Activity regimes for 82Rb cardiac PET: Effects on absolute MBF and MPI.

Camilla M Hoff1, Jens Sørensen2,3, Nana L Christensen1, Kirsten Bouchelouche1,2, Lars Tolbod4,5.   

Abstract

BACKGROUND: Selection of optimal tracer activity for 82Rb PET is based on a trade-off between necessary count-statistics in the late static phase and detector saturation in the early blood-pool phase. Administered tracer activity recommended in prescribing information differs substantially from recommendations in current literature. The present study examines the effect on both absolute myocardial blood flow (MBF), myocardial flow reserve (MFR) and relative myocardial perfusion imaging (MPI) of reducing dose.
METHODS: Forty patients were scanned twice on a PMT-based PET/CT (GE D690): At recommended activity (1110 MBq) and at either 740 or 370 MBq. MBF, MFR, total perfusion deficit (TPD) and ejection fractions (EF) were quantified. Results were compared using linear regression and Bland-Altman plots.
RESULTS: Linear correlation between MBF at 1110 MBq at either reduced activity had an R2 > 0.98. A small bias (± 5%-9%) was observed with opposite signs for 1110/740 and 1110/370. Limits of agreement for MBF were larger for 1110/370. MFR had a lower linear correlation (R2 = 0.96), but wide limits of agreement especially for 1110/370. TPD and EF correlated well at 1110/740 (R2 = 0.96 and 0.99, respectively), but large scatter was observed for 1110/370.
CONCLUSION: Reduction of the tracer activity to 740 MBq, significantly reduced dead-time correction factors, while still producing reliable static and gated images. However, despite large dead-time at 1110 MBq, no systematic bias on absolute MBF was observed compared to reduced activities.
© 2020. American Society of Nuclear Cardiology.

Entities:  

Keywords:  MPI; PET; image analysis; instrumentation; myocardial blood flow

Mesh:

Substances:

Year:  2020        PMID: 32676911     DOI: 10.1007/s12350-020-02266-2

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  2 in total

1.  Body weight-dependent Rubidium-82 activity results in constant image quality in myocardial perfusion imaging with PET.

Authors:  J D van Dijk; M Dotinga; P L Jager; C H Slump; J P Ottervanger; M Mouden; J A van Dalen
Journal:  J Nucl Cardiol       Date:  2019-09-04       Impact factor: 5.952

2.  Automated dynamic motion correction using normalized gradient fields for 82rubidium PET myocardial blood flow quantification.

Authors:  Benjamin C Lee; Jonathan B Moody; Alexis Poitrasson-Rivière; Amanda C Melvin; Richard L Weinberg; James R Corbett; Venkatesh L Murthy; Edward P Ficaro
Journal:  J Nucl Cardiol       Date:  2018-11-07       Impact factor: 5.952

  2 in total
  2 in total

1.  The next step in improving (semi-)quantitative MPI PET.

Authors:  S S Koenders; J A van Dalen; J D van Dijk
Journal:  J Nucl Cardiol       Date:  2020-09-17       Impact factor: 5.952

2.  Effect of temporal sampling protocols on myocardial blood flow measurements using Rubidium-82 PET.

Authors:  S S Koenders; J D van Dijk; P L Jager; M Mouden; A G Tegelaar; C H Slump; J A van Dalen
Journal:  J Nucl Cardiol       Date:  2021-03-02       Impact factor: 3.872