| Literature DB >> 32905265 |
Mohit Tayal1, Anirudh Mukherjee2, Udit Chauhan1, Madhur Uniyal3, Sakshi Garg4, Anjana Singh5, Ajeet Singh Bhadoria6, Ravi Kant7.
Abstract
CONTEXT: Vital parameters including blood oxygen level, respiratory rate, pulse rate, and body temperature are crucial for triaging patients to appropriate medical care. Advances in remote health monitoring system and wearable health devices have created a new horizon for delivery of efficient health care from a distance.Entities:
Keywords: Coronavirus disease; remote health monitoring; wearable devices
Year: 2020 PMID: 32905265 PMCID: PMC7467188 DOI: 10.4103/ijcm.IJCM_317_20
Source DB: PubMed Journal: Indian J Community Med ISSN: 0970-0218
Figure 1Evaluated system architecture: Design as submitted by Electronics Corporation of India Limited, Hyderabad
Intraclass correlation coefficient between the observations of two devices for all the four vital parameters (n=200)
| Vital parameter | Intraclass correlation | |
|---|---|---|
| SpO2 (%) | 0.92 | <0.01 |
| Temperature (°F) | 0.95 | <0.01 |
| Respiration rate per minute | 0.97 | <0.01 |
| Pulse rate per minute | 0.99 | <0.01 |
Bald-Altman comparison among the participants for all the four parameters
| Parameter | Device | Mean±SD | Bald-Altman comparison | |
|---|---|---|---|---|
| MD | 95% CI | |||
| SpO2 (%) | Philips Intellivue MX 500 | 96.96±2.09 | Reference | |
| ECIL device | 97.09±2.27 | −0.12 | −2.36-2.11 | |
| Temperature (°F) | Philips Intellivue MX 500 | 97.67±0.79 | Reference | |
| ECIL device | 97.42±0.80 | 0.25 | −0.25-0.75 | |
| Respiratory rate per minute | Philips Intellivue MX 500 | 18.44±4.05 | Reference | |
| ECIL device | 17.95±3.93 | 0.49 | −2.05-3.04 | |
| Pulse rate per minute | Philips Intellivue MX 500 | 80.87±10.7 | Reference | |
| ECIL device | 80.60±10.61 | 0.27 | −2.69-3.32 | |
ECIL: Electronics Corporation of India Limited, SD: Standard deviation, CI: Confidence interval, MD: Mean difference
Figure 2Bland–Altman comparison between two devices for SpO2(a), temperature (b), respiratory rate (c), and pulse rate (d). X-axis is the mean value for each parameter. Y-axis is the difference in the means from both reference and device under study. Blue line is the mean bias and red dashed lines show 95% limits of agreement