Abstract
Quartz crystal microbalance (QCM) is a highly sensitive mass sensor and has been widely used in many fields. However, the nonuniform distribution of mass sensitivity will lead to poor reproducibility of QCM, which is not conducive to the application of QCM in some fields. Considering the effect of electrode shape, size, and material on mass sensitivity distribution, we found that for an AT-cut QCM with a fundamental frequency of 10 MHz, when the inner and outer diameters of silver ring electrode and the electrode loading factor are 2 and 5 mm and 0.0033, respectively, an approximately uniform mass sensitivity distribution can be obtained. The plating experiment in which rigid silver films were plated on the surface of electrode verified the uniformity. The uniform mass sensitivity distribution will make the application of QCM more convenient; the reproducibility can also be improved. This design of QCM will enrich QCM products and facilitate the application of QCM in various fields.Entities:
Year: 2020 PMID: 32746208 DOI: 10.1109/TUFFC.2020.3008790
Source DB: PubMed Journal: IEEE Trans Ultrason Ferroelectr Freq Control ISSN: 0885-3010 Impact factor: 2.725