Predictors of increase in pacing threshold after transcatheter pacing system implantation due to micro-dislodgement.

Motoaki Higuchi1, Yasutoshi Shinoda1, Tomoaki Hasegawa1, Mayu Ishibashi1, Norihiro Yamada1, Yoshiro Chiba1, Koji Ohira1, Minoru Murata1, Kazutaka Aonuma1.   

Abstract

BACKGROUND: Achieving a favorable pacing threshold with a Micra transcatheter pacing system (Micra-TPS) is needed to reduce battery depletion. In some cases, the threshold increases shortly after the device is implanted, and a higher pacing threshold may be required. This study aims to identify the causes and predictors of the increase in pacing threshold observed shortly after Micra-TPS implantation.
METHODS: The study included 64 consecutive patients who underwent Micra-TPS implantation between 2017 and 2020. The patients were divided into two groups depending on their pacing threshold: the increased pacing threshold (IPT) group (threshold increased by ≥0.5 V/0.24 ms within 1 month of implantation) and the stable pacing threshold (SPT) group.
RESULTS: Excluding four patients who could not be followed up, of the 60 remaining patients, nine (15%) were in the IPT group and 51 (85%) were in the SPT group. The IPT group had significantly lower implant impedance values and higher implant thresholds than the SPT group (582 ± 59 vs 755 ± 167 Ω [P < .001] and 1.29 ± 0.87 vs 0.71 ± 0.40 V/0.24 ms [P = .014]). Implant impedance and threshold may serve as predictors of a threshold increase after implantation (area under the curve: 0.737-0.943 and 0.586-0.926, respectively).
CONCLUSIONS: An IPT was noted shortly after Micra-TPS implantation owing to micro-dislodgement because of insufficient anchoring of the device to the myocardium. Impedance >660 Ω and threshold <1.0 V/0.24 ms may predict an increase in pacing threshold.
© 2020 Wiley Periodicals LLC.

Entities:  

Keywords:  artificial pacemaker; high pacing threshold; implant; implant impedance; micro-dislodgement; transcatheter pacing system

Year:  2020        PMID: 32969504     DOI: 10.1111/pace.14080

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976