| Literature DB >> 32779832 |
Anna Laura Licon1, Ara Alexandrian1, Daniel Saenz1, Pamela Myers1, Karl Rasmussen1, Sotirios Stathakis1, Niko Papanikolaou1, Neil Kirby1.
Abstract
PURPOSE: To create an open-source visualization program that allows one to find potential cone collisions while planning intracranial stereotactic radiosurgery cases.Entities:
Keywords: code; cone collision; open source; stereotactic radiosurgery
Mesh:
Year: 2020 PMID: 32779832 PMCID: PMC7592959 DOI: 10.1002/acm2.12998
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Abbreviation and description of each measurement needed for commissioning, along with examples.
| Name | Description | Our measurement |
|---|---|---|
| SZS | Safety zone distance to superior of isocenter | 165 mm |
| SZI | Safety zone distance to inferior of isocenter | 165 mm |
| SZA | Safety zone distance to the anterior of the isocenter | 155 cm |
| SZP | Safety zone distance to posterior of isocenter | 120 mm |
| SZR | Safety zone distance to the patient right of isocenter | 115 mm |
| SZL | Safety zone distance to the patient left of isocenter | 115 mm |
| COD | Cone outer diameter | 750 mm |
| CHT | Couch head thickness | 20 mm |
| CHAD | Couch head axis distance | 120 mm |
| CAD | Cone axis distance | 256 mm |
| CHW | Couch head width | 282 mm |
| CHSE | Couch head superior extent | 200 mm |
| CHIE | Couch head inferior extent | −113 mm |
| CBIE | Couch body inferior extent (this can be reduced for visualization purposes) | −2240 mm |
| CBW | Couch body width | 530 mm |
| HOD | Head outer diameter | 670 mm |
| HT | Head thickness (this can be reduced for visualization purposes) | 670 mm |
| HAD | Head axis distance | 437 mm |
| GIA | Gantry initial angle (in units of degrees) | 180 |
| GICW | Gantry increases clockwise (1 = true, 0 = false). | 0 |
| CIA | Couch initial angle (in units of degrees) | 180 |
| CICW | Couch increases clockwise (1 = true, 0 = false). | 1 |
| szlatcomm | Lateral position for the center of the safety zone at commissioning | 2 |
| szapcomm | AP position for the center of the safety zone at commissioning | 1.8 |
| szsicomm | Sup/inf position for the center of the safety zone at commissioning | 0.8 |
| reflatcomm | Lateral position for the reference position at commissioning | 103.1 |
| refapcomm | AP position for the reference position at commissioning | 4 |
| refSIcomm | Sup/inf position for the reference position at commissioning | 23.3 |
| APDIR | Anterior–posterior shift sign convention | −1 |
| LATDIR | Lateral posterior shift sign convention | 1 |
| SIDIR | Sup/inf posterior shift sign convention | −1 |
Fig. 1Initial view of the linac, couch, and patient safety zone for the gantry directly overhead, with no couch kick.
Fig. 2An illustration of the relationship between the isocenter, center of the patient safety zone, and the reference point.
Fig. 3Annotated graphical representation of the measurements needed for the gantry, cone, couch, and patient safety zone (shown as an ellipsoid). The center of this ellipsoid is the center of the safety zone.
Fig. 4A demonstration of the software usage. (a–c) show the usage in defining the reference location, treatment isocenter, and beam geometry, respectively. (d) shows the corresponding visualization.
Fig. 5Side‐by‐side demonstration of a collision in the room and on the GUI with the box patient safety zone.
Fig. 6Side‐by‐side demonstration of a noncollision arc in the room and on the GUI with the box patient safety zone.
Patient #1 had a right orbital gyrus lesion. The original plan used beams 1–4. After plan approval, the beams were simulated on the machine. Beams 3 and 4 were found to collide, thus a new plan was created using beams 5–8.
| Beam | Couch angle | Gantry start | Gantry stop | Software (clear, close, collides) | Real (clear, close, collides) |
|---|---|---|---|---|---|
| 1 | 180 | 200 | 350 | Clear | Clear |
| 2 | 255 | 160 | 10 | Close | Close |
| 3 | 205 | 10 | 160 | Collide | Collide |
| 4 | 130 | 200 | 350 | Collide | Collide |
| 5 | 255 | 165 | 45 | Close | Close |
| 6 | 195 | 55 | 165 | Clear | Clear |
| 7 | 165 | 195 | 315 | Close | Clear |
| 8 | 105 | 315 | 195 | Close | Clear |
Patient #2 had a left cerebellar lesion. The original plan used beams 1, 2, and 3. After plan approval, the beams were simulated on the machine. Beams 2 and 3 were found to be too close, thus a new plan was created using beams 1, 4, and 5
| Beam | Couch angle | Gantry start | Gantry stop | Software (clear, close, collides) | Real (clear, close, collides) |
|---|---|---|---|---|---|
| 1 | 180 | 150 | 10 | Clear | Clear |
| 2 | 255 | 10 | 150 | Close | Close |
| 3 | 105 | 220 | 350 | Close | Close |
| 4 | 270 | 10 | 140 | Clear | Clear |
| 5 | 225 | 150 | 10 | Clear | Clear |
Patient #3 had a right cerebellar peduncle lesion. For this treatment, the software was used proactively to find couch angles that were problematic. This showed beams 1 and 2 could be issues. Beams 3 to 7 were chosen to avoid these angles and were used for treatment
| Beam | Couch angle | Gantry start | Gantry stop | Software (clear, close, collides) | Real (clear, close, collides) |
|---|---|---|---|---|---|
| 1 | 125 | 190 | 350 | Clear | Clear |
| 2 | 230 | 10 | 170 | Collides | Clear |
| 3 | 180 | 220 | 350 | Clear | Clear |
| 4 | 245 | 80 | 10 | Clear | Clear |
| 5 | 205 | 10 | 90 | Clear | Clear |
| 6 | 145 | 220 | 350 | Clear | Clear |
| 7 | 115 | 350 | 220 | Clear | Clear |