Zoraida S Gaurisankar1,2, Gwyneth A van Rijn3, Gregorius P M Luyten3, Jan-Willem M Beenakker3,4. 1. Department of Ophthalmology, Leiden University Medical Center, Leiden, the Netherlands. z.s.gaurisankar@lumc.nl. 2. , Leiden, the Netherlands. z.s.gaurisankar@lumc.nl. 3. Department of Ophthalmology, Leiden University Medical Center, Leiden, the Netherlands. 4. Department of Radiology, C.J. Gorter Center for High-Field MRI, Leiden University Medical Center, Leiden, the Netherlands.
Abstract
PURPOSE: To investigate the agreement and reliability of anterior segment optical coherence tomography (AS-OCT) and Scheimpflug imaging in measuring the distance from the anterior edge of an iris-fixated phakic intraocular lens (IF-pIOL) to the corneal endothelium. METHODS: Anterior segment configuration was assessed in a total of 62 eyes of which 25 hyperopic and 37 myopic eyes, all corrected with an IF-pIOL. Measurements were performed by two independent observers using AS-OCT (Visante, Model 1000, Carl Zeiss Meditec Inc.) and Scheimpflug imaging (Pentacam HR, Oculus Optikgerate). The distance from the anterior edge of the pIOL to the endothelium was measured in five different positions using both modalities with their corresponding pIOL software. The measurements as well as the inter- and intra-observer reliability of the two imaging modalities were then compared. RESULTS: Distance measurements for all positions performed by AS-OCT were found to be significantly larger than those performed by Scheimpflug imaging, with mean differences ranging from 0.11 to 0.22 mm. Both instruments exhibited good inter- and intra-observer reliability. CONCLUSION: Anterior pIOL edge to endothelium distance measurements by AS-OCT and Scheimpflug imaging have good intra- and inter-observer reliability. However, as AS-OCT provides larger measurements, these two modalities cannot be used interchangeably. Correction of this difference might be essential for proper decision-making during pre-operative screening for pIOL implantation and post-operative safety monitoring.
PURPOSE: To investigate the agreement and reliability of anterior segment optical coherence tomography (AS-OCT) and Scheimpflug imaging in measuring the distance from the anterior edge of an iris-fixated phakic intraocular lens (IF-pIOL) to the corneal endothelium. METHODS:Anterior segment configuration was assessed in a total of 62 eyes of which 25 hyperopic and 37 myopic eyes, all corrected with an IF-pIOL. Measurements were performed by two independent observers using AS-OCT (Visante, Model 1000, Carl Zeiss Meditec Inc.) and Scheimpflug imaging (Pentacam HR, Oculus Optikgerate). The distance from the anterior edge of the pIOL to the endothelium was measured in five different positions using both modalities with their corresponding pIOL software. The measurements as well as the inter- and intra-observer reliability of the two imaging modalities were then compared. RESULTS: Distance measurements for all positions performed by AS-OCT were found to be significantly larger than those performed by Scheimpflug imaging, with mean differences ranging from 0.11 to 0.22 mm. Both instruments exhibited good inter- and intra-observer reliability. CONCLUSION: Anterior pIOL edge to endothelium distance measurements by AS-OCT and Scheimpflug imaging have good intra- and inter-observer reliability. However, asAS-OCT provides larger measurements, these two modalities cannot be used interchangeably. Correction of this difference might be essential for proper decision-making during pre-operative screening for pIOL implantation and post-operative safety monitoring.
Authors: Muriël Doors; Diana W J K Cals; Tos T J M Berendschot; John de Brabander; Fred Hendrikse; Carroll A B Webers; Rudy M M A Nuijts Journal: J Cataract Refract Surg Date: 2008-12 Impact factor: 3.351
Authors: David Huang; Steven C Schallhorn; Alan Sugar; Ayad A Farjo; Parag A Majmudar; William B Trattler; David J Tanzer Journal: Ophthalmology Date: 2009-11 Impact factor: 12.079
Authors: Ruchi Saxena; Sharmila S Boekhoorn; Paul G H Mulder; Bastiaantje Noordzij; Gabriel van Rij; Gre P M Luyten Journal: Ophthalmology Date: 2007-08-07 Impact factor: 12.079
Authors: Muriël Doors; Tos T J M Berendschot; Fred Hendrikse; Carroll A B Webers; Rudy M M A Nuijts Journal: J Cataract Refract Surg Date: 2009-03 Impact factor: 3.351