| Literature DB >> 32974463 |
Jonathan M Behar1, Malcolm C Finlay1, Edward Rowland1, Vivienne Ezzat1, Simon Sporton1, Mehul Dhinoja1.
Abstract
BACKGROUND: Conventional cardiac resynchronization therapy (CRT) involves the placement of an epicardial left ventricular (LV) lead through the coronary venous tree. However, alternative approaches of delivering CRT have been sought for patients who fail to respond to conventional methods or for those where coronary venous anatomy is unfavourable. Biventricular pacing through an endocardial LV lead has potential advantages; however, the long-term clinical and safety data are not known. CASEEntities:
Keywords: Case series; LV endocardial; Pacing lead; Transvenous extraction
Year: 2020 PMID: 32974463 PMCID: PMC7501919 DOI: 10.1093/ehjcr/ytaa130
Source DB: PubMed Journal: Eur Heart J Case Rep ISSN: 2514-2119
| Case number | 1 | 2 | 3 | 4 |
|---|---|---|---|---|
| Age (years) | 65 | 53 | 77 | 23 |
| Sex | Male | Male | Male | Male |
| Aetiology | Refractory atrial fibrillation and impaired LV function (LVEF 32%) | Non-ischaemic cardiomyopathy (LVEF 25 %) | Ischaemic cardiomyopathy (LVEF 28%) | Hypertrophic cardiomyopathy (LVEF 30%) |
| ECG prior to original implant |
Atrial fibrillation QRSd 110 ms |
Sinus rhythm LBBB, QRSd 160 ms |
Sinus rhythm LBBB, QRSd 156 ms |
Sinus rhythm QRSd 100 ms |
| Heart failure medications |
Ramipril 10 mg Bisoprolol 7.5 mg Spironolactone 25 mg |
Ramipril 10 mg Bisoprolol 10 mg Spironolactone 25 mg |
Ramipril 7.5 mg Bisoprolol 5 mg |
Ramipril 10 mg Bisoprolol 10 mg |
| Device indication | Pace and ablate CRTP | Primary prevention CRTD | Primary prevention CRTD | Primary prevention ICD |
| LV endocardial lead |
Medtronic 5076 CapsureFix, 85 cm |
Medtronic 5076 CapsureFix, 85 cm |
Medtronic 3830 SelectSecure IS-1 Fixed helix |
Medtronic 6935M Quattro Secure DF4 Active fixation |
| Time from implant to extraction (months) | 10 | 30 | 36 | 20 |
| Extraction procedure time (min) | 40 | 60 | 120 | 30 |
| Extraction tools |
Firm stylet LV length Medtronic |
LLD stylet Spectranetics |
LLD stylet Spectranetics |
Firm stylet RV length Medtronic |
| Successful extraction | Yes | Yes | Yes | Yes |
| Total number of device procedures prior to extraction | 3 | 5 | 8 | 2 |
| Follow-up post-extraction (months) | 20 | 18 | 14 | 10 |
| Clinical status | Residual mitral regurgitation, surgical valve replacement | Worsening congestive cardiac failure. Died 6 months later | WiSE CRT implant. Improved NYHA status | No further thromboembolic events |
CRT, cardiac resynchronization therapy; CRTP, cardiac resynchronization therapy pacemaker; ECG, electrocardiogram; ICD, internal cardioverterdefibrillator; LBBB, left bundle branch abnormality; LLD, lead locking device; LV, left ventricular; LVEF, left ventricular ejection fraction; NYHA, New York Heart Association; RV, right ventricular.