Comparing the accuracy of the da Vinci Xi and da Vinci Si for image guidance and automation.

James M Ferguson1,2, Bryn Pitt1,2, Alan Kuntz3, Josephine Granna1,2, Nicholas L Kavoussi2,4, Naren Nimmagadda2,4, Eric J Barth1,2, Stanley Duke Herrell2,4, Robert J Webster1,2,4.   

Abstract

BACKGROUND: Current laparoscopic surgical robots are teleoperated, which requires high fidelity differential motions but does not require absolute accuracy. Emerging applications, including image guidance and automation, require absolute accuracy. The absolute accuracy of the da Vinci Xi robot has not yet been characterized or compared to the Si system, which is now being phased out. This study compares the accuracy of the two.
METHODS: We measure robot tip positions and encoder values assessing accuracy with and without robot calibration.
RESULTS: The Si is accurate if the setup joints are not moved but loses accuracy otherwise. The Xi is always accurate.
CONCLUSION: The Xi can achieve submillimetric average error. Calibration improves accuracy, but excellent baseline accuracy of the Xi means that calibration may not be needed for some applications. Importantly, the external tracking systems needed to account for setup joint error in the Si are no longer required with the Xi.
© 2020 John Wiley & Sons Ltd.

Entities:  

Keywords:  image-guided surgery; surgical automation; surgical robot calibration; surgical robotics

Mesh:

Year:  2020        PMID: 32808429      PMCID: PMC9540612          DOI: 10.1002/rcs.2149

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.483


  30 in total

1.  Dexterity enhancement with robotic surgery.

Authors:  K Moorthy; Y Munz; A Dosis; J Hernandez; S Martin; F Bello; T Rockall; A Darzi
Journal:  Surg Endosc       Date:  2004-04-06       Impact factor: 4.584

2.  Kidney deformation and intraprocedural registration: a study of elements of image-guided kidney surgery.

Authors:  Hernan O Altamar; Rowena E Ong; Courtenay L Glisson; Davis P Viprakasit; Michael I Miga; Stanley Duke Herrell; Robert L Galloway
Journal:  J Endourol       Date:  2010-12-13       Impact factor: 2.942

3.  Augmented reality during robot-assisted laparoscopic partial nephrectomy: toward real-time 3D-CT to stereoscopic video registration.

Authors:  Li-Ming Su; Balazs P Vagvolgyi; Rahul Agarwal; Carol E Reiley; Russell H Taylor; Gregory D Hager
Journal:  Urology       Date:  2009-02-04       Impact factor: 2.649

4.  Transoral robotic surgery using an image guidance system.

Authors:  Shaun C Desai; Chih-Kwang Sung; Eric M Genden
Journal:  Laryngoscope       Date:  2008-11       Impact factor: 3.325

5.  Cadaveric feasibility study of da Vinci Si-assisted cochlear implant with augmented visual navigation for otologic surgery.

Authors:  Wen P Liu; Mahdi Azizian; Jonathan Sorger; Russell H Taylor; Brian K Reilly; Kevin Cleary; Diego Preciado
Journal:  JAMA Otolaryngol Head Neck Surg       Date:  2014-03       Impact factor: 6.223

6.  Predicting error in rigid-body point-based registration.

Authors:  J M Fitzpatrick; J B West; C R Maurer
Journal:  IEEE Trans Med Imaging       Date:  1998-10       Impact factor: 10.048

7.  Intraoperative registered transrectal ultrasound guidance for robot-assisted laparoscopic radical prostatectomy.

Authors:  Omid Mohareri; Joseph Ischia; Peter C Black; Caitlin Schneider; Julio Lobo; Larry Goldenberg; Septimiu E Salcudean
Journal:  J Urol       Date:  2014-08-21       Impact factor: 7.450

8.  Does transition from the da Vinci Si to Xi robotic platform impact single-docking technique for robot-assisted laparoscopic nephroureterectomy?

Authors:  Manish N Patel; Ahmed Aboumohamed; Ashok Hemal
Journal:  BJU Int       Date:  2015-07-18       Impact factor: 5.588

9.  Tracked "Pick-Up" Ultrasound for Robot-Assisted Minimally Invasive Surgery.

Authors:  Caitlin Schneider; Christopher Nguan; Robert Rohling; Septimiu Salcudean
Journal:  IEEE Trans Biomed Eng       Date:  2015-07-07       Impact factor: 4.538

10.  Ergonomics, user comfort, and performance in standard and robot-assisted laparoscopic surgery.

Authors:  R H van der Schatte Olivier; C D P Van't Hullenaar; J P Ruurda; I A M J Broeders
Journal:  Surg Endosc       Date:  2008-10-15       Impact factor: 4.584

View more
  1 in total

1.  Da Vinci robot-assisted pancreato-duodenectomy in a patient with situs inversus totalis: A case report and review of literature.

Authors:  Bai-Bei Li; Shi-Liu Lu; Xiang He; Biao Lei; Jian-Ni Yao; Si-Chen Feng; Shui-Ping Yu
Journal:  World J Gastrointest Oncol       Date:  2022-07-15
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.