Reflection Versus Rebreathing for Administration of Sevoflurane During Minor Gynecological Surgery.

Azzeddine Kermad1, Madeleine Appenzeller1, Emanuela Morinello1, Sven Oliver Schneider1, Stefan Kleinschmidt1, Brian O'Gara2, Thomas Volk1, Andreas Meiser1.   

Abstract

BACKGROUND: Contemporary anesthetic circle systems, when used at low fresh gas flows (FGF) to allow rebreathing of anesthetic, lack the ability for rapid dose titration. The small-scale anesthetic reflection device Anaesthetic Conserving Device (50mL Version; AnaConDa-S) permits administration of volatile anesthetics with high-flow ventilators. We compared washin, washout, and sevoflurane consumption using AnaConDa-S versus a circle system with low and minimal FGF.
METHODS: Forty patients undergoing breast surgery were randomized to receive 0.5 minimal alveolar concentration (MAC) sevoflurane with AnaConDa-S (21 patients, reflection group) or with a circle system (low flow: FGF = 0.2 minute ventilation [V'E], 9 patients; or minimal flow: 0.1 V'E, 10 patients). In the reflection group, syringe pump boluses were given for priming and washin; to simulate an open system, the FGF of the anesthesia ventilator was set to 18 L·min-1 with the soda lime removed. In the other groups, the FGF was increased for washin (1 V'E for 8 minutes) and washout (3 V'E). For all patients, tidal volume was 7 mL·kg-1 and the respiratory rate adjusted to ensure normoventilation. Analgesia was attained with remifentanil 0.3 µg·kg-1·min-1. Sevoflurane consumption was compared between the reflection group and the low- and minimal-flow groups, respectively, using a post hoc test (Fisher Least Significant Difference). To compare washin and washout (half-life), the low- and minimal-flow groups were combined.
RESULTS: Sevoflurane consumption was reduced in the reflection group (9.4 ± 2.0 vs 15.0 ± 3.5 [low flow, P < .001] vs 11.6 ± 2.3 mL·MAC h-1 [minimal flow, P = .02]); washin (33 ± 15 vs 49 ± 12 seconds, P = .001) and washout (28 ± 15 vs 55 ± 19 seconds, P < .001) times were also significantly shorter.
CONCLUSIONS: In this clinical setting with short procedures, low anesthetic requirements, and low tidal volumes, AnaConDa-S decreased anesthetic consumption, washin, and washout times compared to a circle system.
Copyright © 2021 International Anesthesia Research Society.

Entities:  

Mesh:

Substances:

Year:  2021        PMID: 32701542     DOI: 10.1213/ANE.0000000000005074

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108