Yunho Kim1, Jeeyoung Kim1, Heesung Nam1, Hyun Dong Kim1, Mi Ja Eom1, Sang Hoon Jung1, Nami Han1.
Abstract
OBJECTIVE: To examine the correlation between ultrasonographic trunk muscle parameters and balance scales in mild acute stroke patients.Entities:
Keywords: Abdominal muscles; Paraspinal muscles; Postural balance; Stroke; Ultrasonography
Year: 2020 PMID: 32721990 PMCID: PMC7463119 DOI: 10.5535/arm.19125
Source DB: PubMed Journal: Ann Rehabil Med ISSN: 2234-0645
Fig. 1.Flow chart of the study design. A total of 55 patients were selected as study participants. BBS, Berg Balance Scale.
Fig. 2.Ultrasonographic images of the RA muscle (A), lateral abdominal muscle (B), and ES muscle (C). Thickness was measured in the resting and contraction states. RA, rectus abdominis; EO, external oblique; IO, internal oblique; TrA, transversus abdominis; ES, erector spinae.
Demographics and characteristics of patients (n=55)
| Characteristic | High-risk fall group (n=19) | Low-risk fall group (n=36) |
|---|---|---|
| Sex | ||
| Male | 10 | 24 |
| Female | 9 | 12 |
| Age (yr) | 70.00±15.28 | 61.97±12.32 |
| BMI (kg/m2) | 22.45±1.87 | 23.91±2.86 |
| Duration from stroke onset to examination (day) | 9.53±9.82 | 12.44±10.99 |
| Hospitalization period (day) | 17.89±12.14 | 14.86±11.87 |
| Stroke type and lesion | ||
| Infarction | 15 | 27 |
| Cerebellum | 0 | 4 |
| BG or capsula interna | 2 | 0 |
| Thalamus | 1 | 3 |
| Pontine and medullar lesion | 2 | 1 |
| Multiple lesion | 10 | 19 |
| Hemorrhage | 4 | 9 |
| Side of lesion | ||
| Right | 12 | 22 |
| Left | 7 | 14 |
| Smoking | ||
| Yes | 1 | 11 |
| No | 18 | 25 |
| Cardiovascular disease | ||
| Yes | 1 | 4 |
| No | 18 | 32 |
| Left ventricular hypertrophy | ||
| Yes | 0 | 1 |
| No | 19 | 35 |
| Hypertension | ||
| Yes | 9 | 16 |
| No | 10 | 20 |
| Diabetes mellitus | ||
| Yes | 5 | 7 |
| No | 14 | 29 |
| Atrial fibrillation | ||
| Yes | 0 | 4 |
| No | 19 | 32 |
| NIHSS on admission | 4.11±2.10 | 3.06±3.96 |
| mRS on admission | 2.58±0.92 | 1.53±1.44 |
| SARA score | 8.61±4.43 | 2.46±2.14 |
| BBS | 38.42±7.75 | 51.31±2.93 |
| TUG test | 24.29±8.84 | 12.81±53.7 |
| TCT | 95.21±10.76 | 100±0.00 |
Values are presented as mean±standard deviation or as number (%).
BMI, body mass index; BG, basal ganglia; NIHSS, National Institutes of Health Stroke Scale; mRS, modified Rankin Scale; SARA, Scale for the Assessment and Rating of Ataxia; BBS, Berg Balance Scale; TUG, Timed Up and Go Test; TCT, Trunk Control Test.
Ultrasonographic values for the paretic and non-paretic side of trunk muscles
| Non-paretic | Paretic | p-value | |
|---|---|---|---|
| Rectus abdominis | |||
| Resting thickness (mm) | 7.60±2.26 | 7.49±2.42 | 0.337 |
| Contraction thickness (mm) | 8.95±2.54 | 8.22±2.64 | <0.001[ |
| Contractility | 1.19±0.12 | 1.10±0.10 | <0.001[ |
| External oblique | |||
| Resting thickness (mm) | 3.60±1.22 | 3.56±1.04 | 0.675 |
| Contraction thickness (mm) | 4.54±1.43 | 4.14±1.25 | <0.001[ |
| Contractility | 1.28±0.19 | 1.16±0.14 | <0.001[ |
| Internal oblique | |||
| Resting thickness (mm) | 6.70±2.21 | 6.48±2.09 | 0.042[ |
| Contraction thickness (mm) | 8.68±2.71 | 7.74±2.49 | <0.001[ |
| Contractility | 1.30±0.15 | 1.20±1.32 | <0.001[ |
| Transversus abdominis | |||
| Resting thickness (mm) | 3.16±1.35 | 3.00±1.29 | 0.047[ |
| Contraction thickness (mm) | 4.99±2.09 | 4.26±1.95 | <0.001[ |
| Contractility | 1.62±0.49 | 1.43±0.43 | <0.001[ |
| Erector spinae | |||
| Resting thickness (mm) | 37.51±6.19 | 36.71±6.12 | 0.045[ |
| Contraction thickness (mm) | 44.47±6.56 | 41.40±7.00 | <0.001[ |
| Contractility | 1.19±0.11 | 1.13±0.09 | <0.001[ |
Values are presented as mean±standard deviation.
p<0.05 by paired t-test.
Correlation coefficients between ultrasonographic trunk muscle parameters and balance scales in stroke patients
| Pearson correlation coefficient | ||||
|---|---|---|---|---|
| SARA | BBS | TUG | TCT | |
| Rectus abdominis | ||||
| NPS thickness | -0.184 | 0.191 | -0.221 | 0.101 |
| PS thickness | -0.168 | 0.184 | -0.240 | -0.221 |
| NPS contractility | 0.431[ | -0.434[ | 0.340[ | -0.371[ |
| PS contractility | 0.067 | -0.065 | 0.068 | 0.058 |
| Contractility ratio | 0.529[ | -0.534[ | 0.393[ | -0.605[ |
| External oblique | ||||
| NPS thickness | -0.263 | 0.309[ | -0.385[ | 0.118 |
| PS thickness | -0.342 | 0.355 | -0.420 | -0.213 |
| NPS contractility | 0.212 | -0.212 | 0.089 | -0.232 |
| PS contractility | -0.222 | 0.194 | -0.241 | 0.175 |
| Contractility ratio | 0.562[ | -0.533[ | 0.414[ | -0.542[ |
| Internal oblique | ||||
| NPS thickness | -0.115 | 0.161 | -0.220 | 0.060 |
| PS thickness | -0.130 | 0.192 | -0.274 | 0.077 |
| NPS contractility | 0.497[ | -0.491[ | 0.439[ | -0.395[ |
| PS contractility | -0.083 | 0.100 | -0.066 | 0.063 |
| Contractility ratio | 0.659[ | -0.673[ | 0.568[ | -0.541[ |
| Transversus abdominis | ||||
| NPS thickness | -0.118 | 0.091 | -0.119 | 0.200 |
| PS thickness | -0.131 | 0.102 | -0.149 | 0.192 |
| NPS contractility | -0.029 | 0.063 | -0.031 | -0.110 |
| PS contractility | -0.237 | 0.247 | -0.215 | 0.070 |
| Contractility ratio | 0.374[ | -0.337[ | 0.311[ | -0.320[ |
| Erector spinae | ||||
| NPS thickness | -0.313[ | 0.392[ | -0.298[ | 0.134 |
| PS thickness | -0.236 | 0.306[ | -0.243 | 0.035 |
| NPS contractility | 0.170 | -0.092 | 0.014 | 0.024 |
| PS contractility | -0.237 | 0.301[ | -0.354[ | 0.191 |
| Contractility ratio | 0.562[ | -0.533[ | 0.414[ | -0.542[ |
Value of NPS thickness and PS thickness was measured on resting state.
NPS, non-paretic side; PS, paretic side; SARA, Scale for the Assessment and Rating of Ataxia; BBS, Berg Balance Scale; TUG, Timed Up and Go Test; TCT, Trunk Control Test.
p<0.05,
p<0.01 by Pearson correlation coefficient analysis.
Comparison of ultrasonographic trunk muscle parameters between the high- and low-risk fall groups classified according to Berg Balance Scale
| High-risk fall group (n=19) | Low-risk fall group (n=36) | p-value | |
|---|---|---|---|
| Rectus abdominis | |||
| NPS thickness (mm) | 6.94±2.20 | 7.95±2.23 | 0.117 |
| PS thickness (mm) | 6.86±2.45 | 7.81±2.36 | 0.103 |
| NPS contractility | 1.23±0.13 | 1.16±1.10 | 0.061 |
| PS contractility | 1.10±0.09 | 1.10±0.10 | 0.852 |
| Contractility ratio | 1.11±0.09 | 1.06±0.06 | 0.001[ |
| External oblique | |||
| NPS thickness (mm) | 3.09±1.27 | 3.86±1.12 | 0.011[ |
| PS thickness (mm) | 3.12±1.08 | 3.79±0.95 | 0.008[ |
| NPS contractility | 1.32±0.19 | 1.26±0.19 | 0.193 |
| PS contractility | 1.15±0.13 | 1.17±0.15 | 0.710 |
| Contractility ratio | 1.15±0.16 | 1.07±0.08 | 0.048[ |
| Internal oblique | |||
| NPS thickness (mm) | 6.06±1.79 | 7.05±2.36 | 0.159 |
| PS thickness (mm) | 5.73±1.83 | 6.88±2.14 | 0.037[ |
| NPS contractility | 1.37±0.14 | 1.27±0.14 | 0.044[ |
| PS contractility | 1.17±0.13 | 1.20±0.13 | 0.381 |
| Contractility ratio | 1.16±0.14 | 1.05±0.07 | <0.001[ |
| Transversus abdominis | |||
| NPS thickness (mm) | 2.91±1.36 | 3.30±1.34 | 0.125 |
| PS thickness (mm) | 2.76±1.34 | 3.13±1.26 | 0.172 |
| NPS contractility | 1.57±3.55 | 1.65±0.51 | 0.979 |
| PS contractility | 1.28±0.27 | 1.51±0.47 | 0.026[ |
| Contractility ratio | 1.25±0.33 | 1.09±0.14 | <0.001[ |
| Erector spinae | |||
| NPS thickness (mm) | 33.7±4.27 | 39.5±6.15 | 0.001[ |
| PS thickness (mm) | 33.9±4.02 | 38.1±6.55 | 0.019[ |
| NPS contractility | 1.20±0.13 | 1.18±0.09 | 0.818 |
| PS contractility | 1.10±0.07 | 1.14±0.09 | 0.028[ |
| Contractility ratio | 1.10±0.07 | 1.04±0.04 | 0.002[ |
Values are presented as mean±standard deviation.
Value of NPS thickness and PS thickness was measured on resting state.
NPS, non-paretic side; PS, paretic side.
p<0.05,
p<0.01 by Mann-Whitney U-test.