J M Hamilton-Miller, W Brumfitt, S J Brooks, M Bakhtiar.
Abstract
The Microstix system has been tested, in parallel with dipslides, for suitability as a screening technique for the detection of bacteriuria. Results obtained using each method on 228 urine specimens were assessed by comparison with the results of quantitative counts. Dipslides had a higher sensitivity (86-7%) than Microstix. The best specificity (100%) was obtained from those results (77-6%) where both components of the Microstix system were in agreement. Advantages of Microstix are its greater shelf-life and its compactness.Entities:
Mesh:
Year: 1977 PMID: 327452 PMCID: PMC2496566 DOI: 10.1136/pgmj.53.619.248
Source DB: PubMed Journal: Postgrad Med J ISSN: 0032-5473 Impact factor: 2.401