Results From the Assessment of BCVA in the Smart Speaker and Human Administrations of the Snellen Visual Acuity Examination
| Statistical calculations | Eye | ||
|---|---|---|---|
| Right | Left | Both | |
| Sample size, No. of participants | 65 | 65 | 68 |
| Smart speaker BCVA LogMAR, mean (SEM) | 0.112 (0.020) | 0.108 (0.018) | 0.063 (0.014) |
| Human BCVA LogMAR, mean (SEM) | 0.090 (0.018) | 0.079 (0.016) | 0.049 (0.013) |
| .05 | .01 | .04 | |
| Difference between smart speaker and human BCVA LogMAR, mean (SEM) [95% CI] | 0.022 (0.002) [0.000 to 0.043] | 0.029 (0.002) [0.007 to 0.051] | 0.014 (0.001) [0.007 to 0.051] |
| Reliability, Cronbach α | 0.914 | 0.887 | 0.928 |
| Concordance correlation coefficient (95% CI) | 0.8340 (0.7460 to 0.8934) | 0.7799 (0.6680 to 0.8573) | 0.8585 (0.7837 to 0.9088) |
| Pearson ρ (precision) | 0.8502 | 0.8024 | 0.8722 |
| Bias correction factor Cb (accuracy) | 0.9810 | 0.9720 | 0.9843 |
| Bland-Altman limit, mean difference (95% CI) | −0.02 (−0.19 to 0.15) | −0.03 (−0.20 to 0.15) | −0.01 (−0.13 to 0.10) |
Abbreviations: BCVA, best corrected visual acuity; LogMAR, logarithm of the minimum angle of resolution.