Literature DB >> 32957248

Fatigue behavior and damage modes of high performance poly-ether-ketone-ketone PEKK bilayered crowns.

Omar Alsadon1, Duncan Wood2, David Patrick2, Sarah Pollington2.   

Abstract

OBJECTIVES: To evaluate and compare the fatigue behavior (fatigue limit and fatigue life) and damage modes of high-performance poly-ether-ketone-ketone (PEKK), zirconia and alloy bilayered crowns.
MATERIALS AND METHODS: A total of 110 crowns (n = 50 for fatigue limit and n = 60 for fatigue life) were fabricated and used in this study. Pekkton® ivory discs, yttrium stabilized zirconia blanks and NiCr casting alloy were used to produce the respective PEKK, zirconia and alloy copings for crown fabrication. The prepared crowns were veneered with composite resin and subjected to fatigue tests. The fatigue limit was evaluated using the staircase method and the fatigue life of the samples was evaluated by subjecting the crowns to a load lower than the fatigue limit of that particular group, and also with an intermediate load of 522 N. A graphical plot was generated from the shape parameter (β) and life parameter (α) values obtained through the Weibull analysis method. Kruskal-Wallis and Mann-Whitney tests were applied to determine the significance differences in the recorded fracture mode between the study groups. The damage modes of the samples were assessed using Burke's classification.
RESULTS: The recorded fatigue limits of the groups were 442.8 ± 42.1 N, 608.7 ± 7.6 N, and 790.4 ± 29.2 N for zirconia, NiCr and PEKK, respectively. A significant difference in the fatigue limit of the groups was observed (p < 0.05). PEKK samples demonstrated the highest survival cycles of 1,170,000 and the lowest survival cycles was observed with zirconia samples at 100,000 under 522 N loading. The fracture modes in PEKK samples were largely distributed between code 1 and 2 whereas the fracture modes in NiCr group was distributed between code 1 and 4 and YZ crowns exhibited more of code 5 fractures. The difference in fracture modes among the groups was statistically significant (p < 0.05).
CONCLUSION: The PEKK group demonstrated better results compared to zirconia and NiCr based crowns. The PEKK group demonstrated high fatigue limit and survived the highest fatigue life cycles among the tested groups.
Copyright © 2020 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Composite veneer; Damage modes; Fatigue life; Fatigue limit; PEKK; Zirconia

Mesh:

Substances:

Year:  2020        PMID: 32957248     DOI: 10.1016/j.jmbbm.2020.103957

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  4 in total

Review 1.  Polyetherketoneketone (PEKK): An emerging biomaterial for oral implants and dental prostheses.

Authors:  Hatim Alqurashi; Zohaib Khurshid; Azeem Ul Yaqin Syed; Syed Rashid Habib; Dinesh Rokaya; Muhammad Sohail Zafar
Journal:  J Adv Res       Date:  2020-09-18       Impact factor: 10.479

Review 2.  New generation CAD-CAM materials for implant-supported definitive frameworks fabricated by using subtractive technologies.

Authors:  Pinar Cevik; Martin Schimmel; Burak Yilmaz
Journal:  Biomed Res Int       Date:  2022-03-02       Impact factor: 3.411

3.  Comparison of tensile bond strength of ball attachments made of different materials to root canal dentin after chewing simulation.

Authors:  Ahmed Mostafa Abdelfattah Mohamed; Mahmoud El-Moutassem Bellah El Homossany; Sarah Mohamed Abdelmoniem; Tariq Yehia Abdelrahman
Journal:  BMC Oral Health       Date:  2022-06-24       Impact factor: 3.747

4.  Mechanical Response of PEKK and PEEK As Frameworks for Implant-Supported Full-Arch Fixed Dental Prosthesis: 3D Finite Element Analysis.

Authors:  Regina Furbino Villefort; Pedro Jacy Santos Diamantino; Sandra Lúcia Ventorin von Zeidler; Alexandre Luiz Souto Borges; Laís Regiane Silva-Concílio; Guilherme deSiqueira Ferreira Anzaloni Saavedra; João Paulo Mendes Tribst
Journal:  Eur J Dent       Date:  2021-09-24
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.