Literature DB >> 3289520

Intractable hemolysis caused by perivalvular leakage following mitral valve replacement with St. Jude Medical prosthesis.

Y Okita1, S Miki, K Kusuhara, Y Ueda, T Tahata, Y Tsukamoto, K Yamanaka, S Shiraishi.   

Abstract

Nine patients with intractable hemolysis caused by perivalvular leakage following mitral valve replacement with a St. Jude Medical prosthesis are presented. All patients had dark-colored hemoglobinuria, which appeared from 1 day to 44 days after the operation, with moderate or severe hepatorenal insufficiency. One patient died of multiorgan failure. The other 8 patients underwent reoperation, and all survived. Reoperation revealed that all leakages were tiny and had no adverse effect on hemodynamics. In all the patients having reoperation, hemoglobinuria disappeared immediately after the procedure. Surface-scanning electron microscopy of the sewing cuff of the St. Jude Medical prosthesis revealed the rough surface structure of the sewing ring. Because of this irregular, shaggy surface structure, greater shearing forces against erythrocytes can be generated when there is perivalvular leakage.

Entities:  

Mesh:

Year:  1988        PMID: 3289520     DOI: 10.1016/s0003-4975(10)65860-2

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Mid-term surgical results after valve replacement with the CarboMedics valve prosthesis.

Authors:  T Abe; K Morishita; M Tsukamoto; T Tanaka; S Komatsu
Journal:  Surg Today       Date:  1995       Impact factor: 2.549

2.  Intractable mechanical hemolytic anemia complicating mitral valve surgery: a case series study.

Authors:  Jin Wang; Hanlin Zhang; Hongyang Fan; Kang Chen; Yuelun Zhang; Kaicheng Song; Hushan Ao; Chunhua Yu
Journal:  BMC Cardiovasc Disord       Date:  2020-03-03       Impact factor: 2.298

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.