| Literature DB >> 32722397 |
Daniele Bibbo1, Tomas Klinkovsky2, Marek Penhaker2, Petr Kudrna3, Lukas Peter2, Martin Augustynek2, Vladimír Kašík2, Jan Kubicek2, Ali Selamat4, Martin Cerny2, Daniel Bielcik2.
Abstract
In this paper, a new approach for the periodical testing and the functionality evaluation of a fetal heart rate monitor device based on ultrasound principle is proposed. The design and realization of the device are presented, together with the description of its features and functioning tests. In the designed device, a relay element, driven by an electric signal that allows switching at two specific frequencies, is used to simulate the fetus and the mother's heartbeat. The simulator was designed to be compliant with the standard requirements for accurate assessment and measurement of medical devices. The accuracy of the simulated signals was evaluated, and it resulted to be stable and reliable. The generated frequencies show an error of about 0.5% with respect to the nominal one while the accuracy of the test equipment was within ±3% of the test signal set frequency. This value complies with the technical standard for the accuracy of fetal heart rate monitor devices. Moreover, the performed tests and measurements show the correct functionality of the developed simulator. The proposed equipment and testing respect the technical requirements for medical devices. The features of the proposed device make it simple and quick in testing a fetal heart rate monitor, thus providing an efficient way to evaluate and test the correlation capabilities of commercial apparatuses.Entities:
Keywords: cardiotocograph; doppler effect; fetal heart rate; fetal heart rate monitor device; heart movement simulator; tests of medical device
Year: 2020 PMID: 32722397 PMCID: PMC7436177 DOI: 10.3390/s20154139
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1The general scheme of Doppler measuring instruments for heart rate.
Figure 2Block diagram of FHRS.
Figure 3The general scheme of the electrical circuit for the realized FHRS.
Figure 4The scheme of the oscilloscope test.
Figure 5Positioning of the ultrasound CTG probe on FHRS actuator-relay.
Figure 6Displayed waveform on the oscilloscope with the FHRS switch in Position “1”.
Figure 7Displayed waveform on the oscilloscope with the FHRS switch in Position “2”.
Figure 8Screenshot from oscilloscope at the FHRS switch position 0. There are visible switched frequencies 1.56/3.144 of simulated signal every 21 s. The cursors were placed on signal period x and recalculated to frequency as 1/∆x: (a) measured f = 1.56 Hz; and (b) measured f = 3.144).
Measured parameters of FHRS on oscilloscope.
| Switch Position | 1 | 2 | 0 |
|---|---|---|---|
| binary counter settings (input @50 Hz) | 640 ms (1.56 Hz) | 320 ms (3.125 Hz) | 320 ms and 640 ms |
| Measured frequency (Hz) | 1.562 ± 0.001 | 3.144 ± 0.001 | 1.562 ± 0.001 and 3.144 ± 0.001 |
| correspondent BPM (beat/min.) | 93.72 | 188.6 | 93.72 and 188.6 |
Figure 9Displayed waveform on the oscilloscope with the FHRS switch in Position “0”. The yellow curve represents the signal measured on primary coil of relay at set on frequencies. The green curve shows signal measured on switched electrical part of relay, which is realized by mechanical contact of relay tab.
Figure 10Detail of Figure 10, representing the transition phase when the relay is excited.
The comparison outputs of the designed mechanical heart rate mechanical simulator device and settings values.
| Switch Position (Simulated BPM) | |||
|---|---|---|---|
| Ctg Device and Corresponding Average Measured Bpm | 1 | 2 | 0 |
| CADENCE II | 95 | 188 | 94 |
| BFM 800 | 93 | 187 | 93 |
| OCT Baby iFM | 94 | 186 | 94 |
Figure 11Final prototype of the fetal heart rate simulator device: (a) fetal heart rate simulator device. and (b) relay and US probe during simulation tests.