatus (Bio-Rad) according to the manufacturer’s instructions. If less than 96 samples are applied to a membrane, cover any un-used wells with tape and seal well.Connect the vacuum line and leave it open to the atmosphere during sample application.When setting up the Bio-Dot® Apparatus, the membrane should be marked to keep track of the orientation of samples. A notch or small hole(s) made with a clean tweezer along the edge of the membrane is ideal.Add 100 μL TBS to each well of the membrane in the Bio-Dot® apparatus.Turn on the vacuum, and with a gloved finger, gently block the vacuum line (open to the atmosphere) to slowly increase the vacuum on the Bio-Dot® apparatus. Make sure that the TBS is vacuumed thoroughly before applying samples. Detailed information for the operation of the vacuum valve of the apparatus can be found in the product manual.Add 10 μL of sample to each well of the membrane. Take care not to puncture the membrane during application with a pipette tip. Replace the pipette tip between each sample.After all samples are loaded on the membrane, cover the Bio-Dot® apparatus to prevent dust or dirt contamination (foil or a pipette lid box works well). Allow the apparatus to sit for a minimum of 1 h without vacuum to allow protein samples to bind.After 1 h, turn on the vacuum and using a gloved finger to block the vacuum line (open to the atmosphere) to gently pull any remaining sample through the membrane.Use a pair of clean tweezers to remove the membrane from the apparatus. Place the clean membrane on an elevated surface to dry for a minimum of 1 h.We suggest placing the membrane on the hinge of a clean pipette tip box (with lid) to allow optimal air flow and then the membrane placed in a laminar flow hood to prevent contamination from dust or dirt.After the membrane is dried, it can be stained with Revert® 700 Total Protein Stain or it can be stored for up to one week at 4°C.Pause Point: For long term storage, dotted membranes can be stored at 4°C for up to one week prior to staining and immunoprobing.
Revert® 700 Total Protein Stain
Timing: 1 h per membraneThe Revert® 700 Total Protein Stain (LiCor) allows for the normalization of the protein dotted on each well of the membrane. The final antibody signal is normalized against the Revert® signal to allow for more accurate quantification of total amount of applied protein and allows for the simultaneous detection of two protein targets. Re-imaging of the membrane after removing the Revert® stain accounts for residual background, especially when using two antibodies for a sample, or for detection of low-abundance proteins.Once the Revert® staining of the membrane is started, you must continue through each of the steps to completion (i.e., incubation with antibodies and imaging). Repeated drying of the membrane or storage in buffer for an extended period of time is not recommended.Place each membrane in a clean, small container with a lid to allow the membrane/solution to gently move on a rocker. Large containers will require excess reagents (e.g., Revert®, antibodies).For large membranes (e.g., >48 dots), a clean pipette tip box with lid works well. Small membranes (<16 dots) can be stained or immunoprobed in sterile petri dishes (>35 mm). Boxes designed for incubating western blots are suitable for medium-size membranes (16–48 dots). Volumes need to be adjusted for the type of container (e.g., a pipette tip box or container of similar size) used for the membrane.LiCor does not recommend placing multiple membranes in a container for staining with Revert® or antibodies.Add methanol to Revert® kit reagents as described by the manufacturer. Appropriate volumes to add are indicated for each reagent.Revert® 700 Total Protein Stain kit reagents need to be kept in a chemical safety hood when open and the waste disposed to minimize exposure to fumes.Rinse membrane briefly with 15 mL ultrapure water before staining.Add 5–10 mL (enough volume to just cover the membrane) Revert® 700 Total Protein Stain to the membrane and incubate for 5 min with gentle rocking. Keep container lid closed or if possible, place the rocker in a chemical safety hood to reduce exposure to fumes. Blue dots should become visible for protein samples that were applied to the membrane.Lack of visible blue dots could indicate issues with protein application to the membrane.Decant the Revert® stain into a waste container.Add 10–15 mL of Revert® Wash solution to the membrane and incubate for 30–60 s with gentle shaking/agitation.Decant the wash solution into a waste container and wash the membrane for 30–60 s in 15 mL ultrapure water. Decant wash solution into a waste container.Immediately after staining, the membrane(s) are imaged on a LiCor Odyssey® CLx near IR imager. Recommended scan settings are: medium scan quality, 84 μm resolution, and a 0.0 mm offset.For a faster run time, a resolution of 169 μm can be used instead of 84 μm. We prefer the higher resolution and slightly longer run time for better dot visualization.Use a silicone mat and roller (LiCor) to remove any bubbles or liquid on the membrane.Intensities and contrast values for both channels can be optimized for each image/experiment without impacting quantification.See Troubleshooting section for more information on problems with the Revert® stain.Clean imaging surface with alcohol prior to and after imaging the membrane.After imaging the stained membrane, add 5–10 mL of the Revert® 700 Destain solution and incubate for 5–7 min at 20°C–22°C with gentle rocking. The visible blue dots from the Revert® stain should rapidly diminish and disappear during destaining.Decant the Revert® Destain solution and briefly rinse the membrane with 15 mL ultrapure water.Re-image the destained membrane using the same imaging parameters (see step 22).After imaging the membrane, immediately block the membrane and incubate with primary antibody(s).
Incubation with Antibodies
Timing: 1 h, ∼16–24 h overnight incubation, 2+ h next day (+30 min/membrane)Following incubation with blocking buffer, the membrane(s) is incubated sequentially with a primary antibody and then an IRDye® conjugated secondary antibody for detection at 800 nm. If two targets are desired, the secondary antibody to the more-abundant protein is detected at 700 nm to minimize interference from residual Revert® stain (see Troubleshooting section).Incubate membrane(s) with 10 mL Intercept® (TBS) Blocking Buffer (LiCOr) for 45–60 min at 20°C–22°C with gentle rocking.Decant the blocking buffer and add the primary antibody solution to the membrane.Incubate the membrane 16–24 h at 4°C with gentle rocking.Primary antibodies should be made in Intercept® blocking buffer:TBS with 0.01% Tween-20 (50:50).If detection of two protein targets is desired, both primary antibodies can be applied and incubated simultaneously.Dilutions of each antibody should be determined by user. Suggested dilutions for western blotting are often appropriate, but the user should consult the data sheets for each antibody.Primary and secondary antibodies can also be diluted in 100% Intercept® (TBS) Blocking Buffer, but we recommend diluting the Intercept® buffer to minimize reagent use.Decant the primary antibody solution and wash the membrane successively 3 × 10 min with 15 mL of TBS +0.1% Tween-20 (TBST) at 20°C–22°C with gentle rocking.Primary antibody can be saved and re-used depending on dilution concentration and protein abundance, ideally within a few days to minimize antibody degradation or bacterial contamination of the buffer.Add the secondary antibody to the immunoprobed membrane and incubate for 1 h at 20°C–22°C.Dilute the IRDye® secondary antibody(s) in Intercept blocking buffer: TBS with 0.01% Tween-20 (50:50) to a final concentration of 1:10,000.If doing two protein targets, incubation with both secondary antibodies can be done simultaneously.Membranes must be kept in the dark during steps 31 through the imaging step. We recommend covering the container with foil if it is not opaque.For the detection of one protein, we recommend using the 800 nm channel for best sensitivity. For the detection of two proteins, we recommend using the more-abundant protein for the 700 nm channel. Additional support can be found for different applications can be found on the LiCor website.Decant the secondary antibody solution and wash the membrane 2 × 10 min each with 15 mL TBST, and 1 × 15 min with 15 mL TBS.Re-image membrane using the same imaging parameters (see step 22).
Expected Outcomes
Using either protein extraction method should yield solutions with quantifiable amounts of protein. The user may need to adjust the volumes if substantially larger or smaller amounts of cells/tissues need to be used. Examples of protein amounts from cell cultures are 50–100 μg from 2 × 105 hNPCs, 150–300 μg from 2 × 105 hNPCs differentiated into neurons for 28 days, and 15–25 μg protein from 1 × 105 mature astrocytes. From dissected hemi-brains of adult mice, protein amounts averaged 2–3 mg from the cortex, 400–600 μg from the hippocampus, and 1–2 mg from the cerebellum. Optimization of tissue/cell preparation protocols may be necessary for other cell/tissue types and should be optimized by the user.Following completion of this protocol for determination of proteins, the membrane will have been quantified three times: first with Revert® Total Protein Stain for normalization, second to account for residual Revert® background, and third for quantification of protein target(s). Representative images of these three critical steps are shown in Figure 1. Analysis of these three images is described in the Quantification and Statistical Analysis section, and is shown in Table 2.
Figure 1
Representative Images at Each Critical Step of the Dot Blot Assay
A membrane was dotted with neuronal cell lysates (1 μg protein/dot) and imaged following staining with Revert® 700 Total protein stain (shown red) (A), after removal of the Revert® stain (B), and after incubation with primary and secondary antibodies (shown green) (C). Note the intensity of the dots following staining with Revert® are relatively uniform indicating that samples were accurately applied to the membrane. Primary antibody used was mouse-anti Map2 (Invitrogen, MA5-12826) at 1:1,000 dilution. Secondary antibody was goat-anti-mouse IRDye® 800 at a 1:10,000 dilution (LiCor). The quantification of these membranes and an example workflow is demonstrated in Table 2.
Table 2
Example of Data Quantification and Analysis
700 nm Revert® signal
39325.63
47854.84
43274.98
43533.29
41409.51
38262.79
39320.73
40580.71
48671.24
48028.9
40101.15
40803.05
43633.96
46954.63
41187.27
39185.38
41820.01
44945.08
51517.1
49714.33
39130.48
39431.51
41083.66
41974.86
36102.52
35537.79
41391.7
40769.94
44075.72
45503.02
42817.7
36250.27
36829.37
38899.21
33379.94
34572.8
44005.65
40887.78
44576.13
46629.25
46787.11
40068.7
32025.12
32182.57
32564.8
35213.75
34041.37
35230.14
43513.12
38902.95
34201.8
33869.79
38262.18
35114.08
35140.95
35699.27
79585.73
67318.86
36241.09
37316.02
45459.24
41305.18
33846.3
33753.52
45646.42
43645.91
41595.6
43134.48
43042.87
41227.75
40312.23
39034.98
800 nm Revert® signal after destain
542.3281
546.135
541.2319
537.6279
534.8091
533.4812
528.686
530.0818
530.0691
530.1543
537.7957
537.7644
536.2605
534.9106
532.8384
530.3384
527.7983
528.7622
530.957
527.2891
539.6426
540.4126
537.0103
534.4797
534.1409
533.1497
527.7388
527.1843
525.5657
524.0273
536.2266
530.8677
530.1199
528.6519
525.6045
525.4082
526.0254
526.3159
525.0593
525.4895
538.4771
534.8564
534.2458
530.8733
530.2805
528.0095
524.8201
537.4849
540.7908
540.9287
538.4631
534.0232
531.7529
529.0176
528.0061
526.1147
549.5657
550.1379
537.6619
538.8743
535.8604
532.7209
531.6387
533.2056
540.8818
541.7168
542.0713
539.4387
533.8689
532.4343
531.7932
530.7268
800 nm antibody signal
1974.495
2255.098
2605.119
2595.216
2381.56
2364.609
2441.422
2555.386
2762.412
2753.495
1860.999
1983.541
1901.76
1943.954
2120.027
2118.661
2808.831
2844.216
2647.968
2515.487
1698.695
1923.18
2825.785
2832.818
2055.783
2037.267
2011.492
2147.915
2644.119
2452.068
1832.993
1878.122
2455.607
2695.125
1832.613
1865.734
2672.099
2579.503
3100.659
3056.216
1939.647
2147.758
1865.78
1928.732
2486.749
2602.941
1776.811
1739.599
2315.652
2437.313
2128.267
2018.61
2123.796
2142.003
2748.91
2618.123
3267.77
2896.804
1858.057
1969.232
2846.689
2698.491
2026.776
1921.47
2065.46
1988.486
2292.146
2388.771
2870.855
2895.53
2051.392
2357.373
(800 nm antibody signal) minus (700 nm Revert®destain signal)
Quantification of the membranes shown in Figure 2, with the suggested flow for analysis. Final consolidation of replicates and statistical analysis will depend on the particular experimental design. Values in the table are oriented the same as the images of the membranes. Quantification of the images is copied from the exported Excel files following image acquisition, and consolidated to a single spreadsheet for analysis.
Representative Images at Each Critical Step of the Dot Blot AssayA membrane was dotted with neuronal cell lysates (1 μg protein/dot) and imaged following staining with Revert® 700 Total protein stain (shown red) (A), after removal of the Revert® stain (B), and after incubation with primary and secondary antibodies (shown green) (C). Note the intensity of the dots following staining with Revert® are relatively uniform indicating that samples were accurately applied to the membrane. Primary antibody used was mouse-anti Map2 (Invitrogen, MA5-12826) at 1:1,000 dilution. Secondary antibody was goat-anti-mouse IRDye® 800 at a 1:10,000 dilution (LiCor). The quantification of these membranes and an example workflow is demonstrated in Table 2.Example of Data Quantification and AnalysisQuantification of the membranes shown in Figure 2, with the suggested flow for analysis. Final consolidation of replicates and statistical analysis will depend on the particular experimental design. Values in the table are oriented the same as the images of the membranes. Quantification of the images is copied from the exported Excel files following image acquisition, and consolidated to a single spreadsheet for analysis.
Figure 2
Optimal Application of Samples for Accurate Quantification
Following incubation with the Revert® 700 Total Protein Stain (red) and quantification using Image Studio™, dots are selected for quantification (yellow circles) to designate protein signal from background. The membrane was dotted with a dilution series of murine cerebellar extracts (note that protein concentrations for sample 3 are slightly higher). Note the uniform intensities of dots in each column (same amount of protein/dot).
Expected signal intensities following Revert® staining are in the 40–100K range, for 1 μg protein dotted. Following Revert® destain, values should drop to below 5,000. Signals following antibody staining will be dependent on the individual target of interest, but signals in the 1,000–6,000 range were common.
Quantification and Statistical Analysis
Analysis of the scanned images was done using Image Studio™ software from LiCor. Each sample dot is encircled (this can be done using a multiwell plate template provided in the ImageStudio™ software) to designate sample area from background (Figure 2), and the data exported as an Excel spreadsheet. The LiCor website (https://www.licor.com/bio/support/) provides tutorials on the use of this software for image analysis and quantification of stained or immunoprobed membranes.Optimal Application of Samples for Accurate QuantificationFollowing incubation with the Revert® 700 Total Protein Stain (red) and quantification using Image Studio™, dots are selected for quantification (yellow circles) to designate protein signal from background. The membrane was dotted with a dilution series of murine cerebellar extracts (note that protein concentrations for sample 3 are slightly higher). Note the uniform intensities of dots in each column (same amount of protein/dot).Data is presented as (antibody signal - residual Revert® signal)/(Revert® signal) for each dot. An example of the workflow for analysis is shown in Table 2. Values from analysis of membranes will be assay-dependent, examples of expected quantification values are given in the Expected Outcomes section. While the analysis can be completed without the additional imaging step following Revert® destain, this additional quantification allows for more accurate background subtraction, especially if quantification of a protein target in the 700 nm channel is desired.Multiple technical replicates (as reasonably possible) should be used, based on the total number of samples to be analyzed and the volume of each sample available. If the protein dilutions are made and applied to the membrane properly, variability between technical replicates should be low. If high variability is observed, refer to the Troubleshooting section. Variability between technical replicates for properly applied samples should be within + 5% of the average value. Between separately prepared and diluted protein samples applied to different membranes, variability should be below 15%.
Limitations
Quantification of low abundant proteins will be challenging using this method, because the cell or tissue lysate is not separated prior to staining/immunoprobing and the mixture of proteins in each dot is complex. Optimization of the amount of protein applied to the membrane may increase sensitivity. Given the rapidity of this assay and the ability to quickly troubleshoot issues, we suggest trying a dilution series with higher protein loading per dot to optimize the experimental conditions for each protein target.
Troubleshooting
Problem
Inconsistent staining with the Revert® Total Protein Stain.
Potential Solution
If there is inconsistency between technical replicates, this suggests that the problem is with either the dilution of the samples or the pipetting of the samples on the membrane. To minimize inconsistency among technical replicates, ensure samples are well-mixed prior to application to the membrane, and use proper pipetting technique to apply samples to the membrane. Due to the nature of manual pipetting, small variations between technical replicates are expected, and will be accounted for during the Revert® normalization analysis. Deviation between technical replicates should be below 5%. If the technical replicates are consistent, but there is variation among the different samples, the protein quantification or dilution of the samples was done incorrectly. For an example of an incorrectly quantified and loaded membrane, see Figure 3. Consequently, the samples should be re-quantified and applied to another membrane.
Figure 3
Poor Sample Loading Leads to Inconsistent Revert® Staining
Each dot applied to the membrane was thought to contain 1 μg protein, based on previous quantification of murine brain region tissue extracts. Imaging of the Revert® stained membrane revealed variable intensities among samples. Note that adjacent technical replicates (white box) of the samples was relatively uniform indicating that the samples were incorrectly quantified. Re-quantification of the samples is required prior to their application to a new membrane.
Poor Sample Loading Leads to Inconsistent Revert® StainingEach dot applied to the membrane was thought to contain 1 μg protein, based on previous quantification of murine brain region tissue extracts. Imaging of the Revert® stained membrane revealed variable intensities among samples. Note that adjacent technical replicates (white box) of the samples was relatively uniform indicating that the samples were incorrectly quantified. Re-quantification of the samples is required prior to their application to a new membrane.Quantification of the antibody is low or too high.Verify that the primary antibody has been previously used for blotting-applications with the reported concentration of the diluted antibody. Dilution of the protein sample is also an alternative approach. Quantification of the Revert® signal is robust and linear at 500 ng per dot and above (Figure 4). We suggest as a starting amount that 1 μg protein be loaded per dot because higher amounts caused the protein samples to smear between samples for highly-abundant targets (Figure 5). For detection of less-abundant targets, application of higher amounts (> 1 μg) may be required to get good signal and can be tried as needed. The binding capacity of a nitrocellulose membrane is 80–100 μg/cm2. The protein loaded per dot should not exceed this amount (∼10 μg/dot).
Figure 4
Determination of Revert® 700 Total Protein Stain Signal Linearity
Protein homogenates from the cerebral cortex of adult mice were diluted and then applied (in duplicate) to a membrane before analysis using the Revert® Total Protein Stain at a high (A and B) or a low dilution range (C and D) (Note: concentrations in C are approximate whereas concentrations in (D) are exact). Linear fit of the Revert® quantification of the membrane from both dilution ranges are shown (B, D). Linear fits for the higher dilution range were R2=0.96 and higher, whereas R2 values for the lower dilution range were between 0.91 and 0.95. Images (A) and (C) were captured separately so visual comparison should be avoided.
Figure 5
Representative Image of a Protein Overloaded Membrane
Protein homogenates from the cerebral cortex of adult mice were diluted, dotted on a membrane and imaged following detection of tau5 (Invitrogen, MA5-12808) at a 1:1,000 dilution (imaging of the Revert stained membrane in Figure 4). Secondary antibody was goat-anti-mouse IRDye® 800 at a 1:10,000 dilution (LiCor). Note the smearing of the dots in samples 1 and 4 (mice that over-expresses human tau, htau) vs. samples 2 and 3 (wild type mice) indicate a high level of tau protein and thus, quantification will be less accurate. To avoid smearing of dots for samples with high-abundance proteins, we recommend applying 1 μg protein/dot.
Determination of Revert® 700 Total Protein Stain Signal LinearityProtein homogenates from the cerebral cortex of adult mice were diluted and then applied (in duplicate) to a membrane before analysis using the Revert® Total Protein Stain at a high (A and B) or a low dilution range (C and D) (Note: concentrations in C are approximate whereas concentrations in (D) are exact). Linear fit of the Revert® quantification of the membrane from both dilution ranges are shown (B, D). Linear fits for the higher dilution range were R2=0.96 and higher, whereas R2 values for the lower dilution range were between 0.91 and 0.95. Images (A) and (C) were captured separately so visual comparison should be avoided.Representative Image of a Protein Overloaded MembraneProtein homogenates from the cerebral cortex of adult mice were diluted, dotted on a membrane and imaged following detection of tau5 (Invitrogen, MA5-12808) at a 1:1,000 dilution (imaging of the Revert stained membrane in Figure 4). Secondary antibody was goat-anti-mouse IRDye® 800 at a 1:10,000 dilution (LiCor). Note the smearing of the dots in samples 1 and 4 (mice that over-expresses human tau, htau) vs. samples 2 and 3 (wild type mice) indicate a high level of tau protein and thus, quantification will be less accurate. To avoid smearing of dots for samples with high-abundance proteins, we recommend applying 1 μg protein/dot.Residual signal from the Revert® stain was detected after destaining the membrane.We consistently saw 95%–97% removal of the Revert® stain when following the protocol as described (Figure 1). Quantification of the Revert® stain in the 700 nm channel should yield values in the 40–100 K range. Following destain, values in the 700 nm channel should be below 5,000. If the Revert® stain is not efficiently removed (step 23), the incubation time of the reversal solution should be increased to 10 min. More residual signal is routinely detected in the 700 nm than the 800 nm channel, although a small amount will be detected in the 800 nm channel (quantification in the 800 nm channel following destain should be <600). For this reason, antibody detection should be done in the 800 nm channel, as well as the extra imaging step following Revert® destain, so that residual Revert® stain can be accurately determined. Accurate levels of the residual Revert® increases the accuracy of the signal intensities from the primary antibodies in the 700 nm or 800 nm channels.
REAGENT or RESOURCE
SOURCE
IDENTIFIER
Antibodies
IRDye® 700 and 800 Secondary antibodies
LiCor
Species specific, see https://www.licor.com/bio/reagents/irdye-secondary-antibodies for options
Authors: N Guillemin; B Meunier; C Jurie; I Cassar-Malek; J-F Hocquette; H Leveziel; B Picard Journal: J Physiol Pharmacol Date: 2009-10 Impact factor: 3.011