Literature DB >> 32684399

Dissociation of Acetabular Polyethylene Liners With a Morse Taper Design.

Asa W Shnaekel1, Wesley H Mayes1, Jeffrey B Stambough1, Paul K Edwards1, Simon C Mears1, C Lowry Barnes1.   

Abstract

BACKGROUND: Polyethylene liner dissociation is an uncommon complication of hip replacement. Dissociation has been associated with particular acetabular component designs. This study reviewed acetabular liner dissociations in a specific modular cup with a Morse taper locking mechanism that has not been previously reported.
METHODS: The senior author performed 655 primary total hip arthroplasties with one particular design of acetabular component using Class A polyethylene liners and metal head articulation. Cases with revision surgery performed for acetabular liner dissociation were reviewed.
RESULTS: Seven of 655 patients with this cup underwent revision surgery for a dissociated liner. Liner dissociation occurred at a mean of 73 months postoperatively. Patients presented with new-onset hip pain or squeaking, 4 of which developed symptoms acutely. Two patients treated with polyethylene liner exchange into the same cup required a second revision surgery for recurrent dissociation.
CONCLUSION: Polyethylene liner dissociation is an infrequent but possible complication associated with modular acetabular components using a Morse taper locking. Providers should be vigilant with long-term follow-up of patients with this acetabular system for patient complaints of catching or squeaking. Patients treated for liner dissociation should not have a new liner placed into the same acetabular shell given the risk for further dissociation.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Dynasty acetabular system; MicroPort; Wright Medical; modular acetabular components; polyethylene liner dissociation

Year:  2020        PMID: 32684399     DOI: 10.1016/j.arth.2020.06.068

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  1 in total

1.  Liner dissociation in a large-diameter ceramic-bearing acetabular component: a report of five cases.

Authors:  Lazaros Kostretzis; Sagi Martinov; Martin Lavigne; Vincent Massé; Pascal-André Vendittoli
Journal:  BMC Musculoskelet Disord       Date:  2022-02-09       Impact factor: 2.362

  1 in total