Do the tube current and metal artifact reduction influence the diagnosis of vertical root fracture in a tooth positioned in the vicinity of a zirconium implant? A CBCT study.

Rocharles C Fontenele1, Amanda Farias Gomes2, Yuri Nejaim3, Deborah Q Freitas1.   

Abstract

AIM: To evaluate the influence of the tube current and metal artifact reduction (MAR) tool on the diagnosis of vertical root fractures (VRF) in a tooth adjacent to a zirconium implant, in cone-beam computed tomography (CBCT) images.
METHODOLOGY: Thirty single-rooted teeth (15 with VRF and 15 control group) were individually positioned in a mandible, and scanned with the OP300 CBCT unit. Images were acquired using a standardized protocol: 5 × 5 cm field of view, 0.08-mm voxel size, and 90 kVp. Each tooth was scanned with and without a zirconium implant in its vicinity, using different tube currents (4 mA, 8 mA, and 10 mA) and conditions of MAR (enabled × disabled). Diagnostic values were calculated for each protocol, and compared by multi-way analysis of variance.
RESULTS: The ROC curve and sensitivity values did not differ significantly among the tube currents, regardless of the presence of the implant and MAR condition (p > 0.05). There were also no significant differences among the tube currents for the specificity values (p > 0.05); however, the specificity differed significantly between the "with implant" and "without implant" conditions, within the same MAR condition and tube current (p < 0.05). Specificity was significantly lower when the implant was present (p < 0.05).
CONCLUSION: The presence of a zirconium implant impairs the diagnosis of VRF in teeth adjacent to the artifact-generator material. Neither the tube current nor the MAR tool is effective in improving this diagnostic task. Therefore, in this clinical scenario, the use of the lowest tube current (4 mA), without MAR activation, is recommended. CLINICAL RELEVANCE: Considering that the tube current is one of the main factors that influence the radiation dose and image quality in CBCT, and that metal artifacts negatively influence the diagnosis of VRF in areas adjacent to the artifact-generator material, it is important to evaluate the effect of this energetic parameter in the diagnosis of VRF in teeth adjacent to zirconium implants.

Entities:  

Keywords:  Artifact; Cone-beam computed tomography; Tube current; Vertical root fracture; Zirconium implant

Mesh:

Substances:

Year:  2020        PMID: 32827079     DOI: 10.1007/s00784-020-03538-4

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  1 in total

1.  The Evaluation of Root Fracture with Cone Beam Computed Tomography (CBCT): An Epidemiological Study.

Authors:  Mehmet-Sinan Doğan; Michele Callea; Lindawati S Kusdhany; Ahmet Aras; Diah-Ayu Maharani; Masita Mandasari; Melissa Adiatman; Izzet Yavuz
Journal:  J Clin Exp Dent       Date:  2018-01-01
  1 in total
  3 in total

1.  Does the size of an object containing dental implant affect the expression of artifacts in cone beam computed tomography imaging?

Authors:  Mahkameh Moshfeghi; Yaser Safi; Ingrid Różyło-Kalinowska; Shiva Gandomi
Journal:  Head Face Med       Date:  2022-06-29       Impact factor: 2.246

Review 2.  The Impact of Cone-Beam Computed Tomography Exposure Parameters on Peri-Implant Artifacts: A Literature Review.

Authors:  Paweł Sawicki; Paweł J Zawadzki; Piotr Regulski
Journal:  Cureus       Date:  2022-03-10

3.  Effect of the Field of View Size on CBCT Artifacts Caused by the Presence of Metal Objects in the Exomass.

Authors:  Yaser Safi; Mitra Ghazizadeh Ahsaie; Maede Jafarian Amiri
Journal:  Int J Dent       Date:  2022-09-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.