Anodic stripping voltammetry procedure modified for improved accuracy of blood lead analysis.

S M Roda1, R D Greenland, R L Bornschein, P B Hammond.   

Abstract

In evaluating the accuracy and reliability of blood lead (PbB) measurements with the Environmental Science Associates Model 3010A Trace Metal Analyzer, intralaboratory comparison demonstrated that use of the operating conditions recommended by the manufacturer resulted in consistently underestimated PbB concentrations less than 400 micrograms/L and overestimated PbB values greater than 400 micrograms/L. At PbB concentrations less than 50 micrograms/L, measured concentrations were often registered as negative results. However, these negative values could be replicated to within +/- 10 micrograms/L, indicating good precision of the method, but obviously not good accuracy. In addition, lower-than-expected lead (Pb) values were measured in samples containing increased concentrations of copper (Cu), such as may occur in pregnant women. We modified the procedure to eliminate these inaccuracies by substituting manual peak-height measurements for reliance on the integrator and digital display of the instrument. We established the accuracy of the modified procedure by using calibration standards previously quantified by isotope dilution-mass spectroscopy. A quality-control program for monitoring PbB analysis is also described.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3280164

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  10 in total

1.  The influence of age of lead exposure on adult gray matter volume.

Authors:  Christopher J Brubaker; Kim N Dietrich; Bruce P Lanphear; Kim M Cecil
Journal:  Neurotoxicology       Date:  2010-03-11       Impact factor: 4.294

2.  The relationship between surface dust lead loadings on carpets and the blood lead of young children.

Authors:  S Clark; R L Bornschein; W Pan; W Menrath; S Roda; J Grote
Journal:  Environ Geochem Health       Date:  1996-12       Impact factor: 4.609

3.  Improving Reproducibility of Lab-on-a-Chip Sensor with Bismuth Working Electrode for Determining Zn in Serum by Anodic Stripping Voltammetry.

Authors:  Xing Pei; Wenjing Kang; Wei Yue; Adam Bange; William R Heineman; Ian Papautsky
Journal:  J Electrochem Soc       Date:  2014-02-01       Impact factor: 4.316

4.  Screening for lead poisoning in urban school children of southern India using capillary and venous blood samples.

Authors:  Herman S D'Souza; Geraldine Menezes; T Venkatesh
Journal:  Indian J Clin Biochem       Date:  2002-01

5.  Reduced regional volumes associated with total psychopathy scores in an adult population with childhood lead exposure.

Authors:  Travis J Beckwith; Kim N Dietrich; John P Wright; Mekibib Altaye; Kim M Cecil
Journal:  Neurotoxicology       Date:  2018-04-07       Impact factor: 4.294

6.  Developmental lead exposure and adult criminal behavior: A 30-year prospective birth cohort study.

Authors:  John Paul Wright; Bruce P Lanphear; Kim N Dietrich; Michelle Bolger; Lisa Tully; Kim M Cecil; Catherine Sacarellos
Journal:  Neurotoxicol Teratol       Date:  2021-02-20       Impact factor: 3.763

7.  Criminal arrests associated with reduced regional brain volumes in an adult population with documented childhood lead exposure.

Authors:  Travis J Beckwith; Kim N Dietrich; John P Wright; Mekibib Altaye; Kim M Cecil
Journal:  Environ Res       Date:  2021-06-25       Impact factor: 8.431

8.  Lead exposure and neurobehavioral development in later infancy.

Authors:  K N Dietrich; P A Succop; R L Bornschein; K M Krafft; O Berger; P B Hammond; C R Buncher
Journal:  Environ Health Perspect       Date:  1990-11       Impact factor: 9.031

9.  Lead Exposure during Early Human Development and DNA Methylation of Imprinted Gene Regulatory Elements in Adulthood.

Authors:  Yue Li; Changchun Xie; Susan K Murphy; David Skaar; Monica Nye; Adriana C Vidal; Kim M Cecil; Kim N Dietrich; Alvaro Puga; Randy L Jirtle; Cathrine Hoyo
Journal:  Environ Health Perspect       Date:  2015-06-26       Impact factor: 9.031

10.  Disposable copper-based electrochemical sensor for anodic stripping voltammetry.

Authors:  Xing Pei; Wenjing Kang; Wei Yue; Adam Bange; William R Heineman; Ian Papautsky
Journal:  Anal Chem       Date:  2014-04-28       Impact factor: 6.986

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.