| Literature DB >> 32810840 |
Christian Wechselberger1, Susanne Süßner2, Stefan Doppler3, David Bernhard4.
Abstract
BACKGROUND: Serological assays for the determination of the immune status of patients that have tested positive for infection with SARS-CoV-2 by RT-PCR are required for, e.g., contact tracing and epidemiological studies. However, data concerning the performance parameters of commercially available high-throughput ELISA tests are still not available on a large scale. STUDYEntities:
Keywords: COVID-19; ELISA; Evaluation; Immunoglobulin; Point of care; SARS-CoV-2
Mesh:
Substances:
Year: 2020 PMID: 32810840 PMCID: PMC7417972 DOI: 10.1016/j.jcv.2020.104589
Source DB: PubMed Journal: J Clin Virol ISSN: 1386-6532 Impact factor: 3.168
| Assay Manufacturer | Immunoglobulin classes detected | Antigens used | ||||
|---|---|---|---|---|---|---|
| Epitope diagnostics Inc. | IgG | Recombinant full length nucleocapsid protein | ||||
| Abbott Architect | IgG | Recombinant full length nucleocapsid protein | ||||
| Euroimmun AG | IgA | IgG | S1-Domain of Spike protein including receptor-binding domain | |||
| Vircell S.L.U. | IgG | Recombinant antigens from Spike glycoprotein and nucleocapsid protein | ||||
| Mikrogen GmbH | IgG | Recombinant full length nucleocapsid protein | ||||
| IgA | IgG | IgM | Full length spike glycoprotein | |||
| Genspeed Biotech | IgG | Receptor Binding Domain / Full length spike glycoprotein / Nucleoprotein | ||||
Characteristics of different SARS-CoV-2 assays used in this study.
| Epitope diagnostics Inc. | 50 | 13 | 46 | 1 | 0,98 | 0,78 | ||
| Abbott Architect | 48 | 0 | 59 | 3 | 0,94 | 1,00 | ||
| Euroimmun AG | 47 | 1 | 58 | 4 | 0,92 | 0,98 | ||
| Vircell S.L.U. | 50 | 10 | 49 | 1 | 0,98 | 0,83 | ||
| Mikrogen GmbH | 51 | 2 | 57 | 0 | 1,00 | 0,97 | ||
| 49 | 4 | 55 | 2 | 0,96 | 0,93 | |||
| Genspeed Biotech | 51 | 4 | 55 | 0 | 1,00 | 0,93 |
Sensitivity and specificity values for SARS-CoV-2 assays based on the detection of IgG.
| Euroimmun AG | 50 | 1 | 57 | 2 | 0,96 | 0,98 | ||
| 57 | 2 | 49 | 2 | 0,97 | 0,96 |
Sensitivity and specificity values for SARS-CoV-2 assays utilizing the detection of IgA / IgG (Euroimmun AG) and IgA / IgG / IgM (in-house ELISA).
Fig. 1Signal intensities for immunoglobulin classes A, G and M as obtained with the in-house ELISA. Samples designated false positive according to the criteria described in the Results section are indicated with an asterisk (⁕). Samples that are lacking detectable IgG antibodies and tested positive for IgM with our in-house ELISA are labeled with an arrow (↑).
Fig. 2Distribution of the signal intensities of the immunoglobulin data as obtained with the in-house ELISA. The box indicates the interquartile range (IQR), the whiskers the range of values that are within 1.5x IQR and the horizontal lines indicate the respective median.