Masahiro Yuasa1, Hiromasa Kurosaki1,2.
Abstract
BACKGROUND: There are very few studies on noncoplanar radiation in tomotherapy because deformable image registration is not implemented in the TomoTherapy Planning Station, a treatment planning device used in tomotherapy. This study examined whether noncoplanar radiation can be performed on the head using a tilt-type head and neck fixture and deformable image registration.Entities:
Keywords: brain tumor; deformable image registration; noncoplanar radiation; stereotactic radiotherapy; tomotherapy
Year: 2020 PMID: 32734825 PMCID: PMC7406918 DOI: 10.1177/1533033820945776
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
Figure 1.Dose distribution map at a tilt angle of 0° (A). Deformable image registration was performed on the dose distribution map of the computed tomography image with a tilt angle of 40° (B) and that with a tilt angle of 0° (C). (D) is a 1:1 comparison between the dose distribution maps of (A) and (C). All images were for 2-cm spheres of planning target volume.
Ratio of the Phantom Volume to the Irradiation Dose.
| 10% | 20% | 30% | 40% | 50% | 60% | 70% | 80% | 90% | |
|---|---|---|---|---|---|---|---|---|---|
| 2-cm PTV Sphere. | |||||||||
| 1. 0° | 13.90 | 5.56 | 2.54 | 1.48 | 1.01 | 0.74 | 0.57 | 0.43 | 0.30 |
| 2. 40° | 15.01 | 5.29 | 2.64 | 1.54 | 1.04 | 0.77 | 0.59 | 0.44 | 0.30 |
| 3. 0° + 40° | 11.11 | 4.15 | 2.19 | 1.38 | 0.97 | 0.73 | 0.56 | 0.42 | 0.29 |
| 3-cm PTV Sphere | |||||||||
| 1. 0° | 22.43 | 11.50 | 5.90 | 3.45 | 2.33 | 1.73 | 1.36 | 1.07 | 0.79 |
| 2. 40° | 23.36 | 13.32 | 6.48 | 3.75 | 2.52 | 1.87 | 1.14 | 1.16 | 0.88 |
| 3. 0° + 40° | 21.00 | 9.12 | 5.10 | 3.24 | 2.29 | 1.74 | 1.38 | 1.09 | 0.81 |
| 4-cm PTV Sphere | |||||||||
| 1. 0° | 29.20 | 18.66 | 11.00 | 6.64 | 4.49 | 3.35 | 2.65 | 2.13 | 1.65 |
| 2. 40° | 29.88 | 20.27 | 11.64 | 6.68 | 4.58 | 3.41 | 2.70 | 2.15 | 1.64 |
| 3. 0° + 40° | 30.88 | 15.04 | 9.41 | 6.07 | 4.28 | 3.26 | 2.60 | 2.09 | 1.60 |
Abbreviation: PTV, planning target volume.
Details of Dose Parameter (D 95 = 14 Gy).
| CI | HI |
|
|
| DSC | |
|---|---|---|---|---|---|---|
| 2-cm PTV Sphere | ||||||
| 1. 0° | 2.27 | 0.039 | 13.80 | 14.20 | 14.41 | 14.37 |
| 2. 40° | 2.27 | 0.036 | 13.68 | 14.24 | 14.41 | 14.35 |
| 3. 0° + 40° | 2.17 | 0.037 | 13.85 | 14.22 | 14.40 | 14.36 |
| 3-cm PTV Sphere | ||||||
| 1. 0° | 1.94 | 0.031 | 13.79 | 14.19 | 14.38 | 14.22 |
| 2. 40° | 2.16 | 0.019 | 13.99 | 14.15 | 14.30 | 14.07 |
| 3. 0° + 40° | 1.97 | 0.020 | 13.94 | 14.17 | 14.29 | 14.14 |
| 4-cm PTV Sphere | ||||||
| 1. 0° | 1.75 | 0.024 | 13.88 | 14.14 | 14.32 | 14.14 |
| 2. 40° | 1.73 | 0.019 | 13.88 | 14.09 | 14.25 | 14.08 |
| 3. 0° + 40° | 1.67 | 0.021 | 13.91 | 14.12 | 14.25 | 14.11 |
Abbreviations: CI, conformity index; DSC, dose of sphere center; HI, Homogeneity index; PTV, planning target volume.