| Literature DB >> 32754434 |
Samaa A Kasem1, Ashraf Abd Elmawgood Bassiouny2, Doaa Abu Elkassim Rashwan1, Mahmoud Hussein Bahr1.
Abstract
OBJECTIVES: This study compared the efficacy and safety of minimal tourniquet pressure using either determined limb occlusion pressure (LOP) or estimated arterial occlusion pressure (AOP) for elective upper limb surgeries.Entities:
Keywords: Arterial Occlusion Pressure; Tourniquet Pressure; limb Occlusion Pressure
Year: 2020 PMID: 32754434 PMCID: PMC7353217 DOI: 10.5812/aapm.102124
Source DB: PubMed Journal: Anesth Pain Med ISSN: 2228-7523
Figure 1.Consort flow participant diagram
Demographic Data, Arm Circumference, Tourniquet Times, Fentanyl Requirements[a, b, c]
| Group (A) (N = 20) | Group (B) (N = 20) | P Value | |
|---|---|---|---|
|
| 25.6 ± 6.9 | 27.2 ± 9.6 | 0.280 |
|
| 14/6 | 13/7 | |
|
| 27.6 ± 1.6 | 28.3 ± 1.4 | 0.082 |
|
| 55 ± 21 | 58 ± 20 | 0.630 |
|
| 171 ± 16.24 | 175 ± 15.8 | 0.209 |
aData represented as mean ± SD or number.
bGroup (A): tourniquet inflation pressure was based on the AOP; Group (B): tourniquet inflation pressure was being based on LOP
cP value < 0.05 was considered statistically significantly different.
Time to Estimate AOP or Determinate LOP and Set the Tourniquet Cuff Pressure, AOP in Group (A), LOP in Group (B), Tourniquet Inflation Pressures, Systolic Blood Pressure (Initial and Maximal)[a, b, c]
| Group (A) (N = 20) | Group (B) (N = 20) | P Value | |
|---|---|---|---|
|
| 62 ± 15 | 120.8 ± 3.4* | < 0.001 |
|
| 118 ± 2 | 91 ± 2* | < 0.001 |
|
| 133 ± 2 | 132 ± 2 | 0.068 |
|
| 88.1 ± 3.6 | 89.9 ± 3.5 | 0.061 |
|
| 89.1 ± 2.3 | 90.2 ± 2.9 | 0.093 |
aData represented as mean ± SD.
bGroup (A): tourniquet inflation pressure was based on the AOP, Group (B): tourniquet inflation pressure was being based on LOP
cStatistically highly significant compared to group (A), P value of < 0.001.
Figure 2.Heart rate changes. Group (A): tourniquet inflation pressure was based on the AOP. Group (B): tourniquet inflation pressure was being based on LOP
The Performance of the Tourniquets Assessed by the Surgeon During Surgery[a, b, c]
| Group (A) (N = 20) | Group (B) (N = 20) | |||||
|---|---|---|---|---|---|---|
| Initial | Middle | End | Initial | Middle | End | |
|
| 19 (95) | 19 (95) | 18 (95) | 18 (90) | 18 (90) | 18 (90) |
|
| 1 (5) | 1 (5) | 1 (5) | 2 (10) | 2 (10) | 2 (10) |
|
| 0 | 0 | 0 | 0 | 0 | 0 |
|
| 0 | 0 | 0 | 0 | 0 | 0 |
aData represented as number (percentage).
bGroup (A): tourniquet inflation pressure was based on the AOP, Group (B): tourniquet inflation pressure was being based on LOP.
cP value of < 0.05 was considered statistically significantly different