Sunghwan Shin1, Hyeonju Oh1, Hea Ree Park2, Eun Yeon Joo3, Soo-Youn Lee1.
Abstract
Melatonin and cortisol are clinically important for diagnosing sleep and mood disorders. We developed and validated a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for simultaneous measurement of salivary melatonin and cortisol concentrations according to the Clinical and Laboratory Standards Institute guidelines. Additionally, we compared the LC-MS/MS assay with immunoassays, ELISA (Direct Salivary Melatonin Elisa EK-DSM, Bühlmann Laboratories AG, Schönenbuch, Switzerland) and electrochemiluminescence immunoassay (Cortisol II, Roche, Mannheim, Germany), using 121 saliva samples. The LC-MS/MS assay exhibited good performance in terms of linearity, precision, accuracy, limit of detection, lower limit of quantification, extraction recovery, carry-over, and matrix effect. The LC-MS/MS assay and immunoassays showed strong correlation (Pearson's r=0.910 for melatonin, r=0.955 for cortisol), but demonstrated a significant mean bias of 23.2% (range 54.0-143.7%) for melatonin and 48.9% (range 59.7-184.7%) for cortisol. Our LC-MS/MS assay provided more sensitive and reliable salivary melatonin and cortisol quantification results compared with immunoassays.Entities:
Keywords: Cortisol; Immunoassay; Liquid chromatography-tandem mass spectrometry (LC-MS/MS); Melatonin; Performance; Saliva
Mesh:
Substances:
Year: 2020 PMID: 32829586 PMCID: PMC7443521 DOI: 10.3343/alm.2021.41.1.108
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
The performance of LC-MS/MS assay to measure melatonin and cortisol concentrations
| Analyte | Assay range | r | LOD | LLOQ | Accuracy (%) | Precision (CV, %) | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
| |||||||||||
| Intra-assay (N=5) | Inter-assay (N=30) | ||||||||||
|
|
| ||||||||||
| 8.6 (0.55) | 86 (5.52) | 344 (22.07) | 8.6 (0.55) | 86 (5.52) | 344 (22.07) | ||||||
| Melatonin | 2.15–430 pmol/L | 0.997 | 0.43 pmol/L | 2.15 pmol/L | 100.3–102.2 | 3.6 | 3.3 | 4.9 | 6.8 | 5.4 | 3.5 |
|
| |||||||||||
| Cortisol | 0.14–27.59 nmol/L | 0.999 | 0.03 nmol/L | 0.14 nmol/L | 96.9–107.8 | 2.6 | 3.1 | 3.1 | 4.7 | 4.7 | 3.7 |
Melatonin LLOQ: SNR=20.1, CV=5.26%, and percent bias=5.10%; cortisol LLOQ: SNR=192.4, CV=3.91%, and percent bias=6.78%;
Precision was evaluated at concentrations of 8.6, 86, and 344 pmol/L for melatonin and 0.55, 5.52, and 22.07 nmol/L for cortisol.
Abbreviations: r, correlation coefficient; LOD, limit of detection; LLOQ, lower limit of quantification; SNR, signal-to-noise ratio.
Fig. 1Comparison plots between LC-MS/MS and immunoassays. Top: (A) Passing-Bablok regression and (B) Bland-Altman plot to compare salivary melatonin concentrations measured by LC-MS/MS assay and ELISA. Bottom: (C) Passing-Bablok regression and (D) Bland-Altman plot to compare salivary cortisol concentrations measured by LC-MS/MS assay and ECLIA.
Abbreviations: LC-MS/MS, liquid chromatography-tandem mass spectrometry; ECLIA, electrochemiluminescence immunoassay.
Fig. 2Melatonin and cortisol profiles of two participants measured by LC-MS/MS assay and immunoassays.
Abbreviations: ECLIA, electrochemiluminescence immunoassay; LC-MS/MS, liquid chromatography-tandem mass spectrometry.
Mobile phase parameters for LC-MS/MS
| Time (min) | A (%) | B (%) |
|---|---|---|
| 0 | 75 | 25 |
| 1 | 75 | 25 |
| 3 | 10 | 90 |
| 3.5 | 10 | 90 |
| 3.6 | 75 | 25 |
(A) 2 mmol/L ammonium acetate in deionized water;
(B) 0.1% (v/v) formic acid in acetonitrile.
Abbreviation: LC-MS/MS, liquid chromatography-tandem mass spectrometry.
Analyte and internal standard MS parameters
| Analyte | Transition | MRM transition (m/z) | Collision energy (V) | Cell accelerator voltage (V) | RT (min) |
|---|---|---|---|---|---|
| Melatonin | Quantifier | 233.1 → 174.1 | 9 | 4 | 1.3 |
| Qualifier | 233.1 → 159.2 | 33 | 4 | 1.3 | |
|
| |||||
| Melatonin-d4 | Quantifier | 237.1 → 178.1 | 17 | 4 | 1.3 |
|
| |||||
| Cortisol | Quantifier | 363.1 → 120.9 | 25 | 4 | 2.2 |
| Qualifier | 363.1 → 327.4 | 13 | 4 | 2.2 | |
|
| |||||
| Cortisol-d4 | Quantifier | 367.4 → 121.1 | 25 | 4 | 2.2 |
Abbreviations: MS, mass spectrometry; MRM, multiple reaction monitoring; RT, retention time.