| Literature DB >> 32676177 |
Wei-Ke Kuo1, Chih-Yu Huang2, Chung-Chieh Yu1,2, Chung-Ching Hua1,2.
Abstract
Background: A conventional centrally inserted central catheter (CICC) is frequently used to measure central venous pressure (CVP) to monitor the cardiocirculatory status of patients. The tip of the totally implanted port (TIP) is inserted at the same location in the superior vena cava as that of a CICC, and the TIP has been implanted in many patients with cancer. Measurements of CVP using CICC (CICCP) and TIP (TIPP) may be closely related. Material and Methods. Ten patients with TIPs in an intensive care unit were prospectively studied, and 121 records of 4536 paired CICCP and TIPP measurements were collected. A bench test in a static or dynamic setting was performed, and 598 paired measurements taken using CICC and TIP were recorded.Entities:
Mesh:
Year: 2020 PMID: 32676177 PMCID: PMC7346251 DOI: 10.1155/2020/9180856
Source DB: PubMed Journal: J Healthc Eng ISSN: 2040-2295 Impact factor: 2.682
Figure 1Bench test. (a) Static pressure was measured using an inverted T tube with a CICC connected to one side of the base and one of two TIPs connected to the other through a Huber needle set. Positive pressure and negative pressure were simulated by raising the water level in the vertical part of the inverted T tube and by placing the transducer above the central line of the horizontal part of the inverted T tube, respectively. (b) Dynamic pressure was generated using a piston device developed with an Arduino microcontroller (A). CICC (B) and one of two TIPs (C) were inserted through a stopper (D) into a bulb syringe (E) filled with water. A linear stepper (F) motor with a through-type lead screw (G) controlled by the Arduino microcontroller moved to push the bulb of the syringe to generate pressure at one end and was stopped by a microswitch (H) at the other end. The generated pressure waveforms had amplitudes proportional to the stroke length of the lead screw and frequencies changed by the interstroke pause. Abrupt pressure changes were simulated by adding or withdrawing water using a 10-mL syringe (I) attached to a three-way stopcock that had a tube independently connected to the bulb syringe through the stopper. The pressure in the bulb syringe was transmitted to the HP monitor (K) through the pressure transducers (J) that were connected to the CICC and one of the TIPs.
Figure 2Dynamic pressure waveforms recorded by the CICC (CICCP) and TIP (TIPP) were nearly identical in the simulations of abrupt pressure changes.
Characteristic of 10 patients with implanted TIP and CICC insertion.
| Age, years | 59.2 ± 13.4 | |
| Gender | ||
| Male | 8 | |
| Female | 2 | |
| Type of cancer | ||
| Solid | 10 | |
| Hematologic | 0 | |
| Intubated with endotracheal tube | ||
| Yes | 9 | |
| No | 1 | |
| TIP brand | ||
| Powerport | 6 | |
| Polysite | 4 | |
| Side of TIP implantation | ||
| Left | 4 | |
| Right | 6 | |
| Duration of TIP implantation, days | 167.7 ± 133.0 | |
| Side of internal jugular vein for CICC insertion | ||
| Left | 4 | |
| Right | 6 |
TIP, totally implanted port; CICC, centrally inserted central catheter.
The difference of MDP by the characteristics of patients.
| Records | MDP |
| ||
|---|---|---|---|---|
| Gender | 0.13 | |||
| Male | 99 | 1.56 ± 1.58 | ||
| Female | 22 | 2.10 ± 1.28 | ||
| TIP brand | 0.86 | |||
| Powerport | 70 | 1.68 ± 1.76 | ||
| Polysite | 51 | 1.63 ± 1.17 | ||
| Side of TIP implantation | <0.01 | |||
| Left | 34 | 2.54 ± 1.61 | ||
| Right | 87 | 1.31 ± 1.37 | ||
| Side of TIP versus CICC | 0.04 | |||
| Same | 45 | 1.28 ± 1.32 | ||
| Different | 76 | 1.88 ± 1.62 | ||
| In a calm state | <0.01 | |||
| Yes | 93 | 1.08 ± 0.91 | ||
| No | 28 | 3.58 ± 1.63 | ||
| Under sedation | <0.01 | |||
| Yes | 100 | 1.43 ± 1.37 | ||
| No | 21 | 2.74 ± 1.84 | ||
| Control/Assist-control ventilator mode | 0.02 | |||
| Yes | 52 | 1.84 ± 1.70 | ||
| No | 59 | 1.22 ± 1.07 | ||
Data are presented in mean ± standard deviation. CICC, centrally inserted central catheter; TIP, totally implanted port; MDP, the mean difference between pressure measured by CICC and TIP in each record.
Simple linear regression for MDP.
|
| 95% CI of |
| |
|---|---|---|---|
| Age | 0.00 | –0.03∼0.02 | 0.78 |
| Gender | 0.09 | –0.77∼0.96 | 0.83 |
| Side of TIP implantation | –0.62 | –1.66∼0.42 | 0.24 |
| TIP and CICC in the same side | –0.38 | –0.86∼0.11 | 0.13 |
| Control/Assist–control mode | –0.29 | –0.86∼0.27 | 0.30 |
| In a calm state | –2.31 | –2.87 ∼ –1.76 | <0.01 |
| Under sedation | –0.16 | –1.45∼1.12 | 0.80 |
| CVHR | 7.15 | –1.29∼15.59 | 0.10 |
| CVRR | –2.73 | –5.39∼ –0.08 | 0.04 |
β, unstandardized regression coefficient; CI, confidence interval; CICC, central inserted central catheter; TIP, totally implanted port; MDP, the mean difference between pressure measured by CICC and TIP in each record; CVHR, the coefficient of variation of heart rate; CVRR, the coefficient of variation of respiratory rate.
Figure 3The means and correlations of CICCP and TIPP values in each recording were stratified by the calm state of patients. CICCP and TIPP had the slope of regression line between their means closer to the diagonal line and had more significant correlations in the periods in which patients were in a calm state than those periods in which they were not. A closed circle denotes a significant correlation, and the size of the circle is proportional to the correlation coefficient.
The difference of and the correlations between CICCP and TIPP in the records with subjects in a calm state.
| Count of identical values in LP | Record numbers | Percentage of difference between CICCP and TIPP ≤2 mmHg (%) | Spearman's rank correlation | |
|---|---|---|---|---|
| Rho |
| |||
| ≥3 | 1185 | 90.0 | 0.733 | <0.001 |
| ≥4 | 888 | 89.3 | 0.694 | <0.001 |
| ≥5 | 672 | 86.3 | 0.701 | <0.001 |
| ≥6 | 517 | 88.6 | 0.706 | <0.001 |
The data count of the longest episode (LP) in which each of CICCP and TIPP had an individual series of identical values. Series of CICCP and TIPP were included in the analysis if their counts of identical values in LP were equal to or greater than the number listed in the column. CICCP, pressure measured by central inserted central catheter; TIPP, pressure measured by totally implanted port; LP, longest period.