Do directional deep brain stimulation leads rotate after implantation?

Marie T Krüger1,2, Yashar Naseri3,4, Fabian Cavalloni3, Peter C Reinacher5,6, Georg Kägi7, Johannes Weber8, Deborah Brogle3,7, Oliver Bozinov3, Stefan Hägele-Link7, Florian Brugger7.   

Abstract

BACKGROUND: The two middle contacts of directional leads (d-leads) for deep brain stimulation are split into three segments, allowing current steering toward desired axial directions. To facilitate programming, their final orientation needs to be reliably determined. However, it is currently unclear whether d-leads rotate after implantation. Our objective was to assess the degree of d-lead rotation after implantation.
METHODS: We retrospectively analyzed d-lead orientation on intraoperative X-rays, postoperative CT scans (latencies to surgery: 108-189 min postoperatively), and rotational fluoroscopies (4-9 days postoperatively) for a consecutive series of 32 implanted d-leads. For five d-leads, a CT scan with a mean follow-up of 57 days (range 28-182) was available. All d-leads were implanted with the marker facing anterior and the intention to hit an "iron sight" (ISi) on the X-ray, indicating anterior orientation (i.e., 0° ± 6°).
RESULTS: In nine d-leads, an ISi was visible on the final X-ray; median orientation was 1.5° (range 0.5-6.0°) at the first follow-up CT, confirming anterior orientation. In d-leads without ISi or where ISi was not evaluable, the median rotation was 15.5° (9.5-35.0°) and 26.5° (5.5-62.0°), respectively. The orientation of the initial CT was comparable with the orientation determined by the postoperative rotational fluoroscopy and second CT in all d-lead groups.
CONCLUSION: D-lead orientation does not change within the first week after implantation. We provide first indications that d-lead orientation remains stable for several weeks after surgery. Determination of lead orientation using marker-based X-ray alone seems too imprecise; adding the ISi method can increase determination of intraoperative orientation.

Entities:  

Keywords:  Deep brain stimulation; Directional leads; Iron sights; Rotational fluoroscopy

Mesh:

Year:  2020        PMID: 32915306     DOI: 10.1007/s00701-020-04568-3

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  3 in total

1.  Surgical Strategy for Directional Deep Brain Stimulation.

Authors:  Hiroshi Masuda; Hiroshi Shirozu; Yosuke Ito; Masafumi Fukuda; Yukihiko Fujii
Journal:  Neurol Med Chir (Tokyo)       Date:  2021-10-29       Impact factor: 1.742

2.  Local anatomy, stimulation site, and time alter directional deep brain stimulation impedances.

Authors:  Joseph W Olson; Christopher L Gonzalez; Sarah Brinkerhoff; Maria Boolos; Melissa H Wade; Christopher P Hurt; Arie Nakhmani; Bart L Guthrie; Harrison C Walker
Journal:  Front Hum Neurosci       Date:  2022-08-03       Impact factor: 3.473

3.  Deviation of the orientation angle of directional deep brain stimulation leads quantified by intraoperative stereotactic X-ray imaging.

Authors:  Josephiene M Schmidt; Lars Buentjen; Joern Kaufmann; Doreen Gruber; Harald Treuer; Aiden Haghikia; Jürgen Voges
Journal:  Neurosurg Rev       Date:  2022-05-12       Impact factor: 2.800

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.