Minimum 10-Year Follow-Up of Cementless Total Hip Arthroplasty With a 32-mm Cobalt-Chromium Head on Highly Cross-Linked Polyethylene and a Tapered, Fiber Metal Proximally Coated Femoral Stem.

Yohei Naito1, Masahiro Hasegawa1, Shine Tone1, Hiroki Wakabayashi1, Akihiro Sudo1.   

Abstract

BACKGROUND: The clinical and radiographic results of cementless total hip arthroplasty using a 32-mm cobalt-chromium head on remelted highly cross-linked polyethylene (HXLPE) and a tapered, fiber metal proximally coated femoral stem were evaluated at a mean follow-up of 12 years.
METHODS: A total of 57 cementless total hip arthroplasties using remelted HXLPE combined with a 32-mm cobalt-chromium head, and a tapered, fiber metal proximally coated femoral stem were performed from October 2004 to December 2006. Clinical evaluation was performed using the Merle d'Aubigné and Postel scoring system. Standardized anteroposterior and lateral radiographs of the pelvis and femur without weight-bearing were analyzed. Radiographic measurements of two-dimensional femoral head penetration into the polyethylene were performed with a computerized method. The steady-state wear rates were measured based on the radiographs from the first year after surgery to the final follow-up.
RESULTS: The mean Merle d'Aubigné and Postel score improved significantly from 10.7 points preoperatively to 15.6 postoperatively (P < .001). No osteolysis was found around the implant. Stem fixation in all cases showed stable bone ingrowth. Third-degree stress shielding was found in 46% of all stems. The total head penetration rate was 0.05 mm/y, and the steady-state wear rate was 0.01 mm/y. The Kaplan-Meier survivorship with the end point of revision was 95% (95% confidence interval, 85%-98%) at 12 years.
CONCLUSION: A 32-mm cobalt-chromium head on remelted HXLPE demonstrated low wear properties, and a tapered, fiber metal proximally coated femoral stem showed good results at long-term follow-up.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  32-mm cobalt-chromium head; fiber metal proximally coated femoral stem; highly cross-linked polyethylene; long-term results; tapered; total hip arthroplasty; wear

Year:  2020        PMID: 32950339     DOI: 10.1016/j.arth.2020.08.055

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  2 in total

1.  Open-porous magnesium-based scaffolds withstand in vitro corrosion under cyclic loading: A mechanistic study.

Authors:  Roxane Bonithon; Colin Lupton; Marta Roldo; Joseph Nicholas Dunlop; Gordon William Blunn; Frank Witte; Gianluca Tozzi
Journal:  Bioact Mater       Date:  2022-04-29

2.  Effects of porous tantalum on periprosthetic bone remodeling around metaphyseal filling femoral stem: a multicenter, prospective, randomized controlled study.

Authors:  Goro Motomura; Naohiko Mashima; Hiroshi Imai; Akihiro Sudo; Masahiro Hasegawa; Harumoto Yamada; Mitsuhiro Morita; Naoto Mitsugi; Ryosuke Nakanishi; Yasuharu Nakashima
Journal:  Sci Rep       Date:  2022-01-18       Impact factor: 4.379