| Literature DB >> 32942646 |
Timo Schuett1,2, Julian Kimmig1,2, Stefan Zechel1,2, Ulrich S Schubert1,2.
Abstract
The automated dialysis of polymers in synthetic robots is described as a first approach for the purification of polymers using an automated protocol. For this purpose, a dialysis apparatus was installed within a synthesis robot. Therein, the polymer solution could be transferred automatically into the dialysis tube. Afterwards, a permanent running dialysis could be started, enabling the removal of residual monomer. Purification efficiency was studied using chromatography and NMR spectroscopy, showing that the automated dialysis requires less solvent and is faster compared to the classical manual approach.Entities:
Keywords: automation; dialysis; high-throughput experimentations; polymer purification
Year: 2020 PMID: 32942646 PMCID: PMC7569804 DOI: 10.3390/polym12092095
Source DB: PubMed Journal: Polymers (Basel) ISSN: 2073-4360 Impact factor: 4.329
Summary of the utilized amounts and volumes for the Reversible Addition-Fragmentation Chain Transfer (RAFT) polymerizations of methyl methacrylate (MMA).
| Polymer | ||||
|---|---|---|---|---|
|
| 50 | 250 | 736.91 | 136.68 |
|
| 15 | 75 | 221.07 | 41.00 |
Summary of the molar masses of the polymers P1–P3 (molar mass was determined using size exclusion chromatography measurements (SEC); standard PMMA for P1 and P2, polystyrene for P3; solvent: chloroform/isopropanol/triethyl-amine [94/2/4]).
| Polymer | Ɖ | ||
|---|---|---|---|
|
| 6800 | 7800 | 1.14 |
|
| 12,600 | 14,300 | 1.13 |
|
| 42,300 | 64,000 | 1.51 |
Summary of the dialysis experiments.
| Experiment | Modus Operandi | Polymer | Additive | Solvent Change |
|---|---|---|---|---|
|
| Manual | P3 | Sudan 1 | 400 mL after 12 h; 55× in total |
|
| Manual | MMA | 400 mL after 12 h; 2× in total | |
|
| Manual | – | 400 mL after 12 h; 2× in total | |
|
| Automated (modus 1) | Sudan 1 | 50 mL/h; 121 h | |
|
| Automated (modus 1) | MMA | 35 mL/h; 32 h | |
|
| Automated (modus 1) | – | 35 mL/h; 32 h | |
|
| Automated (modus 2) | – | 250 mL after 3 h; 5× in total |
Figure 1Automated dialysis apparatus installed within the synthesis robot. The 250 mL glass vessel was equipped with a special Teflon cap consisting of an injection port for the four-needle head of the robot as well as possibilities for the solvent exchange (solvent pumping). The dialysis tube is located within the vessel filled with a solution of P2.
Figure 2Time-dependent analysis of the dialysis of P3 and Sudan I mixture (automated dialysis modus 1 top; manual dialysis bottom).
Figure 3Representation of the decrease in the residual DMF (M3, A3, A4) or MMA (M2, A2) during dialysis time.
Analysis of the efficiency of the manual (M2, M3) and the automated (A2 to A4) dialysis measured using gas chromatography and NMR spectroscopy. In the case of the reaction solution, the residual solvent (dimethylformamide (DMF)) was measured (M3, A3, A4) (NMR signal at 8.02 ppm compared to signal of anisole at 6.75 to 6.90 ppm), in the case of the other (M2, A2) the additional methyl methacrylate (NMR signal at 5.5 ppm compared to signal of anisole at 6.75 to 6.90 ppm) was analyzed. The values are given in percentages relative to the sample at the beginning of the experiments.
| Time | M2 | M3 | A2 | A3 | A4 | |||||
|---|---|---|---|---|---|---|---|---|---|---|
|
| GC | NMR | GC | NMR | GC | NMR | GC | NMR | GC | NMR |
|
| 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
|
| 44 | 39 | 18 | 16 | 40 | 32 | 11 | 12 | 20 | 18 |
|
| 28 | 16 | 5 | 5 | 23 | 11 | 5 | 5 | 2 | 2 |
|
| 18 | 8 | 4 | 3 | 17 | 6 | 3 | 3 | 0.5 | 0 |
|
| 16 | 5 | 4 | 3 | 15 | 4 | 2 | 3 | 0 | 0 |
|
| 16 | 2 | 2 | 1 | 14 | 2 | 1 | 0 | 0 | 0 |
|
| 11 | 1 | 1 | 1 | 13 | 1 | 1 | 1 | 0 | 0 |
|
| 11 | 0 | 0.05 | 0 | 12 | 0 | 0.5 | 1 | 0 | 0 |
Figure 4NMR kinetic (CDCl3) of the automated dialysis A4 (modus 2) using the DMF signal at 8.02 ppm (green 0 h, blue 3 h, red 9 h, black 24 h) compared to aromatic anisole signals at 6.75 to 6.90 ppm.