| Literature DB >> 32995629 |
Mahendra T A Sampurna1, Risa Etika1, Martono T Utomo1, Siti A D Rani1, Abyan Irzaldy1, Zahra S Irawan1, Kinanti Ayu Ratnasari1, Arend F Bos2.
Abstract
INTRODUCTION: A range of phototherapy devices are commercially available. The American Academy of Pediatrics (2004) recommends routine intensity measurement of phototherapy devices to ensure that babies affected by hyperbilirubinemia receive effective phototherapy.Entities:
Keywords: Biomedical engineering; Hyperbilirubinemia; Intensity; Irradiance; Pediatrics; Phototherapy
Year: 2020 PMID: 32995629 PMCID: PMC7502345 DOI: 10.1016/j.heliyon.2020.e04950
Source DB: PubMed Journal: Heliyon ISSN: 2405-8440
Figure 1Silhouette model illustration. Used with the permission of Van Imhoff et al. (2013) [5].
The characteristics of PT devices: duration of use and irradiation levels (in μW/cm2/nm) and velocity of irradiation decay per hour of use (μW/cm2/nm/hour).
| PT Devices | Brand | Type of PT Lamp | Hours of use | Irradiance measurement | Delta Hours of use | Delta Irradiance | Irradiance Velocity Decay per hour of use | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Initial hour meter | 2nd-hour meter | 3rd-hour meter | 1st Measurement | 2nd Measurement | 3rd Measurement | ||||||
| 1 | XHZ-90 | Fluorescent | 5114.14 | 5114.22 | 5144.23 | 16.75 ± 1.6 | 15.76 ± 1.2 | 14.22 ± 5.3 | 30.09 | 2.53 | 0.08 |
| 2 | XHZ-90 | Fluorescent | 4611.25 | 4671.32 | 4693.26 | 16.83 ± 1.4 | 16.66 ± 1.7 | 15.88 ± 2.0 | 82.01 | 0.94 | 0.01 |
| 3 | My Life/MP71 | Fluorescent | 11643.98 | 11681.79 | 11729.82 | 9.072 ± 1.3 | 6.59 ± 0.7 | 7.14 ± 0.9 | 85.84 | 1.93 | 0.02 |
| 4 | SEEFAR NICE 4000 | LED | 2688.00 | 2763.00 | 2906.00 | 51.63 ± 9.5 | 54.22 ± 8.2 | 50.39 ± 9.3 | 218.00 | 1.24 | 0.01 |
| 5 | SEEFAR NICE 4000 | LED | 2223.00 | 2223.00 | 2446.00 | 59.20 ± 9.8 | 55.64 ± 11.2 | 53.77 ± 12.7 | 223.00 | 5.43 | 0.02 |
| 6 | GEA | Fluorescent | 7.53 | 48.27 | 242.31 | 27.55 ± 2.3 | 21.24 ± 2.6 | 16.61 ± 2.5 | 234.78 | 10.94 | 0.05 |
| 7 | GEA | Fluorescent | 46.46 | 46.50 | 252.56 | 25.69 ± 2.2 | 21.70 ± 2.7 | 22.50 ± 1.4 | 206.10 | 3.19 | 0.02 |
| 8 | Nideal PT Unit 2000 | Fluorescent | 44 | 280.30 | 666.90 | 32.97 ± 1.5 | 29.23 ± 2.3 | 27.76 ± 2.8 | 622.90 | 5.21 | 0.01 |
| 9 | Nideal PT Unit 2000 | Fluorescent | 0 | 155.20 | 534.80 | 30.62 ± 2.3 | 31.73 ± 2.7 | 30.48 ± 2.4 | 534.80 | 0.14 | 0.00 |
| 10 | Nideal PT Unit 2000 | Fluorescent | 2.40 | 223.60 | 518.10 | 34.07 ± 2.1 | 30.82 ± 1.8 | 31.62 ± 2.0 | 515.70 | 2.45 | 0.00 |
| 11 | Nideal PT Unit 2000 | Fluorescent | 71.80 | 321.10 | 627.50 | 32.35 ± 2.3 | 29.16 ± 1.6 | 25.98 ± 2.8 | 555.70 | 6.37 | 0.01 |