Beam monitor calibration of a synchrotron-based scanned light-ion beam delivery system.

Jhonnatan Osorio1, Ralf Dreindl2, Loïc Grevillot2, Virgile Letellier2, Peter Kuess3, Antonio Carlino2, Alessio Elia2, Markus Stock2, Stanislav Vatnitsky2, Hugo Palmans4.   

Abstract

PURPOSE: This paper presents the implementation and comparison of two independent methods of beam monitor calibration in terms of number of particles for scanned proton and carbon ion beams.
METHODS: In the first method, called the single-layer method, dose-area-product to water (DAPw) is derived from the absorbed dose to water determined using a Roos-type plane-parallel ionization chamber in single-energy scanned beams. This is considered the reference method for the beam monitor calibration in the clinically relevant proton and carbon energy ranges. In the second method, called the single-spot method, DAPw of a single central spot is determined using a Bragg-peak (BP) type large-area plane-parallel ionization chamber. Emphasis is given to the detailed characterization of the ionization chambers used for the beam monitor calibration. For both methods a detailed uncertainty budget on the DAPw determination is provided as well as on the derivation of the number of particles.
RESULTS: Both calibration methods agreed on average within 1.1% for protons and within 2.6% for carbon ions. The uncertainty on DAPw using single-layer beams is 2.1% for protons and 3.1% for carbon ions with major contributions from the available values of kQ and the average spot spacing in both lateral directions. The uncertainty using the single-spot method is 2.2% for protons and 3.2% for carbon ions with major contributions from the available values of kQ and the non-uniformity of the BP chamber response, which can lead to a correction of up-to 3.2%. For the number of particles, an additional dominant uncertainty component for the mean stopping power per incident proton (or the CEMA) needs to be added.
CONCLUSION: The agreement between both methods enhances confidence in the beam monitor calibration and the estimated uncertainty. The single-layer method can be used as a reference and the single-spot method is an alternative that, when more accumulated knowledge and data on the method becomes available, can be used as a redundant dose monitor calibration method. This work, together with the overview of information from the literature provided here, is a first step towards comprehensive information on the single-spot method.
Copyright © 2020. Published by Elsevier GmbH.

Entities:  

Keywords:  Beam monitor calibration; Carbon ions; Commissioning; Dose area product; Light ion beam therapy; Protons

Year:  2020        PMID: 32747175     DOI: 10.1016/j.zemedi.2020.06.005

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  1 in total

1.  Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams.

Authors:  Andreas F Resch; Fatima Padilla Cabal; Milovan Regodic; Wolfgang Lechner; Gerd Heilemann; Peter Kuess; Dietmar Georg; Hugo Palmans
Journal:  Med Phys       Date:  2022-07-14       Impact factor: 4.506

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.