Abstract
Ultrasound scanning of surgical wounds is an established procedure for the detection of abscesses. The possible risks of cross-infection resulting from this technique were examined by testing the sterility of the ultrasound probes, the coupling gel and the stand-off medium Kitecko (3 M). The coupling gel was also assessed for any bactericidal properties. Sixty-six per cent of swabs taken from machines in constant use and 33% of the total number of swabs taken were contaminated with skin flora including Staphylococcus aureus. Sterility was achieved using a 70% alcohol wipe. The coupling gel was inherently sterile but had no bactericidal action. The solid stand-off medium Kitecko grew Enterobacteriaceae and Pseudomonas species. The implications of these findings in relation to scanning post-operative wounds are discussed.Entities: