Optic Nerve Sheath Diameter Measured by Point-of-Care Ultrasound and MRI.

Hamid Shokoohi1, Matthew Pyle2, Evan Kuhl2, Michael A Loesche3, Aakshit Goyal4, Maxine A LeSaux2, Keith S Boniface2, M Reza Taheri4.   

Abstract

BACKGROUND AND
PURPOSE: Ultrasound (US) measurement of the optic nerve sheath diameter (ONSD) and optic nerve diameter (OND) is a method frequently used to screen for an increased intracranial pressure. The aim of this study was to assess the accuracy of US measurements of ONSD and OND, when compared to magnetic resonance imaging (MRI) measurements as the criterion standard.
METHODS: In this prospective, single-institution study, orbital US was performed for those patients requiring an emergent brain MRI. ONSD and OND of both eyes were measured in the axial and coronal planes in straight gaze by US. ONSD and OND from brain and orbital MRI were measured by two neuroradiologists. Correlation and agreement between readings were assessed using Pearson's correlations.
RESULTS: Eighty-two patients met inclusion criteria. The mean axial and coronal ONSD in the MRI examinations was 5.6 and 5.7 mm at 3-5.9 mm behind the globe, respectively. The mean ONSD from the US measurements was 6.22 and 5.52 mm in the axial and coronal planes, respectively. The mean OND in US examinations was 4.31 mm (axial) and 3.68 mm (coronal). Axial versus coronal measurements of ONSD had a modest correlation in US assessment with an r2 of .385 (P < .001) but there were no correlations between any of the US and MRI measurements.
CONCLUSIONS: In measuring ONSD and OND, US measurements showed a modest correlation between axial and coronal measurements, but no concordance was found between US and MRI in our setting.
© 2020 American Society of Neuroimaging.

Entities:  

Keywords:  Increased ICP; MRI; optic nerve sheath; point-of-care ultrasound

Mesh:

Year:  2020        PMID: 32896994     DOI: 10.1111/jon.12764

Source DB:  PubMed          Journal:  J Neuroimaging        ISSN: 1051-2284            Impact factor:   2.486


  2 in total

Review 1.  Point-of-Care Ultrasound for the Trauma Anesthesiologist.

Authors:  Eric R Heinz; Anita Vincent
Journal:  Curr Anesthesiol Rep       Date:  2022-01-20

Review 2.  A Review of the Methods of Non-Invasive Assessment of Intracranial Pressure through Ocular Measurement.

Authors:  Jinhui Dong; Qi Li; Xiaofei Wang; Yubo Fan
Journal:  Bioengineering (Basel)       Date:  2022-07-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.