Literature DB >> 32896924

High efficiency and workflow of His bundle pacing and atrioventricular node ablation guided by three-dimensional mapping system.

Yuanjun Sun1, Xiaohong Yu1, Xianjie Xiao1, Xiaomeng Yin1, Lianjun Gao1, Rongfeng Zhang1, Shiyu Dai1, Nan Wang1, Duoduo Zhang1, Yingxue Dong1, Yanzong Yang1, Yunlong Xia1.   

Abstract

BACKGROUND: Atrioventricular node (AVN) ablation combined with His bundle pacing is an effective strategy for permanent atrial fibrillation (AF) with rapid ventricular rate refractory to pharmacological therapy. We aimed to access the feasibility and efficiency of His bundle pacing and AVN ablation guided by three-dimensional (3-D) mapping system throughout the procedure.
METHODS: Eighteen patients with permanent AF with refractory rate and symptoms were referred for His bundle pacing and AVN ablation guided by 3-D mapping (CARTO3). Electroanatomic 3-D mapping of the right atrium and right ventricle was performed by the ablation catheter with CARTO 3 system, followed by the visualization of the leads for implantation and AVN ablation.
RESULTS: Implantation of His bundle and ventricular leads and AVN ablation were achieved successfully with the help of 3-D mapping in 17 patients. Selective His bundle pacing was achieved in five patients (29.4%), and the other (70.6%) were nonselective His bundle pacing. The mean procedure duration was 99.4 ± 16.4 minutes. The mean fluoroscopy time was 7.0 ± 2.6 minutes. The time spent on His lead implantation was 6.1 ± 3.2 minutes. One patient experienced AVN ablation from left side under aortic valves due to no effect of ablation in right atrium.
CONCLUSION: His bundle pacing and AVN ablation guided by throughout real-time 3-D mapping system are of high-efficiency and feasibility.
© 2020 Wiley Periodicals LLC.

Entities:  

Keywords:  His bundle pacing; atrial fibrillation; atrioventricular node ablation; three-dimensional mapping

Year:  2020        PMID: 32896924     DOI: 10.1111/pace.14061

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

Review 1.  Three-dimensional mapping in cardiac implantable electronic device - a feasible and effective alternative to fluoroscopy.

Authors:  Xiaozhen Ge; Mengna Chen; Zhongxin Sha; Junmeng Zhang
Journal:  J Interv Card Electrophysiol       Date:  2022-07-09       Impact factor: 1.900

2.  Zero fluoroscopy atrioventricular node ablation and left bundle branch pacing guided by electroanatomic tridimensional mapping system.

Authors:  Sebastián Giacoman; María Algarra; Ana Delia Ruiz; José Miguel Lozano
Journal:  J Cardiol Cases       Date:  2021-07-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.