Eugenia O'Kelly1, Anmol Arora2, Charlotte Pearson3, James R Ward1, P John Clarkson1.
Abstract
OBJECTIVE: Qualitative fit testing is a popular method of ensuring the fit of sealing face masks such as N95 and FFP3 masks. Increased demand due to the coronavirus disease 2019 (COVID-19) pandemic has led to shortages in testing equipment and has forced many institutions to abandon fit testing. Three key materials are required for qualitative fit testing: the test solution, nebulizer, and testing hood. Accessible alternatives to the testing solution have been studied. This exploratory qualitative study evaluates alternatives to the nebulizer and hoods for performing qualitative fit testing.Entities:
Keywords: behavioral risk factor surveillance; health communication; quality of health care; risk assessment; safety management
Mesh:
Year: 2020 PMID: 32907672 PMCID: PMC7642972 DOI: 10.1017/dmp.2020.352
Source DB: PubMed Journal: Disaster Med Public Health Prep ISSN: 1935-7893 Impact factor: 1.385
FIGURE 1A Chart of the Fit Test Results With Various Diffusers and Enclosures.
A checkmark indicates the correct response to the testing condition, while an x indicates an incorrect response to the test condition. Multiple members of the team tested the aroma diffuser, and their individual results are listed and factored into the accuracy ratings.
FIGURE 2The Procedure Followed to Perform Qualitative Fit Testing With the Aroma Diffuser.
In step 1, the essential equipment was gathered. The required 830 mg of sodium saccharin proved to fill slightly less than ¼ tsp. In step 2, 100 mL of distilled water was mixed with ¼ tsp of sodium saccharin. Half of this mixture was separated and a further 50 mL of water added to make up the sensitivity solution. In step 3, the participant donned the enclosure and breathed as instructed in qualitative fit testing instructions, with the mouth partly open and tongue placed toward the front of the mouth. If the sensitivity solution could be tasted, the sensitivity test was passed. In step 4, the participant donned the mask and performed each fit test procedure for 60 s, taking breaks in between tests for additional safety.