Literature DB >> 32675651

Precision pulse capsulotomy: performance metrics and utility in routine and complex cases.

Cristos Ifantides1, Junhun Lee, Rajy Rouweyha, Mark Vital, David Sretavan.   

Abstract

PURPOSE: To evaluate precision pulse capsulotomy (PPC) performance.
SETTING: University and private practice in the United States and South Korea.
DESIGN: Multicenter retrospective analysis.
METHODS: The surgical videos of 337 cataract surgeries with PPC capsulotomy performed by 4 surgeons at 4 centers were used to assess capsulotomy outcomes including completion rate, diameter, roundness (ovality), and quality of capsular overlap.
RESULTS: PPC use resulted in 99.4% free-floating capsulotomies from 337 cases. Video image analysis in a subset (n = 52) yielded a mean capsulotomy diameter of 5.0 mm ± 0.16 mm SD (95% CI, 4.96-5.04 mm). Capsulotomies were round to slightly oval at the end of the case with a mean ovality of 3.0% ± 2.86% (95% CI, 2.22%-3.78%; 360 degrees capsular overlap was obtained in 98% of cases. The offset of the capsulotomy center with the intraocular lens (IOL) optic center was 197 μm ± 122 μm (SD) (95% CI, 148-246 μm). PPC was used successfully in traumatic cataracts with compromised anterior and posterior capsule, phacodonesis, intumescent cataract with constricted pupil, and zonular dialysis and in penetrating keratoplasty with open-sky extracapsular cataract extraction.
CONCLUSIONS: Surgeons obtained good PPC capsulotomy outcomes in routine and challenging cases. Little variation was observed in achieving free-floating capsulotomies with approximately 5.0 mm diameter and complete capsular overlap. Variation was observed in the amount of offset between the capsulotomy center and the center of the IOL optic. PPC was useful in cases with multiple comorbidities that challenge capsulotomy performance.

Entities:  

Mesh:

Year:  2020        PMID: 32675651     DOI: 10.1097/j.jcrs.0000000000000318

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  Clinical Results After Precision Pulse Capsulotomy.

Authors:  Kjell G Gundersen; Richard Potvin
Journal:  Clin Ophthalmol       Date:  2020-12-29

2.  Fibrotic cataract as a sign of posterior capsule violation post vitrectomy.

Authors:  Eric R Williams; Jesse M Smith; Michael Huvard; Cristos Ifantides
Journal:  Am J Ophthalmol Case Rep       Date:  2021-06-30
  2 in total