Achievable dose reductions with gonadal shielding for children and adults during abdominal/pelvic radiographic examinations: A Monte Carlo simulation.

Elanchezhian Somasundaram1,2, Samuel L Brady1,2, Keith J Strauss1,2.   

Abstract

PURPOSE: Recently, medical professionals have reconsidered the practice of routine gonadal shielding for radiographic examinations. The objective of this study was to evaluate the gonadal dose reduction achievable with gonadal shields in the primary beam during abdominal/pelvic radiographic examinations under ideal and non-ideal shielding placement.
METHODS: CT scans of CIRS anthropomorphic phantoms were used to perform voxelized Monte Carlo simulations of the photon transport during abdominal/pelvic radiographic examinations with standard filtration and 0.1 mm Cu + 1 mm Al added filtration to estimate gonadal doses for an adult, 5 yr old, and newborn phantom with and without gonadal shields. The reduction in dose when the shields were not placed at the ideal locations was also evaluated. The ratio of the number of scattered-to-primary photons (SPR) across the anteroposterior (AP) dimension of the phantoms was also reported.
RESULTS: The simulated dose reduction with ideal shielding placement for the testes and ovaries ranged from 80% to 90% and 55% to 70% respectively. For children, a misalignment of the shield to the gonad of 4 cm reduced the measured dose reduction to the gonads to <10%. For adults, this effect did not occur until the misalignment increased to ~6 cm. Effects of dose reduction with and without the gonadal shields properly placed were similar for standard filtration and added filtration. SPR at the level of the testes was consistently <1 for all phantoms. SPR for ovaries was ~1.5 for the adult and 5-yr old, and ~1 for the newborn phantom.
CONCLUSION: Dose reduction with ideal alignment of the simulated gonadal shield to the gonads in this study was greater for the testes than the ovaries; both reductions were substantial. However, the dose reductions were greatly reduced (to <10%) for both sexes with misalignment of the gonads to the shields by 4 cm for children and 6 cm for adults.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  Monte Carlo; gonadal shielding; pediatric; radiation dose reduction; radiography

Mesh:

Year:  2020        PMID: 32978986     DOI: 10.1002/mp.14495

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Bismuth Pelvic X-Ray Shielding Reduces Radiation Dose Exposure in Pediatric Radiography.

Authors:  Bow Wang; Chien-Yi Ting; Cheng-Shih Lai; Yi-Shan Tsai
Journal:  Biomed Res Int       Date:  2021-10-11       Impact factor: 3.411

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.