Early Failure of a Carbon-Fiber-Reinforced Polyetheretherketone Distal Femur Plate: A Case Report.

W Alexander Goudriaan1, Rutger L Tordoir, Demien Broekhuis, Robert J P van der Wal.   

Abstract

CASE: In this case, we report on a carbon-fiber-reinforced polyetheretherketone plate failure 4 months after implantation, radiation therapy, chemotherapy, and protective weight-bearing in a 75-year-old woman who sustained a nontraumatic pathological distal femur fracture due to lymphoma.
CONCLUSION: Although carbon-fiber composite implants are regularly used and, to date, there have been no reports of early clinical failures, the longevity of the implant's structural integrity after high-dose radiation and/or chemotherapy treatment has not been fully explored. Therefore, we deem it too early to conclude that carbon-fiber-reinforced polyetheretherketone implants are superior to conventional implants in treating (pathological) fractures.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32910577     DOI: 10.2106/JBJS.CC.20.00041

Source DB:  PubMed          Journal:  JBJS Case Connect        ISSN: 2160-3251


  2 in total

Review 1.  Carbon-fiber-reinforced polyetheretherketone orthopedic implants in musculoskeletal and spinal tumors: imaging and clinical features.

Authors:  Jeremiah R Long; Maziyar A Kalani; Krista A Goulding; Jonathan B Ashman; Jonathan A Flug
Journal:  Skeletal Radiol       Date:  2022-05-10       Impact factor: 2.199

2.  Fracture fixation in extremity trauma with carbon fiber-reinforced polyetheretherketone (CFR-PEEK) plates: evidence today.

Authors:  George D Chloros; Apostolos D Prodromidis; Jo Wilson; Peter V Giannoudis
Journal:  Eur J Trauma Emerg Surg       Date:  2021-09-06       Impact factor: 2.374