José Naranjo-Orellana1, José Francisco Ruso-Álvarez1, José Luis Rojo-Álvarez2.
Abstract
The aim of this study was to validate the measurements of the beat intervals taken at rest by the Omegawave® device by comparing them to an ambulatory electrocardiogram system. For this purpose, the electrocardiogram was digitally processed, time-aligned, and scrutinized for its suitable use as gold-standard. Rest measurements were made for 10 minutes on 5 different days to 10 men and 3 women (24.8±5.05 years; 71.82±11.02 kg; 174.35±9.13 cm). RR intervals were simultaneously recorded using the Omegawave device and a Holter electrocardiogram. The processing of Holter electrocardiogram signals included the detrending of baseline noise and a high-pass filtering for emphasizing the QRS complexes and attenuating the T waves. After obtaining the RR intervals from the electrocardiogram, those from the Omegawave device were automatically aligned to them with cross-correlation digital processing techniques and compared to check whether both measurements could be considered superimposable. A Bland-Altman analysis was applied to the 5 measurements made for all subjects. The Omegawave device exhibited very strong agreement with a quality-controlled Holter electrocardiogram. Deviations not exceeding 25 ms could be expected in 95% of the cases, which is within manageable ranges both for clinical practice and for sports. Thieme. All rights reserved.Entities:
Mesh:
Year: 2020 PMID: 32842158 DOI: 10.1055/a-1157-9220