Measuring intradialyser transmembrane and hydrostatic pressures: pitfalls and relevance in haemodialysis and haemodiafiltration.

Alain Ficheux1,2, Nathalie Gayrard1,2, Ilan Szwarc3, Flore Duranton1,2, Fernando Vetromile3, Philippe Brunet4, Marie-Françoise Servel3, Joachim Jankowski5, Àngel Argilés1,2,3.   

Abstract

BACKGROUND: Post-dilutional haemodiafiltration (HDF) with high convection volumes (HCVs) could improve survival. HCV-HDF requires a significant pressure to be applied to the dialyser membrane. The aim of this study was to assess the pressure applied to the dialysers in HCV-HDF, evaluate the influence of transmembrane pressure (TMP) calculation methods on TMP values and check how they relate to the safety limits proposed by guidelines.
METHODS: Nine stable dialysis patients were treated with post-dilutional HCV-HDF with three different convection volumes [including haemodialysis (HD)]. The pressures at blood inlet (Bi), blood outlet (Bo) and dialysate outlet (Do) were continuously recorded. TMP was calculated using two pressures (TMP2: Bo, Do) or three pressures (TMP3: Bo, Do, Bi). Dialysis parameters were analysed at the start of the session and at the end of treatment or at the first occurrence of a manual intervention to decrease convection due to TMP alarms.
RESULTS: During HD sessions, TMP2 and TMP3 remained stable. During HCV-HDF, TMP2 remained stable while TMP3 clearly increased. For the same condition, TMP3 could be 3-fold greater than TMP2. This shows that the TMP limit of 300 mmHg as recommended by guidelines could have different effects according to the TMP calculation method. In HCV-HDF, the pressure at the Bi increased over time and exceeded the safety limits of 600 mmHg provided by the manufacturer, even when respecting TMP safety limits.
CONCLUSIONS: This study draws our attention to the dangers of using a two-pressure points TMP calculation, particularly when performing HCV-HDF.
© The Author(s) 2019. Published by Oxford University Press on behalf of ERA-EDTA.

Entities:  

Keywords:  convection volume; haemodiafiltration; haemodialysis; transmembrane pressure; ultrafiltration

Year:  2019        PMID: 32905251      PMCID: PMC7467581          DOI: 10.1093/ckj/sfz033

Source DB:  PubMed          Journal:  Clin Kidney J        ISSN: 2048-8505


INTRODUCTION

Haemodiafiltration (HDF) is an extracorporeal renal replacement technique that is based on both diffusion (passage of solutes across a permeable membrane according to a concentration gradient) and convection (passage of a fluid across a permeable membrane according to a pressure gradient). In HDF, convection is obtained by extracting fluid from the closed dialysate circuit at a given flow rate, which is then given back to the blood circuit either before the dialyser (pre-dilutional HDF) or after the dialyser (post-dilutional HDF; Figure 1). Recent evidence from randomized controlled trials suggests that post-dilutional HDF is associated with a benefit in survival, which would only be significant provided it is performed with high convection volumes (HCVs), that is, 17.4 L/session for the Turkish study [1-3]. Maduell et al.’s [4] latest study on incident patients confirms this improvement in survival. However, the most recent analysis of the Dialysis Outcomes and Practice Patterns Study (DOPPS) data [5] on a larger number of patients in several European centres does not confirm this benefit. The fact that dialysis conditions and equipment to obtain HCV-HDF vary across DOPPS facilities and are more homogeneous in the Maduell study may explain the discrepancy.
FIGURE 1

Schematic extracorporeal circuit displaying the measurement points and the resulting TMP calculations. The formulas for TMP2, TMP3 and TMP4 are displayed.

Schematic extracorporeal circuit displaying the measurement points and the resulting TMP calculations. The formulas for TMP2, TMP3 and TMP4 are displayed. To obtain HCV-HDF, significant pressure has to be applied to the dialyser membrane, frequently surpassing what is known as the GKD-UF max situation of the dialysis systems [6, 7]. GKD-UF stands for Global Coefficient of Ultra Filtration in the presence of Dialysate fluid. This parameter is not constant; it follows a parabolic function when convection is increased. The GKD-UF max is the vertex of this parabola, which corresponds to the situation where the maximum convection volume is obtained with the least applied pressure [transmembrane pressure (TMP)] [6, 7]. Then, when aiming to obtain as high as possible convection volumes [8], the dialysis systems are submitted to significantly stronger TMP constraints. The question arises to whether the resistance of the extracorporeal circuit to pressure is unlimited and what are the regulations warranting the security of the dialysis systems in terms of intra-dialyser pressure. Manufacturers specify the pressure limits of the dialysers in their notice and the European Renal Best Practice (ERBP) guidelines propose a TMP limited to 300 mmHg, as a safe maximum value [9]. However, TMP calculation is not standardized and commercially available monitors differ in their approaches [10]. Some dialysis monitors provide TMP calculated based on two pressure measurements (TMP2), whereas others use, or can use, a calculation based on three pressure measurements (TMP3). When assessing TMP2, pressures are measured at the blood outlet (Bo) and dialysate outlet (Do); when assessing TMP3, the measurement at the blood inlet (Bi) is also considered [10]. In presently available guidelines, there is no mention of how TMP should be measured and whether the value retained as a limitation applies to the mean observed pressure in the dialyser or the maximum pressure at any point of it [9]. Data on the influence of the two- or three-point (taking into account the pressure at the Bi) measurements on TMP are lacking and the actual values of pressure in the different points of the dialyser are not usually considered by clinicians when performing dialysis treatment. The present work aimed to evaluate the influence of the calculation of TMP (2 and 3 point) on its absolute value in the clinical situation of post-dilutional HDF with different convection volumes. This study focuses on the intra-dialyser pressures in real haemodialysis (HD) and online post-dilution HDF treatments. It analyses the data obtained from a larger study for which technical and clinical aspects have been previously published [11, 12]. In these reports, the sensitivity and reliability of determining GKD-UF max were demonstrated to be in the range of clinical use [11], and its application in clinics showed to be beneficial as it rendered the extracorporeal circuit more stable in terms of pressure constraints and decreased the number of alarms during treatments while allowing to reach sensibly HCVs [12].

MATERIALS AND METHODS

Preliminary study

By definition, TMP measurement requires the determination of four parameters: pressure at the blood line inlet (PBi), at the blood line outlet (PBo), at the dialysate outlet (PDo) and at the dialysate inlet (PDi) of the dialyser. In the clinical setting, PDi is usually unavailable, but some dialysis generators include this parameter such as the DBB-05 generators (Nikkiso, Tokyo, Japan). We analysed data from four patients dialysed on Nikkiso DBB-05 generators with 1.8 m2 HF80 dialysers (Fresenius Medical Care, Bad Homburg, Germany), in post-dilutional HDF. We applied step increases in convection flow [cumulative 10 mL/min increases in the flow rate of the infusion pump (Figure 1)] and recorded the four pressures at the dialysate and Bi and outlet displayed on the generator when reaching the programmed convection at every step. We calculated and compared the resulting TMP2, TMP3 and TMP4 (formulas in Figure 1).

Clinical study

Patients

The population under study is described in Table 1. Nine dialysis patients stable at the moment of the study were assessed. There were four men and five women and mean age was 73 ± 4 years. Their dialysis accesses were arterio-venous fistulae (n = 8) and tunnelled jugular catheter (n = 1). All patients gave their written informed consent. The study protocol was approved by the ‘Comité de Protection des Personnes’ of Nîmes (2011.10.05 bis) with the registration number at the French Agency AFSSAPS 2011-A01092-39. The study was performed in agreement with the declaration of Helsinki.
Table 1.

Patients characteristics

CharacteristicsPatients (n = 9)
Gender5F/4M
Age (years)73 ± 4
Body weight after dialysis (kg)68 ± 5
Haematocrit (%)35.7 ± 1.5
Serum proteins (g/L)62.6 ± 1.5
Serum albumin (g/L)36.0 ± 1.2
Patients characteristics

Dialysis modality

Patients were treated with post-dilutional HDF with 1.8 m2 Amembris membrane dialysers (B BRAUN Avitum, Melsungen, Germany) with three different levels of convection, each applied during three consecutive treatments (total 81 sessions). The convection volumes were: (i) HD (convection volume limited to the weight loss); (ii) post-dilutional HDF with the convection flow corresponding to the maximum of the ultrafiltration flow-to-TMP ratio (HCV-HDF at GKD-UF max) [11]; and (iii) post-dilutional HDF with the highest acceptable convection flow (HCV-HDF at QUF max). Safety alarms were set according to the recommendations of the ERBP guidelines: ultrafiltration rate limited to 30% of the blood flow and TMP3 was limited to 300 mmHg [9]. No automatic system adapting convection was utilized. The nursing staff had the prerogative of modifying the infusion flow if TMP alarms prevented from pursuing the dialysis at the prescribed convection volumes. Blood flow, convection volumes and dialysis times are summarized in Table 2.
Table 2.

Observed characteristics of different dialysis conditions

Dialysis conditionHDQUF at GKD-UF max Q UF maxP-value
Number of sessions262726
Session duration (min)234 ± 3237 ± 3236 ± 40.51
Blood flow (mL/min)362 ± 7360 ± 7365 ± 60.50
Dialysate flow (mL/min)601 ± 2601 ± 1601 ± 10.38
Ultrafiltration volume for weight loss (L)2.6 ± 0.22.8 ±0.22.9 ± 0.20.74
Ultrafiltration volume for infusion (L)017.8 ± 0.322.2 ± 0.5<0.001*
Total ultrafiltration volume (L)2.6 ± 0.220.6 ± 0.425.1 ± 1.0<0.001*
Total ultrafiltration flow over blood flow (%)3.0 ± 0.224.0 ± 0.328.7 ± 0.3<0.001*

All adjusted P-values for pairwise comparisons <0.001.

Observed characteristics of different dialysis conditions All adjusted P-values for pairwise comparisons <0.001.

Variables measured and TMP calculations

HDF machines with three pressure sensors (Dialog+, B BRAUN Avitum, Melsungen, Germany) were used. The pressures PBi, PBo and PDo of the dialyser (Figure 1) were continuously recorded by the dialysis monitor. Two time points of the dialysis session were specifically studied: (i) at the beginning of the treatment (Tstart) and (ii) at the end of dialysis session or when TMP alarms induced a manual intervention to decrease convection (Tend or modif). The average of 10 consecutive readings per patient and time point was considered. TMP was calculated using two pressure points (TMP2) and three pressure points (TMP3) according to formulas displayed in Figure 1.

Statistical analysis

Differences between dialysis conditions were analysed by variance analysis (ANOVA) accounting for the cross-over design. Post hoc tests were performed, adjusting for multiple comparisons using the Bonferroni correction. Statistical analyses were performed using SAS version 9.4 (SAS Institute, Cary, NC, USA). Tests were performed using a Type I error of 5%. Results are given as the mean ± standard error of the mean.

RESULTS

Influence of TMP calculation and convection flow on TMP value

In a preliminary study, the influence of using 2, 3 and 4 point measurements on TMP value was assessed at different levels of convection. Figure 2 shows an illustrative example of the pressure values and resulting TMP obtained, whereas the dialysis system was submitted to a full range of convection. These results, presented for one case, were reproduced in all participating patients. In Figure 2A, it can be observed that PBi increases exponentially with convection flow, while PBo remained constant. The dialysate pressures PDi and PDo decrease exponentially with the increase in QUF, in a near parallel way mirroring changes in PBi.
FIGURE 2

Pressure readings over total ultrafiltration flow variation in a dialysis system equipped with four pressure sensors. (A) The recorded values at the dialyser inlet and outlet at the blood compartment lines (PBi and PBo, respectively) and dialysate compartment side (PDi and PDo, respectively). (B) TMP calculated with TMP2, TMP3 and TMP4. (C) Difference between TMP estimations showing the influence of successively adding a third (TMP3 – TMP2, filled diamonds) and fourth (TMP4 – TMP3, open diamonds) measuring point.

Pressure readings over total ultrafiltration flow variation in a dialysis system equipped with four pressure sensors. (A) The recorded values at the dialyser inlet and outlet at the blood compartment lines (PBi and PBo, respectively) and dialysate compartment side (PDi and PDo, respectively). (B) TMP calculated with TMP2, TMP3 and TMP4. (C) Difference between TMP estimations showing the influence of successively adding a third (TMP3 – TMP2, filled diamonds) and fourth (TMP4 – TMP3, open diamonds) measuring point. The TMP values calculated from the 2, 3 or 4 pressure measurements for the same treatment are presented in Figure 2B. It can be observed that TMP2 underestimates TMP3 of 45–140 mmHg. The difference between TMP3 and TMP2 increases with increasing convection flow due to the exponential increase in PBi, while PBo remains constant (Figure 2A). The patterns of TMP3 and TMP4 are similar and absolute values differ by around 50 mmHg across the QUF range (Figure 2C). This value, according to the TMP formulas, is the half of the pressure drop in the dialysate [(PDi − PDo)/2]. It is around 100 mmHg (Figure 2A), well above the in vitro pressure drop measured at sites strictly corresponding to the dialyser length (25 ± 3 mmHg, in agreement with the values given by the manufacturer). The discrepancy can be explained by the position of the PDi sensor in the Nikkiso DBB-05 machine, away from the dialyser inlet. In this setting, the difference between TMP3 and TMP4 is about 12.5 mmHg (half of 25 mmHg) when dialysate pressure sensors are adjacent to the dialyser. It can be concluded that the fourth pressure measurement (PDi) has a minimal influence on TMP estimation.

Intradialyser transmembrane and hydrostatic pressures during treatment

In the clinical study, we collected pressure records every 30 s from three-pressure sensors generators to assess the clinical relevance of different TMP estimations for dialysis session monitoring. Figure 3 shows the pressure recordings of three treatments representative of HD, HCV-HDF at GKD-UF max and HCV-HDF at maximum convection flow in the same patient. It can be observed that PBi (red curve, Figure 3) is the parameter submitted to more variations during the treatment procedure. It is quite stable in dialysis, increases in HCV-HDF at GKD-UF max and increases even more with HCV-HDF at QUF max, surpassing 600 mmHg (limit given in the manufacturer information for use of dialysers used in this study). These modifications in PBi are reflected by parallel changes in TMP3 (green curve, Figure 3). In HCV-HDF at QUF max, TMP3 reached the limit level of 300 mmHg, forcing decrease of the infusion rate, towards the last third of the treatment time (Tmodif). In contrast, TMP2 did not show the variations occurring in the dialyser and remained stable, well below 300 mmHg. This reveals that safety limits do not apply in a similar manner to a system calculating the TMP2 or TMP3 and it does not render secure the system when applied to dialysis monitors calculating TMP from two pressure measurements (PBo and PDo).
FIGURE 3

Pressure readings during dialysis treatments with different convection flows. Illustrative examples of three treatments performed in HD (left panels), high convection HDF at GKD-UF max (central panels) or at maximum possible convection (QUF max, right panels) in the same patient. The continuous recordings of pressures measured at the Bi and Bo (top panels) and TMP calculated based upon 2 or 3 measurement points (TMP2 and TMP3, lower panels) are displayed. Asterisks indicate infusion modification due to TMP alarms.

Pressure readings during dialysis treatments with different convection flows. Illustrative examples of three treatments performed in HD (left panels), high convection HDF at GKD-UF max (central panels) or at maximum possible convection (QUF max, right panels) in the same patient. The continuous recordings of pressures measured at the Bi and Bo (top panels) and TMP calculated based upon 2 or 3 measurement points (TMP2 and TMP3, lower panels) are displayed. Asterisks indicate infusion modification due to TMP alarms. In Figure 4, the pressure measured at the Bi and Bo at the beginning of the session (Tstart), and either at the moment that the infusion flow was reduced or at the end (Tmodif or end) for the three treatment conditions (upper panels). In both HCV-HDF treatments, PBo remained stable over the session. In HD, PBi was stable during the session. In HDF-HCV with convection flow at GKD-UF max, PBi increased during the session, and in HCV-HDF with maximal convection flow, PBi further increased during treatment. The distributions of TMP2 and TMP3 values at the beginning and at end of the session are plotted for the nine patients. In HD treatments, the TMP calculated from two or three sensors remained relatively stable throughout the treatment (Figure 4, left panel). In the case of HCV-HDF at GKD-UF max, TMP2 and TMP3 were greater and increased during session (Figure 4, central panel). On average, TMP3 was 3.7 ± 0.2 times the value of TMP2 for the same session. There were three reductions in ultrafiltration rates due to TMP alarms (TMP3 > 300 mmHg) over 27 sessions. In HCV-HDF with QUF max, TMP2 and TMP3 increased further during the sessions (Figure 4, right panel). The mean difference between TMP3 and TMP2 was 148 ± 5 mmHg (range 74–247), and TMP3 was on average 2.8 ± 0.1 times the value of TMP2. There were ultrafiltration reductions due to TMP3 reaching the alarm threshold of 300 mmHg in 21 out of 26 sessions.
FIGURE 4

Pressure readings at two time points of dialysis treatments performed with different convection flows. Distribution of blood line pressures (top panels) and TMP (lower panels) observed in 79 treatments from 9 patients in HD (left panels), high convection HDF at GKD-UF max (central panels) or at maximum possible convection (QUF max, right panels). Values were obtained at the start of the treatment (Tstart) and at the end of the treatment (Tend) or at the moment of adapting the convection because of TMP alarms (Tmodif or end), whatever was first. Box plots display median, first and third quartile, Tukey’s fences and extreme values.

Pressure readings at two time points of dialysis treatments performed with different convection flows. Distribution of blood line pressures (top panels) and TMP (lower panels) observed in 79 treatments from 9 patients in HD (left panels), high convection HDF at GKD-UF max (central panels) or at maximum possible convection (QUF max, right panels). Values were obtained at the start of the treatment (Tstart) and at the end of the treatment (Tend) or at the moment of adapting the convection because of TMP alarms (Tmodif or end), whatever was first. Box plots display median, first and third quartile, Tukey’s fences and extreme values.

Pitfalls in TMP calculation with two sensors

Figure 5 presents two examples where TMP2 overlooks a problem occurring at the blood side during the dialysis procedure and shows a decrease in TMP2, while TMP3 increased. In the left panel, Figure 5 displays the recordings of an HD treatment. A marked increase in pressure at the Bi was observed for an unexpected reason (very likely clotting of the fibres) after ∼150 min from session initiation. The increase in hydrostatic pressure at the blood side translates into an increase in TMP3 (green curve), whereas TMP2 displayed an artefactual decrease.
FIGURE 5

Examples of pressure variations observed during two treatments in the same patient. During an HD treatment (left panel), an unexpected pressure rise at the Bi (PBi) was observed after 150 min. TMP3 increased consequently, while TMP2 tended to slowly decrease regardless pressure variations in the blood compartment. In a high convection HDF performed at the maximum convection flow (QUF max, right panel), TMP2 and TMP3 initially followed the changes in PBi until TMP3 reached 300 mmHg and infusion had to be turned down (asterisks, infusion modification). After this point, only TMP3 followed PBi variations, while TMP2 remained constant, regardless of the variations observed in the blood compartment.

Examples of pressure variations observed during two treatments in the same patient. During an HD treatment (left panel), an unexpected pressure rise at the Bi (PBi) was observed after 150 min. TMP3 increased consequently, while TMP2 tended to slowly decrease regardless pressure variations in the blood compartment. In a high convection HDF performed at the maximum convection flow (QUF max, right panel), TMP2 and TMP3 initially followed the changes in PBi until TMP3 reached 300 mmHg and infusion had to be turned down (asterisks, infusion modification). After this point, only TMP3 followed PBi variations, while TMP2 remained constant, regardless of the variations observed in the blood compartment. In the right panel, Figure 5 presents an example of HCV-HDF in which convection had to be reduced during the treatment procedure due to TMP3 reaching the safety limit 300 mmHg. A further increase in TMP3 was observed after resuming with a lower convection flow and was associated with a marked increase in pressure measured at the Bi, reaching values over the advised limit of use given by the manufacturer of 600 mmHg. TMP2 values (yellow curve) were much lower than TMP3 values and TMP2 variations did not reflect the changes in pressure occurring in the dialyser. The pressure changes that should trigger nurse interventions were overlooked when monitoring TMP2.

DISCUSSION

TMP was the main parameter that allowed the pioneers in dialysis to control ultrafiltration and thereby weight loss and extracellular volume. With the advent of ultrafiltration controllers [13], weight loss was accurately driven by automatic setting of TMP by dialysis monitors. TMP turned into an informative value and lost its protagonism. Subsequently, with the introduction and spread of high permeability dialysers and of the use of convective techniques, TMP regained interest. Among other things, properly assessing TMP has allowed identification of the fact that the global ultrafiltration coefficient of a dialysis system is not constant, but follows a parabolic function when convection increases [6]. It is known that TMP varies greatly along the fibres of the dialyser [14]. However, in practice, it is not possible to measure the pressure of the fibres within the dialyser and external pressure sensors located outside the dialyser are used instead. To estimate an average TMP value, dialysis machines are equipped with two, three or four pressure sensors measuring pressures at the inlet and outlet of the dialyser, at both the blood and dialysate sides. The present work was performed to evaluate the influence of TMP calculations using two and three pressure measurements on its absolute value and on its variations during sessions in the clinical situation of post-dilutional HCV-HDF. Almost no information is available on these aspects, and the rare criteria included in dialysis guidelines are not adequately described or sustained by solid data. Our study provides information on the direct effects of convection on the pressures at the dialysate and blood sides, on the resulting TMP value. It also provides a description of longitudinal changes in pressure observed throughout the treatment, and how they are induced by convection. When performing HD without added convection, the pressures of the system remain particularly stable. When convection is added, the pressures change during the procedure and the greatest changes are observed at the blood side at the entrance of the dialyser: pressure increases in a proportional manner to the rate of convection QUF. Since the pressure at the outlet at the blood side of the dialyser remains constant during dialysis, we observe an increase in the pressure drop in the blood compartment of the dialyser. Therefore, any method to assess the pressures in the dialysis system performing HCV-HDF that does not take into account the pressure values at the Bi of the dialyser is not accurate and does not reflect the phenomena occurring inside the dialyser. This is the case when calculating TMP from only two points (PBo and PDo). Since the blood pressure variations occurring during dialysis are overlooked, TMP2 should not be used to monitor a dialysis system performing HCV-HDF. Only machines performing at least a three-point TMP measurement including the Bi site should be used to monitor the dialysis systems, particularly in HCV-HDF. Is it necessary to monitor the four points? Our preliminary study shows that TMP3 and TMP4 follow the same pattern and that the difference between the two is relatively constant. The value of this difference is the result of adding the readings from the dialysate inlet into the formula (PDi will be averaged with the PDo value). In our study, this value was further increased by the location of the sensor on the dialysis machine, situated at a distance of the entrance of the dialyser. The pressure drop in the dialysate compartment is stable during the dialysis treatment, relatively independent of the convection rate and available from manufacturer documentations. In consequence, measuring the pressure at the dialysate inlet of the dialyser is not mandatory. However, should the reading of the pressure at the dialysate inlet be incorporated in the formula, it has to be corrected by the pressure drop before reaching the dialyser. Otherwise, using the crude value given by the sensor would artefactually decrease the estimated TMP4. The relevance of our observations is supported by the tendency to advise and spread the use of HCV-HDF based upon most of the recent studies in dialysis patients [1-4], whereas the knowledge on the behaviour of the filtrating system under HCV-HDF is rather scant. Our data show precisely that in HCV-HDF, considerable modifications in TMP occur, and more particularly when convection rates overpass GKD-UF max. Secondly, we aim to increase safety and want to avoid problems occurring during treatment. Focussing on the blood side of the extracorporeal circuit, the pressure measured in the blood outlet reflects the return or venous pressure. Therefore, any sudden changes in PBo require immediate attention to the needle and blood access. The pressure at the Bi (PBi) reflects changes on the resistance to flow between the blood pump and the entrance of the dialyser. Therefore, with constant blood and convection flows, a sudden increase in PBi may be due to a decrease in effective dialysate surface or an increase in blood viscosity, both markers of initiated or premonitory of clotting in the dialyser fibres, respectively [15]. As we have shown in Figure 5, left panel, this complication may be prevented if monitored TMP is calculated using three or four pressure points, as they include the pressure measured at the PBi. TMP3 and TMP4 will warn about a sudden increase in pressure at the Bi and preventive measures may be implemented, whereas TMP2 will not. Some machines presently available still use a TMP2 monitoring though. Furthermore, with the advent of the HCV-HDF, the knowledge on the pressure changes in the system will help in establishing safety limits for this type of treatment. The ERBP has emitted an advice limiting the TMP to 300 mmHg [9]. However, our data clearly show that any given TMP limit should be associated to a type of TMP calculation. We have seen that 300 mmHg of TMP3 is already associated with increases in pressure at the Bi that may be beyond those advised by the manufacturer and that the use of TMP2 overlooks them. We think it may also be appropriate to monitor pressure at the Bi and propose to set a limit to be the dialyser manufacturer limit. In summary, our study provides new information on the behaviour of dialysis systems, particularly relevant when performing HCV-HDF, which may help to establish pressure safety. We show that a strong discrepancy exists across TMP calculations, whereas appropriately monitoring pressures should be sought.
  12 in total

1.  The ultrafiltration coefficient of a dialyser (KUF) is not a fixed value, and it follows a parabolic function: the new concept of KUF max.

Authors:  Alain Ficheux; Peter G Kerr; Philippe Brunet; Àngel Argilés
Journal:  Nephrol Dial Transplant       Date:  2010-09-08       Impact factor: 5.992

2.  A new scintigraphic method to characterize ultrafiltration in hollow fiber dialyzers.

Authors:  C Ronco; A Brendolan; M Feriani; M Milan; P Conz; A Lupi; P Berto; M Bettini; G La Greca
Journal:  Kidney Int       Date:  1992-05       Impact factor: 10.612

3.  EBPG guideline on dialysis strategies.

Authors:  James Tattersall; Alejandro Martin-Malo; Luciano Pedrini; Ali Basci; Bernard Canaud; Denis Fouque; Patrick Haage; Klaus Konner; Jeroen Kooman; Francesco Pizzarelli; Jan Tordoir; Marianne Vennegoor; Christoph Wanner; Piet ter Wee; Raymond Vanholder
Journal:  Nephrol Dial Transplant       Date:  2007-05       Impact factor: 5.992

4.  Mortality and cardiovascular events in online haemodiafiltration (OL-HDF) compared with high-flux dialysis: results from the Turkish OL-HDF Study.

Authors:  Ercan Ok; Gulay Asci; Huseyin Toz; Ebru Sevinc Ok; Fatih Kircelli; Mumtaz Yilmaz; Ender Hur; Meltem Sezis Demirci; Cenk Demirci; Soner Duman; Ali Basci; Siddig Momin Adam; Ismet Onder Isik; Murat Zengin; Gultekin Suleymanlar; Mehmet Emin Yilmaz; Mehmet Ozkahya
Journal:  Nephrol Dial Transplant       Date:  2012-12-09       Impact factor: 5.992

5.  Hemodiafiltration Reduces All-Cause and Cardiovascular Mortality in Incident Hemodialysis Patients: A Propensity-Matched Cohort Study.

Authors:  Francisco Maduell; Javier Varas; Rosa Ramos; Alejandro Martin-Malo; Rafael Pérez-Garcia; Isabel Berdud; Francesc Moreso; Bernard Canaud; Stefano Stuard; Adelheid Gauly; Pedro Aljama; Jose Ignacio Merello
Journal:  Am J Nephrol       Date:  2017-10-17       Impact factor: 3.754

6.  Patient- and treatment-related determinants of convective volume in post-dilution haemodiafiltration in clinical practice.

Authors:  E Lars Penne; Neelke C van der Weerd; Michiel L Bots; Marinus A van den Dorpel; Muriel P C Grooteman; Renée Lévesque; Menso J Nubé; Piet M Ter Wee; Peter J Blankestijn
Journal:  Nephrol Dial Transplant       Date:  2009-06-10       Impact factor: 5.992

7.  Consequences of increasing convection onto patient care and protein removal in hemodialysis.

Authors:  Nathalie Gayrard; Alain Ficheux; Flore Duranton; Caroline Guzman; Ilan Szwarc; Fernando Vetromile; Chantal Cazevieille; Philippe Brunet; Marie-Françoise Servel; Àngel Argilés; Moglie Le Quintrec
Journal:  PLoS One       Date:  2017-02-06       Impact factor: 3.240

8.  Mortality risk in patients on hemodiafiltration versus hemodialysis: a 'real-world' comparison from the DOPPS.

Authors:  Francesco Locatelli; Angelo Karaboyas; Ronald L Pisoni; Bruce M Robinson; Joan Fort; Raymond Vanholder; Hugh C Rayner; Werner Kleophas; Stefan H Jacobson; Christian Combe; Friedrich K Port; Francesca Tentori
Journal:  Nephrol Dial Transplant       Date:  2018-04-01       Impact factor: 5.992

9.  A reliable method to assess the water permeability of a dialysis system: the global ultrafiltration coefficient.

Authors:  A Ficheux; N Gayrard; F Duranton; C Guzman; I Szwarc; F Vetromile; P Brunet; M F Servel; A Argilés
Journal:  Nephrol Dial Transplant       Date:  2017-02-01       Impact factor: 5.992

Review 10.  The ultrafiltration coefficient: this old 'grand inconnu' in dialysis.

Authors:  Alain Ficheux; Claudio Ronco; Philippe Brunet; Àngel Argilés
Journal:  Nephrol Dial Transplant       Date:  2013-12-19       Impact factor: 5.992

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.