Literature DB >> 32967784

Vie scope® laryngoscope versus Macintosh laryngoscope with personal protective equipment during intubation of COVID-19 resuscitation patient.

Maciej Maslanka1, Jacek Smereka2, Lukasz Czyzewski3, Jerzy Robert Ladny4, Marek Dabrowski5, Lukasz Szarpak6.   

Abstract

Entities:  

Year:  2020        PMID: 32967784      PMCID: PMC7471714          DOI: 10.1016/j.ajem.2020.08.085

Source DB:  PubMed          Journal:  Am J Emerg Med        ISSN: 0735-6757            Impact factor:   2.469


× No keyword cloud information.
Dear Editor, Airway management is a key element of resuscitation procedures. However, due to the current COVID-19 pandemic, medical personnel should complete medical procedures wearing full personal protective equipment (PPE) for aerosol-generating procedures (AGP) [1]. PPE-AGP should be used especially during cardiopulmonary resuscitation, including airway management [2]. Emergency intubation using direct laryngoscopy carries a significant risk of failure. As many authors indicate, the effectiveness of the first intubation attempt with a Macintosh laryngoscope is from 57.6% to 89.9% [3,4]. The use of PPE-AGP may reduce the efficiency of medical procedures and prolong their duration [5,6]. intubation methods alternative to Macintosh laryngoscope may be a suitable solution. An example of a new type of laryngoscope is Vie Scope® (Adroit Surgical, Oklahoma City, USA), which is a self-contained, battery-powered, disposable scope that takes advantage of a closed circular tube with a beveled end to visualize the vocal cords (Fig. 1 ). The light is transmitted through the sidewall of the tube from end to end as well as within the lumen of the tube. The intubation procedure using Vie Scope® should be based on the following steps – the intubator should: (1) insert scope and identify glottis opening between vocal cords, (2) pass the bougie under direct vision between the vocal cords into the trachea, (3) remove the scope leave the bougie in place, (4) pass the endotracheal tube over the bougie into the trachea, (5) remove the bougie, (6) ventilate.
Fig. 1

Intubation using Vie Scope laryngoscope

Intubation using Vie Scope laryngoscope The aim of this study was to evaluate intubation performance by paramedics wearing PPE-AGP using Macintosh laryngoscope and Vie Scope® laryngoscope under simulated resuscitation of COVID-19 patient. The study was designed as a prospective randomized crossover simulation trial. The study protocol was approved by the Institutional Review Board of the Polish Society of Disaster Medicine (No. 15.11.2019.IRB). 42 working paramedics took part in the study. Voluntary informed consent was obtained from each participant. None of the participants had previous experience with the Vie Scope® laryngoscope. Before the examination, the participants took part in the presentation of the laryngoscope and the method of correct intubation using it. The study participants then attended a 20-minute practical training course during which they performed endotracheal intubation using Vie Scope® under normal airway conditions. During the target study, paramedics dressed in full PPE-AGP were to perform endotracheal intubation with continuous chest compression. Endotracheal intubation was performed using Macintosh laryngoscope (MAC; blade no.3) as well as with Vie Scope® laryngoscope (VSC). The advanced SimMan 3G adult patient simulator (Stavanger, Norway) was used to simulate a patient requiring endotracheal intubation. A standard intubation guide was used for MAC and a bougie guide for VSC. Each participant had a maximum of 3 attempts to intubate using each device. Both the order of participants and intubation methods were random. A detailed randomization procedure is presented in Supplementary Fig. 1. The data were blinded for the team interpreting the results. The results were analyzed using the statistical package STATISTICA 13.3EN (Tibco Inc., USA) or Review Manager 5.4EN (Cochrane Collaboration, Oxford, UK). Group differences in dichotomous data are expressed as odds ratios (ORs) and group differences in continuous data as mean differences (MDs), both with 95% confidence intervals (CIs). The fixed-effect model was used to pool the results.
Supplementary Fig. 1

Randomization flow chart.

The study was designed as a prospective randomized crossover simulation trial. The study protocol was approved by the Institutional Review Board of the Polish Society of Disaster Medicine (No. 15.11.2019.IRB). 42 working paramedics took part in the study. Voluntary informed consent was obtained from each participant. None of the participants had previous experience with the Vie Scope® laryngoscope. Before the examination, the participants took part in the presentation of the laryngoscope and the method of correct intubation using it. The study participants then attended a 20-minute practical training course during which they performed endotracheal intubation using Vie Scope® under normal airway conditions. During the target study, paramedics dressed in full PPE-AGP were to perform endotracheal intubation with continuous chest compression. Endotracheal intubation was performed using Macintosh laryngoscope (MAC; blade no.3) as well as with Vie Scope® laryngoscope (VSC). The advanced SimMan 3G adult patient simulator (Stavanger, Norway) was used to simulate a patient requiring endotracheal intubation. A standard intubation guide was used for MAC and a bougie guide for VSC. Each participant had a maximum of 3 attempts to intubate using each device. Both the order of participants and intubation methods were random. A detailed randomization procedure is presented in Supplementary Fig. 1. The data were blinded for the team interpreting the results. The results were analyzed using the statistical package STATISTICA 13.3EN (Tibco Inc., USA) or Review Manager 5.4EN (Cochrane Collaboration, Oxford, UK). Group differences in dichotomous data are expressed as odds ratios (ORs) and group differences in continuous data as mean differences (MDs), both with 95% confidence intervals (CIs). The fixed-effect model was used to pool the results. Duration of intubation when one attempt needed between distinct intubation methods varied and amounted to 44 (IQR; 40.5–53)sec. For MAC and 28.5(IQR; 25–38.5)sec. For VSC (MD = 15.30; 95%CI: 13.69, 16.91; p < 0.001). Median overall intubation time with MAC was 47 s (IQR; 41.5–95.5) and was statistically significantly longer than with VSC – 29 s. (IQR; 25–39; MD = 27.30; 95%CI: 22.43, 32.17; p < 0.001). First attempt success rate with MAC was 50.0% vs. 92.9% for VSC (MD = 0.08; 95%CI: 0.02, 0.29; p < 0.001), and overall success rate was 90.5 vs. 100% respectively (MD = 0.10; 95%CI: 0.01, 1.93; p = 0.13). The ease of intubation is based on the VAS scale (0 – „easy procedure” to „10″ – difficult procedure) varied and amouted to 5 (IQR, 4–8) vs. 2 (IQR; 2–5) points (MD = 2.75; 95%CI: 2.34, 3.16; p < 0.001). In conclusion, under the conditions of intubation performed by paramedics wearing PPE-AGP with continuous chest compression, the results of the study indicate higher efficiency of intubation with Vie Scope® compared to Macintosh laryngoscope in terms of both the efficiency of the first intubation attempt and the time of the procedure. Further studies are required to confirm the results. The following is the supplementary data related to this article. Randomization flow chart. Supplementary data to this article can be found online at https://doi.org/10.1016/j.ajem.2020.08.085.

Declaration of Competing Interest

None.
  3 in total

1.  Comparison of Vie Scope® and Macintosh laryngoscopes for intubation during resuscitation by paramedics wearing personal protective equipment.

Authors:  Lukasz Szarpak; Frank W Peacock; Zubaid Rafique; Jerzy R Ladny; Klaudiusz Nadolny; Marek Malysz; Marek Dabrowski; Francesco Chirico; Jacek Smereka
Journal:  Am J Emerg Med       Date:  2022-01-04       Impact factor: 2.469

2.  Elective Tracheal Intubation With the VieScope-A Prospective Randomized Non-inferiority Pilot Study (VieScOP-Trial).

Authors:  Martin Petzoldt; Yasmin Engels; Zohal Popal; Pischtaz A Tariparast; Phillip B Sasu; Andrés Brockmann; Mark A Punke; Jörn Grensemann
Journal:  Front Med (Lausanne)       Date:  2022-03-15

3.  Direct vs. Video-Laryngoscopy for Intubation by Paramedics of Simulated COVID-19 Patients under Cardiopulmonary Resuscitation: A Randomized Crossover Trial.

Authors:  Leszek Gadek; Lukasz Szarpak; Lars Konge; Marek Dabrowski; Dominika Telecka-Gadek; Maciej Maslanka; Wiktoria Laura Drela; Marta Jachowicz; Lukasz Iskrzycki; Szymon Bialka; Frank William Peacock; Jacek Smereka
Journal:  J Clin Med       Date:  2021-12-08       Impact factor: 4.241

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.