Literature DB >> 32769599

Safety and Performance of Bone-Anchored Prostheses in Persons with a Transfemoral Amputation: A 5-Year Follow-up Study.

D Reetz1, R Atallah, J Mohamed, H van de Meent, J P M Frölke, R Leijendekkers.   

Abstract

BACKGROUND: For almost 30 years, bone-anchored prostheses have offered an alternative solution to prosthetic sockets by attaching the artificial limb directly to the femoral residuum by means of an osseointegration implant. Osseointegration implant surgery was introduced in our center in 2009. The aim of the present study is to report on safety, prosthesis-wearing time, and health-related quality-of-life (HRQoL) for patients with femoral bone-anchored prostheses during a 5-year follow-up period.
METHODS: All patients who underwent implantation of a press-fit osseointegration implant between May 2009 and November 2013 were eligible for the present study. Implantation was performed in 2 stages. Adverse events included infectious complications (grade 1 to 4), aseptic loosening, breakage, stoma-redundant tissue, and stoma hypergranulation. Prosthesis-wearing time and HRQoL were measured with the Questionnaire for Persons with a Transfemoral Amputation (Q-TFA) prosthetic use score and global score, respectively.
RESULTS: Thirty-nine of 42 eligible patients were included. Thirty patients (77%) presented with some kind of infection (156 events in total), with 148 (95%) events being classified as grade 1 or 2 and 8 events (5%) being classified as grade 3; the latter 8 events occurred in 4 patients. There were no instances of septic loosening. The intramedullary stem of the osseointegration implant broke in 2 patients. In total, soft-tissue refashioning had to be done 30 times in 14 patients. The Q-TFA median prosthetic use and global scores improved significantly from 71 to 100 and from 33 to 75, respectively (p < 0.001).
CONCLUSIONS: Despite the adverse events, patient prosthetic use and HRQoL improved significantly. Grade-1 and 2 infections were frequent but could mostly be treated with nonoperative measures. Most infections seemed to occur in the first 2 years and did not lead to deep infections. Two broken intramedullary stems were revised successfully. Current developments focus on reduction of infectious complications and prevention of osseointegration implant breakage. LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.

Entities:  

Year:  2020        PMID: 32769599     DOI: 10.2106/JBJS.19.01169

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  6 in total

1.  Comparison of functional outcome and patient satisfaction between patients with socket prosthesis and patients treated with transcutaneous osseointegrated prosthetic systems (TOPS) after transfemoral amputation.

Authors:  Marcel Winkelmann; Alexander Ranker; Marcus Örgel; Frederik Schwarze; Tilman Graulich; Christian Krettek; Friederike Weidemann; Horst-Heinrich Aschoff
Journal:  Eur J Trauma Emerg Surg       Date:  2022-06-18       Impact factor: 3.693

2.  CORR Insights®: What Are the Risk Factors for Mechanical Failure and Loosening of a Transfemoral Osseointegrated Implant System in Patients with a Lower-limb Amputation?

Authors:  Benjamin K Potter
Journal:  Clin Orthop Relat Res       Date:  2022-04-01       Impact factor: 4.755

3.  What Are the Risk Factors for Mechanical Failure and Loosening of a Transfemoral Osseointegrated Implant System in Patients with a Lower-limb Amputation?

Authors:  Jamal Mohamed; David Reetz; Henk van de Meent; Hendrik Schreuder; Jan Paul Frölke; Ruud Leijendekkers
Journal:  Clin Orthop Relat Res       Date:  2022-04-01       Impact factor: 4.755

4.  Extramedullary Osseointegration-A Novel Design of Percutaneous Osseointegration Prosthesis for Amputees.

Authors:  Yingying Sun; Jinying Xu; Shuang Lv; Ziran Xu; Lisha Li; Yan Li; Yulin Li
Journal:  Front Bioeng Biotechnol       Date:  2022-02-10

5.  Bone Bricks: The Effect of Architecture and Material Composition on the Mechanical and Biological Performance of Bone Scaffolds.

Authors:  Evangelos Daskalakis; Boyang Huang; Cian Vyas; Anil A Acar; Fengyuan Liu; Ali Fallah; Glen Cooper; Andrew Weightman; Gordon Blunn; Bahattin Koç; Paulo Bartolo
Journal:  ACS Omega       Date:  2022-02-22

6.  Elective transfemoral amputation and simultaneous implantation of a transcutaneous osseointegrated prosthesis stem as salvage treatment after knee joint arthrodesis with poor function: A case report.

Authors:  Katharina Krause; Katherina Richter; Thomas Beyer; Horst Heinrich Aschoff; Dagmar-Christiane Fischer; Thomas Mittlmeier
Journal:  Front Surg       Date:  2022-08-30
  6 in total