| Literature DB >> 32933070 |
Alicia Hauke1, Susanne Oertel1, Leona Knoke2, Vanessa Fein2, Christoph Maier2, Folke Brinkmann2, Michael P M Jank1.
Abstract
Analysis of sweat chloride levels in cystic fibrosis (CF) patients is essential not only for diagnosis but also for the monitoring of therapeutic responses to new drugs, such as cystic fibrosis transmembrane conductance regulator (CFTR) modulators and potentiators. Using iontophoresis as the gold standard can cause complications like burns, is uncomfortable, and requires repetitive hospital visits, which can be particularly problematic during a pandemic, where distancing and hygiene requirements are increased; therefore, it is necessary to develop fast and simple measures for the diagnosis and monitoring of CF. A screen-printed, low-cost chloride sensor was developed to remotely monitor CF patients. Using potentiometric measurements, the performance of the sensor was tested. It showed good sensitivity and a detection limit of 2.7 × 10-5 mol/L, which covered more than the complete concentration range of interest for CF diagnosis. Due to its fast response of 30 s, it competes well with standard sensor systems. It also offers significantly reduced costs and can be used as a portable device. The analysis of real sweat samples from healthy subjects, as well as CF patients, demonstrates a proper distinction using the screen-printed sensor. This approach presents an attractive remote measurement alternative for fast, simple, and low-cost CF diagnosis and monitoring.Entities:
Keywords: chloride; cystic fibrosis; ion-selective electrode; screen-printed sensor; sweat analysis
Mesh:
Substances:
Year: 2020 PMID: 32933070 PMCID: PMC7558764 DOI: 10.3390/bios10090123
Source DB: PubMed Journal: Biosensors (Basel) ISSN: 2079-6374
Figure 1(a) Explosion view of screen-printed chloride sensor depicting the layers. (i) Silver (dark grey), (ii) silver/silver chloride (light grey), (iii) carbon (black), and (iv) insulation/encapsulation (green). (b) Manufactured sample sheet of sensors before deposition of the selective and reference membranes.
Membrane compounds of the chloride-selective membrane using the chloride ionophore I –cocktail A with its corresponding portion in parentheses.
| Chloride Ionophore I—Cocktail A | Polymer | Solvent | ||
|---|---|---|---|---|
| Ionophore | Plasticizer | Anionic Site | ||
| Chloride ionophore I | 2-nitrophenyl octyl ether + 1-Decanol | Tridodecylmethyl-ammonium chloride | PVC | 1 mL THF on 100 mg compounds |
| (3.5%) | (65.8%) | (0.7%) | (30.0%) | |
Figure 2(a) Sweat inducement by pilocarpine iontophoresis using two electrodes attached to the forearm; (b) irritation of skin after iontophoresis due to the applied electric current; (c) test tube pressed onto the forearm in the former position of the pilocarpine electrode to collect sweat; and (d) small amount of collected sweat with test tube after pilocarpine iontophoresis.
Figure 3(a) Potentiometric response curve of the screen-printed chloride sensor measured in NH4Cl solutions with different concentrations. (b) Resulting calibration plot measured in NH4Cl solutions of four different chloride-selective electrodes sharing one reference electrode.
Comparison of screen-printed chloride sensor to the standard sensor system.
| Parameters | Screen-Printed Chloride Sensor | Coulometric Titration |
|---|---|---|
| Detection limit | 0.03 mmol/L | 10 mmol/L |
| Response time | <1 min | <1 min |
| System preparation | 5 min | 5 min |
| Sample preparation | - | + |
| Cost of equipment | <50 € | 8200 € |
| Cost of assay | <3 € | ca. 5 € |
| Portability | + | - |
| Usability | + | - |
1 by Kreienbaum.
The determined chloride concentration of sweat samples and categorization in health status.
| Sample | c(Cl−) in mmol/L | Status 1 (Healthy or CF) |
|---|---|---|
| 1 | 0.9 | healthy |
| 2 | 1.0 | healthy |
| 3 | 1.9 | healthy |
| 4 | 3.0 | healthy |
| 5 | 5.7 | healthy |
| 6 | 0.4 | healthy |
| 7 | 0.6 | healthy |
| 8 | 1.7 | healthy |
| 9 | 129.2 | CF |
| 10 | 100.0 | CF |
| 11 | 63.0 | CF |
1 Derived from different accepted test protocols.
Figure 4Measurement of sweat samples (samples 1, 2, 3, and 9) using the screen-printed chloride sensor for the detection of cystic fibrosis (CF). The dashed line displays the threshold of chloride concentration. Concentrations >60 mmol/L indicates CF disease.