Matthias Halper1, Marc-André Delsuc2,3, Kathrin Breuker1, Maria A van Agthoven1. 1. Institute for Organic Chemistry, University of Innsbruck, Innrain 80/82, 6020 Innsbruck, Austria. 2. Institut de Génétique et de Biologie Moléculaire et Cellulaire, INSERM U596, UMR 7104, Université de Strasbourg, 1 rue Laurent Fries, 67404 Illkirch-Graffenstaden, France. 3. CASC4DE, Pôle API, 300 Bd. Sébastien Grant, 67400 Illkirch-Graffenstaden, France.
Abstract
Two-dimensional mass spectrometry (2D MS) on a Fourier transform ion cyclotron resonance (FT-ICR) mass analyzer allows for tandem mass spectrometry without requiring ion isolation. In the ICR cell, the precursor ion radii are modulated before fragmentation, which results in modulation of the abundance of their fragments. The resulting 2D mass spectrum enables a correlation between the precursor and fragment ions. In a standard broadband 2D MS, the range of precursor ion cyclotron frequencies is determined by the lowest mass-to-charge (m/z) ratio to be fragmented in the 2D MS experiment, which leads to precursor ion m/z ranges that are much wider than necessary, thereby limiting the resolving power for precursor ions and the accuracy of the correlation between the precursor and fragment ions. We present narrowband modulation 2D MS, which increases the precursor ion resolving power by reducing the precursor ion m/z range, with the aim of resolving the fragment ion patterns of overlapping isotopic distributions. In this proof-of-concept study, we compare broadband and narrowband modulation 2D mass spectra of an equimolar mixture of histone peptide isoforms. In narrowband modulation 2D MS, we were able to separate the fragment ion patterns of all 13C isotopes of the different histone peptide forms. We further demonstrate the potential of narrowband 2D MS for label-free quantification of peptides.
Two-dimensional mass spectrometry (2D MS) on a Fourier transform ion cyclotron resonance (FT-ICR) mass analyzer allows for tandem mass spectrometry without requiring ion isolation. In the ICR cell, the precursor ion radii are modulated before fragmentation, which results in modulation of the abundance of their fragments. The resulting 2D mass spectrum enables a correlation between the precursor and fragment ions. In a standard broadband 2D MS, the range of precursor ion cyclotron frequencies is determined by the lowest mass-to-charge (m/z) ratio to be fragmented in the 2D MS experiment, which leads to precursor ion m/z ranges that are much wider than necessary, thereby limiting the resolving power for precursor ions and the accuracy of the correlation between the precursor and fragment ions. We present narrowband modulation 2D MS, which increases the precursor ion resolving power by reducing the precursor ion m/z range, with the aim of resolving the fragment ion patterns of overlapping isotopic distributions. In this proof-of-concept study, we compare broadband and narrowband modulation 2D mass spectra of an equimolar mixture of histone peptide isoforms. In narrowband modulation 2D MS, we were able to separate the fragment ion patterns of all 13C isotopes of the different histone peptide forms. We further demonstrate the potential of narrowband 2D MS for label-free quantification of peptides.
Histones are 11–21
kDa proteins that make up the chief components
of chromatin, the complex around which DNA is wound in the nucleus.[1] Histones frequently carry numerous post-translational
modifications (PTMs), which have been found to play a central role
in gene regulation, DNA repair, chromosome condensation, and spermatogenesis.
A functional understanding of histone modifications and the proposed
“histone code” requires far more experimental data than
are available so far.[2]Mass spectrometry
(MS) is ideally suited to address this challenge
as it can directly detect all mass-altering modifications and does
not require laborious biochemical techniques for analysis. Specifically,
MS can provide mass values of both the biomolecules under study and
their fragments from dissociation in tandem mass spectrometry (MS/MS)
experiments, for which prefractionation by chromatography is typically
required.[3,4] However, it is often challenging—if
not impossible—to fractionate very similar compounds, for example,
post-translationally modified proteins that differ only in their modification
patterns. Likewise, standard MS/MS experiments, in which each ionized
compound (precursor ion) in a sample is isolated before dissociation
and
detection of its fragment ions in the mass spectrometer, can be limited
by overlapping isotopic distributions of different peptide or protein
forms. Although isolation and fragmentation of a single histone isoform,[5] as well as the label-free, direct localization
and relative quantitation of histone[6−8] and ribonucleic acid,[8] modifications by MS/MS have been demonstrated,
a more general approach that does not rely on precursor ion isolation
would significantly advance all fields of PTM research.Two-dimensional
mass spectrometry (2D MS) is a method for Tandem
MS that enables the correlation between the precursor and fragment
ions without ion isolation.[9−11] 2D MS is a useful technique for
samples that are difficult to separate through chromatographic methods,
such as protein isoforms with relatively small modifications (e.g.,
methylations). Another advantage of 2D MS is the fact that all analytes
are subjected to the same experimental conditions, which makes 2D
MS a good candidate to perform quantification studies.2D MS
can be performed on Fourier transform ion cyclotron resonance
(FT-ICR) or linear ion trap mass spectrometers and has been successfully
applied to the analysis of small molecules, polymers, and in bottom-up
and top-down proteomics.[12−23] In most studies using 2D MS, the analytes were within an m/z range of several hundred units. In
the case of modified histones, the m/z range is typically less than 10 units depending on the charge state
of the protein.High-resolution mass spectrometers, that is,
the FT-ICR mass spectrometer
and Orbitrap, are used to resolve complex fragmentation patterns of,
for example, large peptides or proteins, with high mass accuracy.[24−26] Efficient 2D MS in a FT-ICR mass spectrometer requires a gas-free
fragmentation method such as infrared multiphoton dissociation, which
preferentially fragments the most labile bonds including those of
the PTMs of peptides and proteins, or electron capture dissociation
(ECD), which preferentially fragments the C-α backbone bonds
of peptides and proteins.[27−29] These characteristics make ECD
the method of choice for PTM analysis of peptides and proteins in
2D MS.[30]In standard broadband 2D
MS, the cyclotron frequency range for
precursor ions is determined by the lowest m/z ions
to be fragmented in the 2D MS experiment, which leads to m/z ranges that are much wider than necessary, thereby
limiting the resolving power of the precursor ions (corresponding
to the isolation m/z window in standard
MS/MS). In one-dimensional FT-ICR MS, heterodyne or “narrowband”
techniques have been developed to enable ultrahigh resolution.[31−33] We present narrowband modulation 2D MS, which is conceptually different
from one-dimensional narrowband MS and does not use heterodyne detection.
Narrowband modulation 2D MS increases the precursor ion resolving
power with the aim of differentiating the fragmentation patterns of
overlapping isotopic distributions (e.g., peptides and proteins with
and without disulfide bonds).In this study, we compare the
analysis of histone peptide isoforms
with broadband and narrowband modulation 2D MS. We further discuss
the use of 2D MS for the localization and assignment of PTMs, as well
as their relative quantification. In order to simplify this article,
narrowband modulation 2D MS will be referred to as “narrowband
2D MS”, although this is not meant to indicate an equivalence
to one-dimensional narrowband MS techniques.
Experimental Methods
Sample
Preparation
Milli-Q water (Merck Millipore,
Darmstadt), methanol, and acetic acid (VWR, Vienna, Austria) were
used in all experiments. As model peptides, C-terminal GK-biotinylated
histone H3 sequences (amino acid residues 21 to 44) without and with
native modifications (mono-, di-, and trimethylation) at K27 were
purchased (AnaSpec, Fremont, CA, USA) with a purity of >95%, in
which
K27 of the full-length histone corresponds to K7 of the model peptides.Peptides were desalted using MWCO 2000 Vivaspin centrifugal concentrators
(Sartorius, Göttingen, Germany) at 7900 rcf, 6× ammonium
acetate 100 mM (Sigma, Vienna, Austria), and 6× H2O. Peptide concentration was determined by UV absorption at 280 nm
using an Implen Nano PhotometerTM (Implen, München, Germany).
For electrospray ionization, equimolar mixtures (0.1 μM each)
of unmodified (referred to as “wild-type” in the text),
mono-, di-, and trimethylated peptides (referred to as “K7
1m”, “K7 2m”, and “K7 3m”, respectively)
in 50:50 H2O/CH3OH and 1% vol CH3COOH, pH ∼ 3.0 were prepared from 100 μM stock solutions
of each peptide in H2O.
Instrument Parameters
2D mass spectra were acquired
on a 7 T Apex Ultra FT-ICR mass spectrometer (Bruker Daltonik, GmbH,
Bremen, Germany) with an electrospray ion source operated in the positive
mode and direct injection at a flow rate of 90 μL/h. Ions were
accumulated for 0.1 s in the first hexapole and 0.1 s in the second
hexapole.[34] In the quadrupole, ions were
isolated at m/z 490 with an isolation
window of m/z 30 and transferred
to the Infinity ICR cell through a series of focusing lenses.[35]The experimental script for the acquisition
of 2D mass spectra is shown in Scheme . All excitation pulses (P1, P2, and P3) have a frequency range of 357,144.31–71,669.99
Hz (corresponding to m/z 301.08–1500).
Scheme 1
. Experimental Script of 2D MS
The frequency list was composed of 458 frequencies with a 624.71
Hz decrement. For the encoding pulses P1 and P2, the power level (corresponding to the amplitude of the pulses)
was set at 106 Vpp with a pulse length of 1.0 μs
per frequency. For the excitation pulse P3, the power level
was set at 80 Vpp with a pulse length of 20 μs per
frequency (default settings in the Apex Control software, Bruker Daltonik,
GmbH, Bremen, Germany).[36] The fragmentation
mode used was ECD, using a hollow cathode.[37] The heater was at 1.3 A, the lens at 20 V, the ECD bias at 2.0 V,
and the irradiation pulse at 0.03 s. The transients were recorded
with 1 Mwords (16 bits) for 1.4680 s, with a 714,288.62 Hz sampling
frequency.For the broadband 2D mass spectrum, the encoding
period t1 was incremented 4096 times with
a 2.0 μs
increment, leading to a 250 kHz Nyquist frequency (mass range: m/z 334.439–1500 with the high m/z limit set by the frequency range of
P1). The longest encoding period was 8.192 ms. For the
narrowband 2D mass spectrum, the encoding period t1 was incremented 4096 times with an 8.0 μs increment,
leading to a 62.5 kHz Nyquist frequency (mass range: m/z 415.00–546.825). The longest encoding
period was 32.768 ms. For neither 2D mass spectrum was the detection
heterodyne.
Data Processing
The 2D mass spectra
were processed
and visualized using the SPIKE software developed independently by
the University of Strasbourg and CASC4DE (Illkirch-Graffenstaden,
France) in a 64-bit Python 3.7 programming language on an open-source
platform distributed by the Python Software Foundation (Beaverton,
OR, USA).[38] Processed data files were saved
using the HDF5 file format. Both 2D mass spectra were digitally demodulated
with a demodulation frequency of 71,669.99 Hz and denoised using the
support selection for noise elimination (SANE) algorithm with a SANE
rank of 50.[10,39−41]For the
broadband 2D mass spectrum, cyclotron frequencies in the vertical
precursor ion dimension were calculated using the following equationin
which fICR is
the reduced cyclotron frequency of the precursor ion, f is the measured modulation frequency of the precursor ion, and f0 is the lowest frequency in the excitation
pulses (71,669.99 Hz). For the narrowband 2D mass spectrum, cyclotron
frequencies in the vertical precursor ion dimension were calculated
using the following equationin which fN is
the Nyquist frequency in the vertical precursor ion dimension (62,500.00
Hz).Each 2D mass spectrum was internally calibrated in the
horizontal
fragment ion m/z dimension using
a quadratic frequency-to-mass conversion.[42] For each precursor ion species, four to five fragment ion spectra
with the highest fragment ion intensities were added before peak-picking.
Theory of Narrowband 2D MS
Scheme shows the experimental script for 2D FT-ICR
MS. A detailed description of the mechanisms of the experiment can
be found in previous articles.[10,43,44] The radius modulation of precursor ions at the end of the encoding
sequence can be modeled according to the incremental delay t1in which r is the
radius of the ion packet in the ICR cell after the encoding sequence, r0 is the radius of the ion packet at the end
of the first pulse, fICR is the reduced
cyclotron frequency of the precursor ion, f0 is the lowest frequency in the encoding pulses, and T is the duration of the first encoding pulse.[9,10,43,44]Scheme (left)
shows the schematic of a broadband 2D spectrum in the frequency domain.
The modulation frequency in the vertical precursor frequency is fICR–f0. The
highest modulation frequency fmax1 in
the vertical dimension is determined by the encoding period (t1 in Scheme ) increment Δt1in which fS1 is
the sampling frequency in the vertical precursor dimension. As a result,
the range of precursor cyclotron frequencies in the 2D spectrum is
[f0, fS1/2].
Scheme 2
Schematics of 2D Mass Spectra Obtained in Broadband (Left) and Narrowband
Mode (Right)
According to the Shannon–Nyquist
theorem, periodically sampling
a signal at a sampling frequency of fS produces an aliasing of the signal frequencies such that all frequencies fN separated by fS of a multiple of fS produce the same
measurement f(45)The ± sign is due to the cosine in the modulation equation
for 2D MS.[10,43] Signals in 2D MS experiments
that have frequencies higher than fS1/2
can therefore be represented in aliased form in the 2D spectrum.In Scheme (left),
the autocorrelation line has modulation signals up to fS1/2 (precursor ion cyclotron frequencies are between f0 and f0 + fS1/2). If the sampling frequency is reduced
in the experiment by increasing the encoding period increment Δt1 (Scheme center), then the autocorrelation line is aliased
and can be folded back into the spectrum multiple times and appears
to zig–zag.The sampling frequency fS1′ and
the lowest excitation frequency f0 can
be chosen so that the modulation frequency
range of interest is folded back into the spectrum an even number
of times. The precursor ion cyclotron frequencies can be recovered
by offsetting all frequencies in the precursor ion dimension by NfS1′, where N is the number
of times that the signal of interest has been folded back into the
spectrum (Scheme right).
Subsequent frequency-to-mass conversion yields the narrowband 2D mass
spectrum. Although folding over the signal does not truncate the precursor
mass range, in practice quadrupole isolation is desirable for effective
data analysis.In cases where only a small m/z range is necessary in the vertical precursor
ion dimension, such
as a single charge state of a protein and its isoforms, narrowband
2D MS significantly reduces the vertical precursor m/z range, which significantly increases the resolving
power of each peak in the vertical precursor ion dimension and enables
more accurate precursor–fragment correlation.The total
acquisition time of a 2D MS experiment is proportional
to the number of t1 increments. As a result,
narrowband 2D MS can either be used to increase the vertical resolving
power by keeping a constant number of data points, or decreasing experiment
times and sample consumption and retaining the same resolving power.
Results and Discussion
Figure shows the
full broadband (Figure a) and narrowband (Figure b) 2D mass spectra of a mixture of the four isoforms of the
H3 histone peptide. In both 2D mass spectra, fragment ion m/z ratios are plotted horizontally. Precursor
ion m/z ratios, calculated after
conversion from measured frequencies to reduced cyclotron frequencies
with eq (broadband)
and eq (narrowband),
are plotted vertically. In both 2D mass spectra, the autocorrelation
line has been extracted and shows the isotopic distribution of the
[M + 6H]6+ ions of the four isoforms, evenly spaced by m/z 2.335942 (corresponding to 14.015652
Da for CH2 divided by z = 6). Because
of the quadrupole isolation, these species are the only ones detected
on the autocorrelation line.
Figure 1
. (a) Broadband 2D ECD mass spectrum of histone
peptide isoforms
(wild-type, 7K monomethylated or K7 1m, 7K dimethylated
or K7 2m, and 7K trimethylated or K7 3m). Inset: extracted
autocorrelation line, sequence coverage of wild-type histone peptide.
(b) Narrowband 2D ECD mass spectrum of histone peptide isoforms (wild-type, 7K monomethylated, 7K dimethylated, and 7K trimethylated). Inset: extracted autocorrelation line, sequence
coverage of wild-type histone peptide.
. (a) Broadband 2D ECD mass spectrum of histone
peptide isoforms
(wild-type, 7K monomethylated or K7 1m, 7K dimethylated
or K7 2m, and 7K trimethylated or K7 3m). Inset: extracted
autocorrelation line, sequence coverage of wild-type histone peptide.
(b) Narrowband 2D ECD mass spectrum of histone peptide isoforms (wild-type, 7K monomethylated, 7K dimethylated, and 7K trimethylated). Inset: extracted autocorrelation line, sequence
coverage of wild-type histone peptide.The resolving power on the autocorrelation line is approximately
147,000 at m/z 490 in both 2D mass
spectra.In both 2D mass spectra, the ECD fragmentation patterns
of the
peptide isoforms are represented in the horizontal fragment ion spectra.
The sequence coverage of the unmodified (wild-type) peptide is shown
in Figure a,b for
the respective 2D mass spectra. In both cases, the sequence coverage
is nearly complete. The peak assignments for all fragment ion spectra
can be found in the Supporting Information for the broadband 2D mass spectrum (Tables S1–S4) and the narrowband 2D mass spectrum (Tables S5–S8). The sequence coverage for all four isoforms
can be found in Figure S1 of the Supporting Information for the broadband 2D mass spectrum and Figure S2 for the narrowband 2D mass spectrum. For all isoforms, in
both 2D mass spectra, the sequence coverage is nearly complete (over
20 out of 25 cleavages).In the broadband 2D mass spectrum,
the vertical range is m/z 334.439–1500.
Only 0.77% of
the vertical m/z range contains
analytically useful information (m/z 486–495 for precursor ions) because it is determined by the
Nyquist frequency, which has to be higher than the cyclotron frequency
of the lowest m/z ratio of interest.
In the narrowband 2D mass spectrum, the vertical mass range is m/z 415.00–546.825, which means
that 6.8% contains useful analytical information. In the narrowband
2D mass spectrum, the vertical m/z range is still determined by the Nyquist frequency, but the foldover
effect is used to reduce it.In both 2D mass spectra, vertical
harmonics are clearly visible.
The vertical harmonics, as well as residual scintillation noise, are
clearly visible in the 2D mass spectra in Figure because of the low relative abundance of
fragment ion peaks compared to the precursor ion peaks in ECD.[40,46] The vertical harmonics in 2D mass spectra arise from the non-sinusoidal
modulation of signals during the 2D MS experiment.[36,43,44,47] Their relative
intensity can be minimized by adjusting the modulation zone in the
ICR cell to the fragmentation zone during the encoding sequence (Scheme ), but they cannot
be eliminated.[36] A previous study has shown
that a careful examination of the harmonics in precursor ion spectra
can uncover additional peaks for those fragment ions whose fragmentation
zones differ significantly from a Gaussian profile (e.g., in cases
of secondary fragmentation or combined fragmentation methods, such
as IR-ECD).[48]In the narrowband 2D
mass spectrum (Figure b), the 2f harmonic is visible
(the 3f harmonic has a too low signal-to-noise ratio
to be visible). The absence of precursor ion peaks on the autocorrelation
line clearly marks the fragment ion spectra as harmonics. In the broadband
2D mass spectrum (Figure a), the aliased 2f and 3f harmonics are visible. The foldover effect is marked by the fact
that, in both cases, the slopes of the dissociation lines are negative.
The signal from vertical harmonics can be used, in a similar way to
the signal from harmonics in one-dimensional FT-ICR mass spectra,
to increase the resolving power in the vertical dimension and therefore
to improve the accuracy of the precursor–fragment correlation.[49−51] Although the smaller mass range in the precursor ion dimension of
narrowband 2D mass spectra may lead to a decrease in harmonic signals,
they may be folded back into the spectrum by aliasing.Figure shows the
vertical precursor ion spectra extracted from both the broadband and
the narrowband 2D mass spectrum. Figure a shows the vertical precursor ion spectrum
of the monoisotopic c5 fragment at m/z 460.2879. Fragment c5 does not contain any modification, and therefore, the
precursor ion spectrum shows peaks for all four isoforms. In the vertical
precursor ion spectrum extracted from the broadband 2D mass spectrum,
the isotopic peaks of the four precursor ions of c5 are not resolved (black circles show the theoretical
isotopic peaks in the vertical dimension). In contrast, the isotopic
distribution of the precursor ions is resolved in the precursor ion
spectrum extracted from the narrowband 2D mass spectrum, albeit not
to the baseline.
Figure 2
(a) Normalized precursor ion spectrum of c5 extracted from the broadband and the narrowband 2D mass
spectra
(black dots: theoretical isotopic distribution). (b) Normalized precursor
ion spectrum of z244+ extracted
from the broadband and the narrowband 2D mass spectra at the first
harmonic (black dots: theoretical isotopic distribution).
(a) Normalized precursor ion spectrum of c5 extracted from the broadband and the narrowband 2D mass
spectra
(black dots: theoretical isotopic distribution). (b) Normalized precursor
ion spectrum of z244+ extracted
from the broadband and the narrowband 2D mass spectra at the first
harmonic (black dots: theoretical isotopic distribution).The vertical resolving power in both spectra can be calculated
from the extracted precursor ion spectra shown in Figure b, which correspond to the
monoisotopic z244+ fragment
of the wild-type peptide at m/z 682.8849
(to help compare between the two peaks, the spectra have been normalized,
which makes the noise level of the spectrum extracted from the narrowband
2D mass spectrum appear higher).The z244+ fragment contains
the K7 residue on which all modifications (methylation, dimethylation,
and trimethylation) are located, so the precursor ion spectrum only
shows a peak for the wild-type peptide. In addition, the z244+ fragment contains 24 out of 26 residues.
Any 13C isotope in the precursor ion has a probability
close to 100% to remain in the z244+ fragment instead of its complement c2.[10,52] Therefore, the monoisotopic z244+ fragment is the product of the monoisotopic
peak of the wild-type precursor with negligible contribution from
any 13C isotopes of the precursor ion, and the extracted
precursor ion spectra in Figure b can be used to accurately determine the vertical
resolving power and the accuracy of the precursor–fragment
correlation.The isotopic distributions of fragment ions in
2D mass spectra
are further illustrated in Figure S4 in the Supporting Information.As expected, the vertical resolving power
in the narrowband 2D
mass spectrum (3400 at m/z 487)
is almost 4 times the vertical resolving power in the broadband 2D
mass spectrum (960 at m/z 487).
The vertical resolving power measures the accuracy of the precursor–fragment
correlation. The horizontal resolving power corresponds to the standard
resolving power of one-dimensional mass spectra. This result is consistent
with the fact that the frequency range in the broadband 2D mass spectrum
is 4 times the frequency range in the narrowband 2D mass spectrum
with an equal number of data points in the vertical dimension.Figure shows the
vertical precursor ion spectra of all four isotopes of the z244+ fragment of the [M + 6H]6+ charge state of the wild-type histone peptide. Figure a shows the first
harmonic peak (i.e., corresponding to the modulation frequency of
the precursor ion radius) and Figure b shows its 2f harmonic peak (or second
harmonic) with an adapted frequency-to-mass conversion as extracted
from the broadband 2D mass spectrum. Figure c shows the first harmonic peak and Figure d shows its second
harmonic peak as extracted from the narrowband 2D mass spectrum.
Figure 3
Precursor
ion spectra of the isotopes of the z244+ fragment of the wild-type histone peptide,
with frequency-to-mass conversion extracted from the broadband 2D
mass spectrum for the (a) first harmonic peak and the (b) second harmonic
peak and extracted from the narrowband 2D mass spectrum for the (c)
first harmonic peak and the (d) second harmonic peak. The resolving
power is measured on the 12C isotopic peak, and the signal-to-noise
ratio is measured for the 1 × 13C isotopic peak.
Precursor
ion spectra of the isotopes of the z244+ fragment of the wild-type histone peptide,
with frequency-to-mass conversion extracted from the broadband 2D
mass spectrum for the (a) first harmonic peak and the (b) second harmonic
peak and extracted from the narrowband 2D mass spectrum for the (c)
first harmonic peak and the (d) second harmonic peak. The resolving
power is measured on the 12C isotopic peak, and the signal-to-noise
ratio is measured for the 1 × 13C isotopic peak.In Figure a (first
harmonic of the broadband 2D mass spectrum), the isotopic distribution
is not resolved (R = 960 for the 12C isotopic
peak). In Figure b,
the use of the second harmonic enables an increase in revolving power
(R = 2200), but the isotopic distribution is still
not resolved. The signal-to-noise ratio is slightly lower for the
second harmonic than in the first harmonic. The third harmonic is
shown in the Supporting Information, Figure
S5.In Figure c, while
the isotopic distribution of the [M + 6H]6+ precursor is
resolved (R = 3400 for the 12C isotopic
peak), the peaks are not baseline-resolved. In Figure d, the isotopic distribution is resolved
(R = 7700 for the 12C isotopic peak).
In the first harmonic, the signal-to-noise ratio of the 1 × 13C isotopic peak was calculated to be 110, but in the second
harmonic, it was calculated to be 50. The horizontal fragment ion
spectra for the second harmonic of all four isoforms (visible between
vertical m/z 435–450 in Figure b) were peak-picked.
The peak assignments are shown in the Supporting Information (Tables S9–S12).The sequence coverage
of all four isoforms in the second harmonic
(Supporting Information Figure S3) was
approximately 50%, compared to approximately 90% in the first harmonic.
Unlike in one-dimensional FT-ICR MS, where the relative intensity
of 2f–6f harmonics can be
maximized by adapting the design of ICR cells,[49−51] increasing
the relative intensity of higher harmonics in 2D FT-ICR MS comes at
the expense of overall signal-to-noise ratio and thus analytical information.[36] However, in 2D MS, higher harmonics in the vertical
dimension can be useful to establish precursor–fragment correlations
when first harmonic fragment ion peaks have insufficient resolving
power.Figure also highlights
the necessity to add up fragment ion spectra in the narrowband 2D
mass spectrum to obtain complete isotopic distributions. Further illustration
can be found in Figure S6 in the Supporting Information.Figure shows
how
the narrowband 2D mass spectrum enables the sequencing of the four
peptide isoforms, the identification and location of the modifications,
and their relative quantification. Figure a shows a zoom-in of the narrowband 2D mass
spectrum between m/z 600–700.
For each peptide isoform, horizontal fragment ion spectra were extracted
for each isotope and added up before peak-picking and peak assignment
(Tables S5–S8 and Figure S2 in the Supporting Information).
Figure 4
(a) Zoom-in of the narrowband 2D mass spectrum of the
histone peptide
mixture with various dissociation lines highlighted. (b) Extracted
dissociation line of the c142+ fragment. (c) Extracted dissociation line of the z244+ fragment. (d) Label-free relative quantification
of the four histone peptide isoforms using fragment abundances for
each residue (N-terminus and C-terminus combined). (e) Label-free
relative quantification of the four histone peptide isoforms using
the relative abundances of each charge state (6+ precursor ion, 5+
and 4+ charge-reduced states) and average of the quantification using
fragment ion abundances.
(a) Zoom-in of the narrowband 2D mass spectrum of the
histone peptide
mixture with various dissociation lines highlighted. (b) Extracted
dissociation line of the c142+ fragment. (c) Extracted dissociation line of the z244+ fragment. (d) Label-free relative quantification
of the four histone peptide isoforms using fragment abundances for
each residue (N-terminus and C-terminus combined). (e) Label-free
relative quantification of the four histone peptide isoforms using
the relative abundances of each charge state (6+ precursor ion, 5+
and 4+ charge-reduced states) and average of the quantification using
fragment ion abundances.Figure a illustrates
how to locate peptide modifications. In the 2D mass spectrum, precursor
ions of different m/z with fragments
of identical m/z (i.e., fragments
not containing the modified lysine) can all be found on the same vertical
precursor ion spectrum. In Figure a, c6 (m/z 616.3888), z163+ (m/z 640.3481), and z183+ (m/z 696.3790) all form vertical precursor ion spectra. Because
the methylations are situated on residue number 7 counting from the
N-terminus or residue number 19 counting from the C-terminus, these
vertical precursor ion spectra enable the accurate localization of
the modifications.The complementary ions for fragments with
identical m/z ratios can be found
on diagonal dissociation
lines following eq (10,16,48,53)in which (m/z)p is the m/z ratio
of the precursor ion, (m/z)f is the m/z ratio of the
fragment, n is the charge state of the precursor, p is the charge state of the fragment, and Δm is the mass difference between the precursor and the fragment.
In Figure a, c122+, c132+, and c142+ are
observed on dissociation lines with a slope of 2/6 = 0.33, c193+ on a dissociation line with
a slope of 3/6 = 0.5 and finally z214+, z224+, z234+, z244+, c234+, c244+, and c254+ on dissociation lines with a slope of 4/6 = 0.67 (linear
regressions available in Table S13 in the Supporting Information). The use of vertical precursor ion spectra and
diagonal dissociation lines can therefore be used to locate the sites
of peptide modifications.Figure b,c shows
the extracted dissociation lines of c142+ (Figure b) and z244+ (Figure c). The resolving power on
both dissociation lines is sufficient to calculate the exact mass
differences between the isoforms for correct assignments of the peptide
modifications. In Figure b, the modifications are measured to have masses of 14.0155,
28.0313, and 42.0465 Da, corresponding to mono-, di-, and trimethylation,
respectively. The unusual pattern for the K7 2m isotopic distribution
results from the overlap of the c142+ and z234+ fragment
ion distributions (see Figure a). Figure a–c therefore shows that a visual analysis of narrowband 2D
mass spectra can be used for peptide sequencing and to identify and
locate the modifications.Figure d,e shows
different methods to quantify the four peptide isoforms. In label-free
relative quantification studies with one-dimensional MS/MS, relative
intensities of ions with given mass differences are compared to locate
modifications and quantify isoforms on one graph.[8,54] With
2D MS, locating modifications can be accomplished visually (cf. Figure a) and relative intensities
can be plotted without distinguishing m/z ratios. In Figure d, the relative intensities are given for both N-terminus and C-terminus
fragments.Figure e shows
the comparison of the quantification obtained by using peaks on the
autocorrelation line (6+), the one-electron capture line (5+), the
two-electron capture line (4+), and the average quantification used
with the fragment ion peaks (fragment average). The proportions of
the four peptide isoforms are consistent between all the four methods.
The only discrepancy was in the proportions of the two-electron capture
line, where K7 2m was not detected, because of the low abundance of
the [M + 6H]4+•• charge state.
Conclusions
This study shows that decreasing the sampling frequency along the t1 axis (i.e., precursor ion modulation) in 2D
MS can substantially increase the resolving power in the vertical precursor ion m/z dimension for an unchanged acquisition time. Narrowband
2D MS can resolve isotopic distributions of the [M + 6H]6+ ions in a mixture of four histone peptide isoforms (26 residues).
The precursor–fragment correlation was high enough to enable
the detailed examination of fragment isotopic distributions in the
2D mass spectrum. Finally, the study shows the capability of 2D MS
to identify and locate histone modifications and to quantify their
relative abundances.As this study demonstrates, narrowband
2D MS has the potential
for top-down MS of biomolecules. The high precursor–fragment
correlation can be used to distinguish the fragmentation patterns
of overlapping isotopic distributions (e.g., for proteins with and
without disulfide bonds) and easily identify and locate modifications
by extracting dissociation lines. The capacity of 2D MS to analyze
all analytes in a sample with identical experimental parameters in
a single experiment enables label-free relative quantification. These
characteristics of 2D MS enable the analysis of heavily modified biomolecules,
such as histones, for the elucidation of the histone code.[55]
Authors: R A Zubarev; D M Horn; E K Fridriksson; N L Kelleher; N A Kruger; M A Lewis; B K Carpenter; F W McLafferty Journal: Anal Chem Date: 2000-02-01 Impact factor: 6.986
Authors: H J Simon; M A van Agthoven; P Y Lam; F Floris; L Chiron; M-A Delsuc; C Rolando; M P Barrow; P B O'Connor Journal: Analyst Date: 2015-11-16 Impact factor: 4.616
Authors: Maria A van Agthoven; Marie-Aude Coutouly; Christian Rolando; Marc-André Delsuc Journal: Rapid Commun Mass Spectrom Date: 2011-06-15 Impact factor: 2.419
Authors: Maria A van Agthoven; Marc-André Delsuc; Geoffrey Bodenhausen; Christian Rolando Journal: Anal Bioanal Chem Date: 2012-10-18 Impact factor: 4.142
Authors: Donald F Smith; Greg T Blakney; Steven C Beu; Lissa C Anderson; Chad R Weisbrod; Christopher L Hendrickson Journal: Anal Chem Date: 2020-02-03 Impact factor: 6.986
Authors: Maria A van Agthoven; Yuko P Y Lam; Peter B O'Connor; Christian Rolando; Marc-André Delsuc Journal: Eur Biophys J Date: 2019-03-13 Impact factor: 1.733