| Literature DB >> 32851513 |
Takahiro Tamura1, Atsushi Mori2, Akira Ishii3, Masahiko Ando4, Yoko Kubo5, Kimitoshi Nishiwaki6.
Abstract
PURPOSE: Volatile anesthetics (VAs) protect myocardial cells in cardiovascular surgery. A recent clinical trial of cardiopulmonary bypass (CPB) surgery reported no significant difference in mortality rates between the use of VAs and total intravenous anesthetics at 1 year postoperatively. However, oxygenator function may affect the VA pharmacokinetics. Thus, we measured the VA blood concentrations during CPB in patients managed with four different microporous polypropylene hollow fiber membrane oxygenators.Entities:
Keywords: Cardiopulmonary bypass; Concentration; Desflurane; Oxygenator; Sevoflurane
Mesh:
Substances:
Year: 2020 PMID: 32851513 PMCID: PMC7449527 DOI: 10.1007/s00540-020-02844-1
Source DB: PubMed Journal: J Anesth ISSN: 0913-8668 Impact factor: 2.931
Fig. 1CONSORT flow diagram. TE CAPIOX (Terumo Co. Ltd.), LI INSPIRE (LivaNova Japan K.K.), ME Fusion (Medtronic Japan Co. Ltd.), SE EXELUNG (Senko Medical Instrument Mfg. Co. Ltd.), and HL human lung
Patients’ demographic and surgical characteristics
| Desflurane ( | Sevoflurane ( | |
|---|---|---|
| Demographic information | ||
| Age (years) | 63.3 (14.6) | 59.2 (17.2) |
| Height (cm) | 163.8 (9.5) | 165.9 (3.4) |
| Body weight (kg) | 61.0 (10.6) | 66.7 (10.4) |
| Body mass index (kg/m2) | 22.6 (2.5) | 24.3 (3.6) |
| Male:female | 11:1 | 9:3 |
| Surgical information | ||
| Single valvea | 2 | 4 |
| Coronary arteryb | 4 | 1 |
| Aortac | 4 | 3 |
| Congenitald | 0 | 1 |
| Complexe | 2 | 3 |
Data are expressed as means (standard deviation) or number of patients. Twenty-four patients scheduled for elective cardiopulmonary bypass were randomly allocated to one of the two VA groups (desflurane and sevoflurane groups) and, then, randomly divided into one of the four groups (TE, LI, ME, and SE) consisted of four individuals within each VA group. The same number of patients was also randomly and redundantly allocated to group HL in each VA group. Therefore, the patients in the HL group are not included in this table to avoid duplication
aaortic valve repair, aortic valve replacement, mitral valve repair, or mitral valve replacement
bcoronary artery bypass grafting
ctotal arch replacement ± elephant trunk, ascending aorta replacement
datrial septal defect
evalve + maze procedure, valve + coronary artery bypass grafting
Fig. 2Blood concentrations of desflurane and sevoflurane via oxygenator. a desflurane. b sevoflurane. The HL sampling for both anesthetics was taken 20 min after starting the delivery of 6.0 vol% desflurane or 1.7 vol% sevoflurane. Blood samples via each oxygenator were collected 20 min after starting the delivery of 6.0 vol% desflurane or 1.7 vol% sevoflurane to the oxygenator. Plots and horizontal lines in each group correspond to the measured values (n = 3) and the mean, respectively. HL human lung, TE CAPIOX (Terumo Co. Ltd.), LI INSPIRE (LivaNova Japan K.K.), ME Fusion (Medtronic Japan Co. Ltd.), and SE EXELUNG (Senko Medical Instrument Mfg. Co. Ltd.)
Fig. 3The microstructure of microporous polypropylene fibers from oxygenators under SEM. Fiber images from both unused (a) and clinically used (b) oxygenators were captured at 10,000 times magnification. Definite fiber images from unused SE could not be acquired due to its unique coating. Therefore, no further observation was performed. TE CAPIOX (Terumo Co. Ltd.); LI INSPIRE (LivaNova Japan K.K.), ME Fusion (Medtronic Japan Co. Ltd.), and SE EXELUNG (Senko Medical Instrument Mfg. Co. Ltd.). SEM scanning electron microscopy