Literature DB >> 32967932

Factors Determining Continuous Infusion Aerosol Delivery During Mechanical Ventilation.

Michael McPeck1, Sunya Ashraf2, Ann D Cuccia3, Gerald C Smaldone2.   

Abstract

BACKGROUND: Continuous nebulization of prostacyclins and albuterol by infusion pump during mechanical ventilation evolved as a popular off-label treatment for severe hypoxemic respiratory failure and asthma. Most institutions use a vibrating mesh nebulizer. A new breath-enhanced jet nebulizer is a potential alternative. This study was designed to compare these devices to better define factors influencing continuous infusion aerosol delivery. Device function, ventilator settings, and infusion pump flow were studied in vitro.
METHODS: Using a bench model of adult mechanical ventilation, radiolabeled saline was infused at 6 flows (1.5-12 mL/h) into test nebulizers; 4 examples of each were used in rotation to test device reproducibility. Four breathing patterns with duty cycles (percentage of inspiratory time) ranging from 0.13 to 0.34 were tested. The vibrating mesh nebulizer was installed on the "dry" side of the heated humidifier (37°C). The breath-enhanced jet nebulizer, installed on the "wet" side, was powered by air at 3.5 L/min and 50 psi. Infusion time was 1 h. Inhaled mass of aerosol was collected on a filter at the airway opening. Inhaled mass was expressed as the percentage of the initial syringe radioactivity delivered per hour. Radioactivity deposited in the circuit was measured with a gamma camera. Data were analyzed with multiple linear regression.
RESULTS: Variation in inhaled mass was significantly explained by pump flow and duty cycle (R2 0.92) and not by nebulizer technology. Duty cycle effects were more apparent at higher pump flow. Vibrating mesh nebulizers failed to nebulize completely in 20% of the test runs. Mass balance indicated that vibrating mesh nebulizers deposited 15.3% in the humidifier versus 0.2% for breath-enhanced jet nebulizer.
CONCLUSIONS: Aerosol delivery was determined by infusion pump flow and ventilator settings with comparable aerosol delivery between devices. The breath-enhanced jet nebulizer was more reliable than the vibrating mesh nebulizer; 10-12 mL/h was the maximum infusion flow for both nebulizer technologies.
Copyright © 2021 by Daedalus Enterprises.

Entities:  

Keywords:  aerosols; bias flow; drug delivery; epoprostenol; humidifiers; inhalation administration; mechanical ventilators; nebulizers and vaporizers

Year:  2020        PMID: 32967932     DOI: 10.4187/respcare.07715

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  3 in total

1.  The impact of nebulized epoprostenol and iloprost on hemoglobin oxygen affinity: an ex vivo experiment.

Authors:  Simon Woyke; Norbert Mair; Thomas Haller; Marco Ronzani; David Plunser; Herbert Oberacher; Hannes Gatterer; Christopher Rugg; Mathias Ströhle
Journal:  Am J Physiol Lung Cell Mol Physiol       Date:  2022-05-03       Impact factor: 6.011

Review 2.  The Unfulfilled Promise of Inhaled Therapy in Ventilator-Associated Infections: Where Do We Go from Here?

Authors:  Lucy B Palmer; Gerald C Smaldone
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2022-01-28       Impact factor: 3.440

3.  Real-Time In Vitro Assessment of Aerosol Delivery During Mechanical Ventilation.

Authors:  Michael McPeck; Janice A Lee; Ann D Cuccia; Gerald C Smaldone
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2021-07-06       Impact factor: 2.849

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.