Clinical application of an intelligent oropharyngeal swab robot: implication for the COVID-19 pandemic.

Shao-Qiang Li1,2, Wen-Liang Guo1,2, Hao Liu3,2, Tao Wang1, Yuan-Yuan Zhou3, Tao Yu3, Chong-Yang Wang3, Yong-Ming Yang3, Nan-Shan Zhong1, Nuo-Fu Zhang1,4, Shi-Yue Li5,4.   

Abstract

Entities:  

Mesh:

Year:  2020        PMID: 32675204      PMCID: PMC7366183          DOI: 10.1183/13993003.01912-2020

Source DB:  PubMed          Journal:  Eur Respir J        ISSN: 0903-1936            Impact factor:   16.671


× No keyword cloud information.
To the Editor: Coronavirus disease 2019 (COVID-19), caused by infection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), is transmitted through respiratory droplets and close contact [1-4]. To diagnose COVID-19, oropharyngeal swab (OP swab) sampling is widely used for viral nucleic acid detection [3]. However, healthcare workers who perform OP swab are at high risk of infection due to aerosol from patients during the process of sampling. And the quality of manual OP swabs is inconsistent among different collectors, which may lead to misdiagnosis [5]. Use of a remote-controlled OP swab robot has the potential to avoid close contact between healthcare workers with patients, and thus reduce the risk of SARS-CoV-2 infection during sampling. Here, we invented a robotic sampling (RS) system and evaluated the safety and efficacy of this system on OP swab sampling during the period of pandemic. A first-generation semi-automatic OP swab robot, jointly developed by Shenyang Institute of Automation of the Chinese Academy of Science and First Affiliated Hospital of Guangzhou Medical University, was used to take OP swabs in this study (figure 1a–d). The exposure of OP swab sampling using robot is classified according to the visible structures as follows [6]. Class I: soft palate, fauces, uvula, pillars; class II: soft palate, fauces, uvula; class III: soft palate, base of uvula; class IV: soft palate not visible at all. Class I was defined as good field of view (FOV) of posterior pharyngeal wall, and class II–IV was defined as limited FOV of posterior pharyngeal wall. Research protocol was registered (ChiCTR2000030861) and approved by the ethics committee of the First Affiliated Hospital of Guangzhou Medical University (2020-46). Suspected COVID-19 patients were determined based on the criteria of the National Health Commission of China [7].
FIGURE 1

Representative images of the oropharyngeal swab robot and its sampling processes. b) The force control mode was initiated, with the Force Ctrl state shown as “Start.” The swab moved forward automatically until it touched the posterior pharyngeal wall and reached the pre-set force range. The current force was 26.2, as shown in the upper right corner. d) The force control mode was stopped, with the Force Ctrl state shown as “Stop.” The robotic arm was controlled to withdraw the swab and retreat. e) Pathogen test results of oropharyngeal swabs collected with two different sampling methods in the 20 suspected coronavirus disease 2019 patients. RS: robotic sampling MS: manual sampling.

Representative images of the oropharyngeal swab robot and its sampling processes. b) The force control mode was initiated, with the Force Ctrl state shown as “Start.” The swab moved forward automatically until it touched the posterior pharyngeal wall and reached the pre-set force range. The current force was 26.2, as shown in the upper right corner. d) The force control mode was stopped, with the Force Ctrl state shown as “Stop.” The robotic arm was controlled to withdraw the swab and retreat. e) Pathogen test results of oropharyngeal swabs collected with two different sampling methods in the 20 suspected coronavirus disease 2019 patients. RS: robotic sampling MS: manual sampling. Sampling parameters, safety and effectiveness of OP swab robot were investigated in 20 healthy individuals using four different sampling forces (10–60 g). Effectiveness was determined in terms of sampling success rate (completion in first attempt), and swab quality was determined based on the threshold cycle (Ct) value for glyceraldehyde 3-phosphate dehydrogenase (GAPDH) by RT-PCR [8]. OP swabs with Ct values of ≤37 and >37 was considered as qualified and unqualified samples, respectively. Sampling adverse reactions were determined in terms of retching, sore throat, pharyngeal congestion, or injury of posterior pharyngeal wall after sampling. Comparison study of the safety and effectiveness between RS and manual sampling (MS) was performed in 30 healthy individuals in cross-over trials. Clinical application study was performed in 20 consecutive suspected COVID-19 patients from a fever clinic to compare the pathogen test results of the two methods. OP swab sample collection in the MS group was achieved through wiping of posterior pharyngeal wall with sterile cotton swab by experienced collectors. Data normality was tested via Kolmogorov–Smirnov test and presented as mean±sd or median (interquartile range). Two-sided independent sample t-test, one-way ANOVA or Fisher's exact test were used as appropriate; a p-value <0.05 was considered statistically significant. Sampling success rate was 95% in 20 healthy subjects with different force groups when using the RS system. The Ct values showed 97% specimens as qualified, and no significant difference in the quality of the specimens was found in the four sampling force groups based on the sample GADPH Ct value. (Group 10–30 g: 31.50±3.5; Group 20–40 g: 32.93±2.9; Group 30–50 g: 32.37±2.1; and Group 40–60 g: 32.97±2.1; p=0.346). There was no pharyngeal congestion or injury in any participant; however, incidence of sore throat was higher when the sampling force is >40 g (15.8%, 6/38) when compared to low-force (≤40 g) group (0/38; p=0.025). Thus, 20–40 g force was selected for OP swab robot sampling in subsequent experiments. Next, we compared RS and MS system and showed 95% and 100% success rates, respectively, when measuring GAPDH in the swab samples in the crossover study. All specimens were qualified based on GAPDH Ct values of ≤37; however, variation of Ct value from MS specimens was more significant compared to RS specimens (27.14±2.0 versus 27.86±2.02; p=0.008). However, the difference in Ct values from the data did not exceed 1, and although there was statistical difference, no clinical significance was indicated. No significant differences in adverse reactions were observed between the two groups. Swab quality was affected by limited FOV as Ct value of swab specimens with good FOV group was significantly lower than the limited FOV group (Ct value of good FOV group: 27.33±1.99; Ct value of limited FOV group: 28.61±1.87; p=0.018). There were seven and nine cases of limited FOV in the MS and RS groups, respectively. However, after eliminating the influence factor of limited FOV, no significant difference in swab quality was found between the MS and RS methods (27.05±2.0 versus 27.61±1.97; p=0.156). RS sampling time (9±4.1 s) was significantly shorter than MS group (18±5.1 s; p=0.032). There was no significant difference in adverse reactions between the RS and MS groups (10% versus 10%; p=1). Lastly, OP swabs were collected using MS and RS in 20 suspected COVID-19 patients, including eight males and 12 females with the mean age of 45±15 years. OP swab samples were subjected to pathogen testing. SARS-CoV-2 nucleic acid was negative for all suspected cases. The results of viral and mycoplasma tests were negative. There were two cases tested positive for bacteria, including one case of Staphylococcus infection and another case of Moraxella catarrhalis infection. The consistency of the pathogen discovery results between the two sampling methods for the 20 suspected cases of COVID-19 reached 95% (Kappa=0.875). No differences in the success rate, swab quality, and pathogen discovery results were found between the OP swab robot and manual sampling methods. Standard sampling of OP swab, including the precise delivery of swab to target tissue, appropriate force and touch avoidance of surrounding tissue, has confirmed to be prerequisite for the enhancement of positive detection rate [9-11]. MS swab sampling may potentially lead to false-negative results due to lack of experience and psychological anxiety of sampling personnel due to the contagious nature of SARS-CoV-2 [5]. OP swab robot with remote camera can help medical staffs to have clear vision during sampling without close contact to patients, thus resulting in collection of high-quality swab specimens consistently. Though greater force is believed to be capable of collecting higher number of cells per swab, our results showed that there was no correlation between the quality of swab specimens and sampling force. Thus, the RS system showed an advantage in the quality control of OP swab sampling, and have the potential to minimise the risk of infection to healthcare workers [11]. Time from research and development to application was pressed in light of the crisis of the emerging infectious disease. Therefore, the first generation of pharyngeal swab robots still required medical assistants to manually replace the isolation sleeves and swabs after collecting the specimens, which would minimise but not avoid the exposure to SARS-COV-2. However, ultraviolet disinfection and air disinfectant machine were performed for sterilisation after each specimen collection to avoid the exposure risk to the utmost extent. Additionally, in view of this situation, a second-generation intelligent robot has been developed to achieve intelligent movement and automatically replace the swab and protective cover, and is currently preparing for the next phase of clinical trials. It can further improve intelligence of the process and reduce the degree of manual participation. Though this was a preliminary, single-centre study in limited COVID-19 patients, the sampling process of the intelligent OP swab robot is safe, with a high success rate of sampling. The quality of OP swabs and the detection rate of pathogens were not inferior to those collected manually. The OP swab robot presents a safe and effective sampling method, and could avoid cross infection to operators. During epidemics or pandemics, the OP swab robot has good prospects for application in clinical practice. This one-page PDF can be shared freely online. Shareable PDF ERJ-01912-2020.Shareable
  9 in total

1.  Guideline to reference gene selection for quantitative real-time PCR.

Authors:  Aleksandar Radonić; Stefanie Thulke; Ian M Mackay; Olfert Landt; Wolfgang Siegert; Andreas Nitsche
Journal:  Biochem Biophys Res Commun       Date:  2004-01-23       Impact factor: 3.575

2.  Rapid detection of pathogenic virus genome sequence from throat and nasal swab samples using an exhaustive gene amplification method.

Authors:  Junji Hosokawa-Muto; Hiroki Sakai; Yukiko Sassa; Yoshihito Fujinami; Mai Kishimoto; Hiroaki Nakahara
Journal:  Forensic Sci Med Pathol       Date:  2019-06-27       Impact factor: 2.007

3.  Optimal site for throat culture: tonsillar surface versus posterior pharyngeal wall.

Authors:  E L van der Veen; E A M Sanders; W J M Videler; B K van Staaij; P P G van Benthem; A G M Schilder
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-05-04       Impact factor: 2.503

4.  Difficult tracheal intubation: a retrospective study.

Authors:  G L Samsoon; J R Young
Journal:  Anaesthesia       Date:  1987-05       Impact factor: 6.955

5.  Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China.

Authors:  Chaolin Huang; Yeming Wang; Xingwang Li; Lili Ren; Jianping Zhao; Yi Hu; Li Zhang; Guohui Fan; Jiuyang Xu; Xiaoying Gu; Zhenshun Cheng; Ting Yu; Jiaan Xia; Yuan Wei; Wenjuan Wu; Xuelei Xie; Wen Yin; Hui Li; Min Liu; Yan Xiao; Hong Gao; Li Guo; Jungang Xie; Guangfa Wang; Rongmeng Jiang; Zhancheng Gao; Qi Jin; Jianwei Wang; Bin Cao
Journal:  Lancet       Date:  2020-01-24       Impact factor: 79.321

6.  A familial cluster of pneumonia associated with the 2019 novel coronavirus indicating person-to-person transmission: a study of a family cluster.

Authors:  Jasper Fuk-Woo Chan; Shuofeng Yuan; Kin-Hang Kok; Kelvin Kai-Wang To; Hin Chu; Jin Yang; Fanfan Xing; Jieling Liu; Cyril Chik-Yan Yip; Rosana Wing-Shan Poon; Hoi-Wah Tsoi; Simon Kam-Fai Lo; Kwok-Hung Chan; Vincent Kwok-Man Poon; Wan-Mui Chan; Jonathan Daniel Ip; Jian-Piao Cai; Vincent Chi-Chung Cheng; Honglin Chen; Christopher Kim-Ming Hui; Kwok-Yung Yuen
Journal:  Lancet       Date:  2020-01-24       Impact factor: 79.321

7.  Epidemiological and clinical characteristics of 99 cases of 2019 novel coronavirus pneumonia in Wuhan, China: a descriptive study.

Authors:  Nanshan Chen; Min Zhou; Xuan Dong; Jieming Qu; Fengyun Gong; Yang Han; Yang Qiu; Jingli Wang; Ying Liu; Yuan Wei; Jia'an Xia; Ting Yu; Xinxin Zhang; Li Zhang
Journal:  Lancet       Date:  2020-01-30       Impact factor: 79.321

8.  Comparison among nasopharyngeal swab, nasal wash, and oropharyngeal swab for respiratory virus detection in adults with acute pharyngitis.

Authors:  Li Li; Qiao-Yan Chen; Yun-Ying Li; Yan-Fang Wang; Zi-Feng Yang; Nan-Shan Zhong
Journal:  BMC Infect Dis       Date:  2013-06-20       Impact factor: 3.090

9.  False-Negative Results of Real-Time Reverse-Transcriptase Polymerase Chain Reaction for Severe Acute Respiratory Syndrome Coronavirus 2: Role of Deep-Learning-Based CT Diagnosis and Insights from Two Cases.

Authors:  Dasheng Li; Dawei Wang; Jianping Dong; Nana Wang; He Huang; Haiwang Xu; Chen Xia
Journal:  Korean J Radiol       Date:  2020-03-05       Impact factor: 3.500

  9 in total
  13 in total

1.  Design and Control of a Highly Redundant Rigid-flexible Coupling Robot to Assist the COVID-19 Oropharyngeal-Swab Sampling.

Authors:  Yingbai Hu; Jian Li; Yongquan Chen; Qiwen Wang; Chuliang Chi; Heng Zhang; Qing Gao; Yuanmin Lan; Zheng Li; Zonggao Mu; Zhenglong Sun; Alois Knoll
Journal:  IEEE Robot Autom Lett       Date:  2021-02-25

2.  Robots Under COVID-19 Pandemic: A Comprehensive Survey.

Authors:  Yang Shen; Dejun Guo; Fei Long; Luis A Mateos; Houzhu Ding; Zhen Xiu; Randall B Hellman; Adam King; Shixun Chen; Chengkun Zhang; Huan Tan
Journal:  IEEE Access       Date:  2020-12-18       Impact factor: 3.367

3.  COVID-19 Isolation Control Proposal via UAV and UGV for Crowded Indoor Environments: Assistive Robots in the Shopping Malls.

Authors:  Muhammet Fatih Aslan; Khairunnisa Hasikin; Abdullah Yusefi; Akif Durdu; Kadir Sabanci; Muhammad Mokhzaini Azizan
Journal:  Front Public Health       Date:  2022-05-31

Review 4.  Review on the COVID-19 pandemic prevention and control system based on AI.

Authors:  Junfei Yi; Hui Zhang; Jianxu Mao; Yurong Chen; Hang Zhong; Yaonan Wang
Journal:  Eng Appl Artif Intell       Date:  2022-07-11       Impact factor: 7.802

5.  Tele-Operated Oropharyngeal Swab (TOOS) Robot Enabled by TSS Soft Hand for Safe and Effective Sampling.

Authors:  Wei Chen; Jianshu Zhou; Shing Shin Cheng; Yiang Lu; Fangxun Zhong; Yuan Gao; Yaqing Wang; Lingbin Xue; Michael C F Tong; Yun-Hui Liu
Journal:  IEEE Trans Med Robot Bionics       Date:  2021-10-27

6.  A Flexible Transoral Robot Towards COVID-19 Swab Sampling.

Authors:  Changsheng Li; Xiaoyi Gu; Xiao Xiao; Chwee Ming Lim; Xingguang Duan; Hongliang Ren
Journal:  Front Robot AI       Date:  2021-04-12

7.  Analysis of IgM, IgA, and IgG isotype antibodies Directed against SARS-CoV-2 spike glycoprotein and ORF8 in the course of COVID-19.

Authors:  Denise Meinberger; Manuel Koch; Annika Roth; Gabriele Hermes; Jannik Stemler; Oliver A Cornely; Thomas Streichert; Andreas R Klatt
Journal:  Sci Rep       Date:  2021-04-26       Impact factor: 4.379

Review 8.  Healthcare Workers' Point of View on Medical Robotics During COVID-19 Pandemic - A Scoping Review.

Authors:  Irma Ruslina Defi; Shelly Iskandar; Septiana Charismawati; Arjon Turnip; Dessy Novita
Journal:  Int J Gen Med       Date:  2022-04-06

9.  Telerobotic Operation of Intensive Care Unit Ventilators.

Authors:  Balazs P Vagvolgyi; Mikhail Khrenov; Jonathan Cope; Anton Deguet; Peter Kazanzides; Sajid Manzoor; Russell H Taylor; Axel Krieger
Journal:  Front Robot AI       Date:  2021-06-24

Review 10.  Immunological considerations for COVID-19 vaccine strategies.

Authors:  Mangalakumari Jeyanathan; Sam Afkhami; Fiona Smaill; Matthew S Miller; Brian D Lichty; Zhou Xing
Journal:  Nat Rev Immunol       Date:  2020-09-04       Impact factor: 53.106

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.