| Literature DB >> 32742185 |
Christian Mbohou Nchetnkou1, Hervé Nyabeyeu Nyabeyeu1, Loick P Kojom Foko1, Leopold G Lehman1,2.
Abstract
BACKGROUND: Malaria has a negative impact on the activities of companies in endemic countries especially in Cameroon. In this regard, an increasingly growing number of companies have started to include management of malarious patients in their health policies. In the present study, we will evaluate the diagnostic performances of a fluorescence microscopy (FM), Cyscope® microscope, in the detection of malaria parasites.Entities:
Keywords: Cameroon; Company; Diagnosis performances; Fluorescence microscopy; Malaria
Year: 2020 PMID: 32742185 PMCID: PMC7385968 DOI: 10.1186/s41182-020-00234-7
Source DB: PubMed Journal: Trop Med Health ISSN: 1348-8945
Fig. 1a–c Cyscope® microscope (Partec-Sysmex, Japan). d Positive malaria slides under observation with CyScope®. Bigger spots correspond to WBC and the smaller spots to malaria parasites (Photographs are provided by the authors)
Baseline characteristics of study population
| Parameters | Frequencies (%) or mean ± SD |
|---|---|
| CON | 279 (36.9) |
| BPT | 163 (63.1) |
| [19–36[ | 210 (47.5) |
| [36–60[ | 200 (45.2) |
| ≥ 60 | 32 (7.2) |
| Mean age (SD) | 37.8 (9.7) |
| Male:female ratio | 3.86:1 |
| Yes | 56 (12.7) |
| No | 386 (87.3) |
| Primary | 46 (10.4) |
| Secondary | 158 (35.7) |
| University | 128 (29.0) |
| Missing data | 110 (24.9) |
| Workers | 261 (59.0) |
| Managers | 71 (16.1) |
| Missing data | 110 (24.9) |
| Douala 1 | 20 (4.5) |
| Douala 2 | 17 (3.8) |
| Douala 3 | 208 (47.1) |
| Douala 4 | 15 (3.4) |
| Douala 5 | 66 (14.9) |
| Out of Douala | 1 (0.2) |
| Missing data | 115 (26) |
SD Standard deviation
Malariometric variables using light microscopy
| Factors | |
|---|---|
| 72 (96) | |
| 3 (4.0) | |
| Gametocytes | 51 (68.0) |
| Trophozoites | 34 (45.3) |
| Schizonts | 1 (1.3) |
| Low (1–500) | 26 (35.1) |
| Moderate (501–5000) | 45 (60.8) |
| High (> 5000) | 3 (4.1) ) |
Cyscope® FM results compared to LM results with respect to clinical status
| Cyscope® FM | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| All | Febrile | Afebrile | ||||||||
| Positive | Negative | Total | Positive | Negative | Total | Positive | Negative | Total | ||
| Positive | 63 | 12 | 75 | 4 | 3 | 7 | 59 | 9 | 68 | |
| Negative | 110 | 256 | 366 | 17 | 32 | 49 | 93 | 224 | 317 | |
FM Fluorescence microscopy, LM Light microscopy
Data are presented as frequency
aParasitological result was missing for one individual
Diagnostic performances of Cyscope® FM with regard to clinical status
| Parameters | All ( | Febrile ( | Afebrile ( | ||||
|---|---|---|---|---|---|---|---|
| Value | 95% CI | Value | 95% CI | Value | 95% CI | ||
| Se (%) | 84.00 | 74.08–90.60 | 57.14 | 25.04–84.18 | 86.76 | 76.71–92.87 | 0.74 |
| Sp (%) | 69.95 | 65.07–74.42 | 65.3 | 51.32–77.08 | 70.66 | 65.42–75.40 | 0.92 |
| PPV (%) | 63.58 | 56.19–70.38 | 19.05 | 7.67–40.0 | 38.81 | 31.44–46.75 | 0.3 |
| NPV (%) | 95.55 | 88.31–94.72 | 91.43 | 77.62–97.04 | 96.13 | 92.83–97.96 | 1 |
| PLR | 2.8 | – | 1.65 | NA | 2.95 | NA | – |
| NLR | 0.23 | – | NA | NA | NA | NA | – |
| Accuracy (%) | 89.36 | 83.34–90.21 | 10.08 | 51.2–75.85 | 79.27 | 68.89–77.67 | 0.63 |
| Reliability (%) | 53.95 | – | 22.45 | – | 57.43 | – | – |
| Kappa | 0.45 | – | – | – | - | – | – |
CI Confidence interval, FM Fluorescence microscopy, Se Sensitivity, Sp Specificity, PPV Positive predictive value, NPV Negative predictive value, PLR Positive likelihood ratio, NLR Negative likelihood ratio, NA Not applicable. Pearson’s chi-square test was used to make comparisons. Significance was set at p value < 0.05