| Literature DB >> 32887546 |
Sirui Liu1, Bo Hou1, Yiwei Zhang1, Tianye Lin1, Xiaoyuan Fan1, Hui You1, Feng Feng2.
Abstract
BACKGROUND: The inter-scanner reproducibility of brain volumetry is important in multi-site neuroimaging studies, where the reliability of automated brain segmentation (ABS) tools plays an important role. This study aimed to evaluate the influence of ABS tools on the consistency and reproducibility of the quantified brain volumetry from different scanners.Entities:
Keywords: Automated brain volumetry; Coefficient of variation; Inter-scanner reproducibility; Magnetic resonance imaging
Year: 2020 PMID: 32887546 PMCID: PMC7472704 DOI: 10.1186/s12868-020-00585-1
Source DB: PubMed Journal: BMC Neurosci ISSN: 1471-2202 Impact factor: 3.288
Imaging protocols of the tested MRI scanners
| Imaging protocols | GE Discovery MR750 | Siemens Skyra | Philips Ingenia CX |
|---|---|---|---|
| Field strength (T) | 3.0 | 3.0 | 3.0 |
| Sequence name | BRAVO | tfl3d1 | T1W_3D_TFE |
| Sequence type | T1 inversion-prepared FSGRE | MPRAGE | 3D TFE |
| TR (ms) | 6.7 | 1900 | 8.2 |
| TE (ms) | 2.9 | 2.5 | 3.7 |
| TI (ms) | 400.0 | 900.0 | 964.9 |
| FA (°) | 12 | 9 | 8 |
| Matrix | 256 × 256 | 256 × 256 | 256 × 256 |
| FOV (mm) | 256 × 256 | 256 × 256 | 256 × 256 |
| Slice thickness(mm) | 1 | 1 | 1 |
| Angulation | Sagittal | Sagittal | Sagittal |
| Voxel size (mm) | 1 × 1 | 1 × 1 | 1 × 1 |
| Number of slices | 190 | 192 | 190 |
| TA (min:s) | 05:51 | 04:31 | 05:34 |
FA flip angle, FOV field of view, TA acquisition time, TE echo time, TI inversion time, TR repetition time
Fig. 1Analysis method of coefficient of variation of left hippocampus. CV: coefficient of variance; SD: standard deviation of volumetric results of left hippocampus from three scanners; Mean: arithmetic mean of volumetric results of left hippocampus from three scanners
Fig. 2FreeSurfer segmentation results of different MRI acquisitions
Fig. 3FSL-FIRST segmentation results of different MRI acquisitions
Fig. 4AccuBrain® segmentation results of different MRI acquisitions
Coefficient of variation (CV) for inter-scanner volumetric measurements among GE, Siemens and Philips
| Coefficient of variation | p-value | |||||
|---|---|---|---|---|---|---|
| FreeSurfer | FSL-FIRST | AccuBrain | FreeSurfer vs. FSL-FIRST | AccuBrain vs. FreeSurfer | AccuBrain vs. FSL-FIRST | |
| WM | 8.117 | N.A. | 1.348 | N.A. | 0.005 | N.A. |
| GM | 6.946 | N.A. | 1.927 | N.A. | < 0.001 | N.A. |
| Hippocampus L | 7.706 | 9.600 | 8.800 | 0.151 | 0.978 | 0.679 |
| Hippocampus R | 8.459 | 10.57 | 8.144 | 0.489 | 0.561 | 0.064 |
| Amygdala L | 10.35 | 17.15 | 3.768 | 0.010 | < 0.001 | < 0.001 |
| Amygdala R | 10.23 | 20.98 | 3.557 | 0.004 | 0.002 | < 0.001 |
| Lateral ventricle L | 9.123 | N.A. | 2.547 | N.A. | 0.001 | N.A. |
| Lateral ventricle R | 8.631 | N.A. | 2.892 | N.A. | 0.008 | N.A. |
| VentralDC L | 9.809 | N.A. | 4.298 | N.A. | 0.018 | N.A. |
| VentralDC R | 9.350 | N.A. | 3.685 | N.A. | 0.022 | N.A. |
| Thalamus L | 8.632 | 8.919 | 1.635 | 0.679 | < 0.001 | < 0.001 |
| Thalamus R | 8.440 | 9.739 | 1.811 | 0.524 | < 0.001 | < 0.001 |
| Caudate L | 9.597 | 12.91 | 2.750 | 0.083 | < 0.001 | < 0.001 |
| Caudate R | 9.331 | 14.11 | 2.314 | 0.073 | 0.001 | < 0.001 |
| Putamen L | 10.77 | 11.61 | 2.755 | 0.720 | < 0.001 | < 0.001 |
| Putamen R | 9.793 | 11.64 | 2.471 | 0.330 | 0.001 | < 0.001 |
| Pallidum L | 12.29 | 7.245 | 4.025 | 0.018 | < 0.001 | 0.007 |
| Pallidum R | 11.77 | 9.306 | 2.316 | 0.389 | 0.001 | < 0.001 |
| Accumbens L | 11.72 | 13.93 | 4.859 | 0.679 | < 0.001 | 0.001 |
| Accumbens R | 10.43 | 18.75 | 5.014 | 0.015 | 0.002 | < 0.001 |
| Average | 9.577 | 12.60a | 3.546 | 0.041 | 0.001 | < 0.001 |
The mean values of region-specific CV (i.e. inter-scanner variability in brain volumetry of the three scanners) of the examined 15 subjects are displayed for each ABS tool, and the associated pairwise comparison results (i.e. p values) of the coefficients of variation among the ABS tools are also provided
GM gray matter, N.A. not available, WM white matter, VentralDC ventral diencephalon, L Left, R Right
The average CV of FSL-FIRST for different brain substructures is the mean over the structures available for quantification in FSL-FIRST (i.e. hippocampus, amygdala, thalamus, caudate, putamen, pallidum and accumbens)
Coefficient of variation (CV) for inter-scanner volumetric measurement between each pair of scanners
| Freesurfer | FSL-FIRST | AccuBrain | |||||||
|---|---|---|---|---|---|---|---|---|---|
| GE vs. PHILIPS | GE vs. SIEMENS | PHILIPS vs. SIEMENS | GE vs. PHILIPS | GE vs. SIEMENS | PHILIPS vs. SIEMENS | GE vs. PHILIPS | GE vs. SIEMENS | PHILIPS vs. SIEMENS | |
| WM | 8.393 | 0.903 | 7.874 | N.A. | N.A. | N.A. | 0.999 | 0.818 | 1.138 |
| GM | 6.835 | 1.142 | 7.223 | N.A. | N.A. | N.A. | 1.581 | 1.357 | 1.376 |
| Hippocampus L | 8.165 | 2.583 | 6.471 | 7.802 | 7.603 | 7.814 | 8.810 | 5.916 | 6.615 |
| Hippocampus R | 8.947 | 2.062 | 7.723 | 8.160 | 8.431 | 8.949 | 6.636 | 5.958 | 6.374 |
| Amygdala L | 9.192 | 7.257 | 7.429 | 15.23 | 8.490 | 18.15 | 2.029 | 3.246 | 3.233 |
| Amygdala R | 9.765 | 2.699 | 9.996 | 13.86 | 13.62 | 21.76 | 3.000 | 1.756 | 3.385 |
| Lateral ventricle L | 7.307 | 4.174 | 9.454 | N.A. | N.A. | N.A. | 2.332 | 1.352 | 2.419 |
| Lateral ventricle R | 8.131 | 2.478 | 8.356 | N.A. | N.A. | N.A. | 2.684 | 2.086 | 2.763 |
| VentralDC L | 9.536 | 2.908 | 9.250 | N.A. | N.A. | N.A. | 4.573 | 4.217 | 1.268 |
| VentralDC R | 9.621 | 2.952 | 8.060 | N.A. | N.A. | N.A. | 3.918 | 3.222 | 1.459 |
| Thalamus L | 8.538 | 2.611 | 7.701 | 3.052 | 8.805 | 8.357 | 1.569 | 0.984 | 1.354 |
| Thalamus R | 8.576 | 2.672 | 9.367 | 2.603 | 9.811 | 9.126 | 0.928 | 1.676 | 1.676 |
| Caudate L | 8.058 | 4.347 | 9.700 | 7.599 | 12.30 | 10.25 | 2.875 | 1.308 | 1.627 |
| Caudate R | 8.678 | 2.803 | 9.218 | 9.815 | 11.28 | 12.37 | 2.147 | 1.463 | 1.727 |
| Putamen L | 10.70 | 4.854 | 8.713 | 7.811 | 13.66 | 6.006 | 2.663 | 2.022 | 1.788 |
| Putamen R | 9.662 | 3.231 | 8.733 | 8.854 | 13.26 | 6.034 | 2.231 | 2.590 | 1.499 |
| Pallidum L | 9.381 | 7.032 | 12.41 | 5.857 | 7.079 | 4.253 | 4.174 | 3.178 | 2.232 |
| Pallidum R | 8.093 | 7.788 | 11.46 | 6.331 | 9.790 | 6.236 | 2.158 | 1.418 | 1.889 |
| Accumbens L | 9.714 | 7.977 | 9.544 | 7.800 | 10.96 | 13.44 | 3.572 | 3.583 | 4.223 |
| Accumbens R | 8.844 | 6.046 | 9.091 | 16.31 | 12.92 | 14.93 | 4.093 | 2.793 | 4.671 |
| Average. | 8.807 | 3.926 | 8.789 | 8.650 | 10.57 | 10.55 | 3.149 | 2.547 | 2.636 |
The mean values of region-specific CV (i.e. inter-scanner variability in brain volumetry of GE vs. PHILPS, GE vs. SIEMENS and PHILIPS vs. SIEMENS) of the examined 15 subjects are displayed for each ABS tool
GM gray matter, N.A: not available, WM white matter, VentralDC ventral diencephalon, L Left, R Right
The average CV of FSL-FIRST for different brain substructures is the mean over the structures available for quantification in FSL-FIRST (i.e. hippocampus, amygdala, thalamus, caudate, putamen, pallidum and accumbens)