Literature DB >> 32818330

In vitro optical quality assessment of a monofocal IOL sutured to an artificial iris.

Christian Mayer1, Hyeck-Soo Son, Grzegorz Łabuz, Timur M Yildirim, Gerd U Auffarth, Ramin Khoramnia.   

Abstract

PURPOSE: To assess whether the combined implantation of a monofocal IOL and an artificial iris had an effect on the IOL's optical performance.
SETTING: David J. Apple International Laboratory for Ocular Pathology, Heidelberg, Germany.
DESIGN: In vitro laboratory study.
METHODS: IOL optical quality was assessed using an OptiSpheric IOL Pro II to measure the IOL's modulation transfer function (MTF) at 3.0 mm pupil size and spatial frequency of 100 lp/mm. Three ASPIRA-aAY IOLs with different base powers, 10.0 diopter (D) (IOL A), 20.0 D (IOL B), and 30.0 D (IOL C) were measured before and after suturing the IOL to an ArtificialIris (AI). The degree of IOL decentration about the center of the AI was also evaluated.
RESULTS: The mean MTF values prior to suturing were 0.57, 0.65, and 0.63 for IOLs A, B, and C, respectively. After suturing to the AI, the mean MTF values were 0.52, 0.54, and 0.55 for IOLs A, B, and C, respectively. The decentration values in vertical direction were 0.20 mm, 0.00 mm, and 0.02 mm for IOLs A, B, and C, respectively. In horizontal direction, the decentration values were 0.42 mm, 0.10 mm, and 0.03 mm for IOLs A, B, and C, respectively.
CONCLUSIONS: The MTF decreased slightly in all 3 IOLs after they were sutured to the AI. The small differences, however, should be clinically irrelevant. This laboratory assessment showed that suturing of the IOL to the AI can be performed in a reliable and reproducible manner without deteriorating optical quality.

Entities:  

Year:  2020        PMID: 32818330     DOI: 10.1097/j.jcrs.0000000000000287

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  7 in total

1.  "Double Prosthesis Implantation": Biometry and Refractive Outcomes in Combined Intraocular Lens and Artificial Iris Surgery.

Authors:  Christian Mayer; Ramin Khoramnia
Journal:  Clin Ophthalmol       Date:  2021-02-24

2.  Surgical Management for Silicone Oil Barrier of Traumatic Aniridia with Aphakia: Suturing of Temporary Iris-Diaphragm Prior to Final Iris-Lens-Diaphragm Implantation.

Authors:  Christian S Mayer; Isabella Baur; Julia Storr; Alexander Markard; Ramin Khoramnia
Journal:  Clin Ophthalmol       Date:  2020-12-21

3.  Traumatic iridodialysis and mydriasis: Surgical reconstruction of the iris-lens-diaphragm with an iris implant and Intraocular lens.

Authors:  Isabella Diana Baur; Christian Steffen Mayer; Julia Storr; Ramin Khoramnia
Journal:  Am J Ophthalmol Case Rep       Date:  2022-04-20

4.  Functional outcomes after combined iris and intraocular lens implantation in various iris and lens defects.

Authors:  Christian S Mayer; Andrea M Hoffmann; Phillipp Prahs; Lukas Reznicek; Ramin Khoramnia
Journal:  BMC Ophthalmol       Date:  2020-09-15       Impact factor: 2.209

5.  Aesthetics of iris reconstruction with a custom-made artificial iris prosthesis.

Authors:  Timur M Yildirim; Ramin Khoramnia; Michael Masyk; Hyeck-Soo Son; Gerd U Auffarth; Christian S Mayer
Journal:  PLoS One       Date:  2020-08-13       Impact factor: 3.240

6.  Complete anterior segment reconstruction: Corneal transplantation and implantation of an iris prosthesis and IOL in a single surgery.

Authors:  Christian Mayer; Isabella Diana Baur; Julia Storr; Ramin Khoramnia
Journal:  Eur J Ophthalmol       Date:  2021-01-28       Impact factor: 2.597

Review 7.  [Pupil reconstruction with an artificial iris].

Authors:  Christian Mayer; Ramin Khoramnia
Journal:  Ophthalmologe       Date:  2021-06-28       Impact factor: 1.059

  7 in total